Assessment of combined use of ArcCheck® detector and portal dosimetry for delivery quality assurance of head and neck and prostate volumetric-modulated arc therapy

被引:9
作者
Moliner, Gilles [1 ,2 ]
Sorro, Lise [1 ,2 ]
Verstraet, Rodolfe [1 ,2 ]
Daviau, Paul Alexandre [1 ,2 ]
Casas, Melanie [1 ,2 ]
Piron, Berengere [1 ,2 ]
Dubois, Karine [1 ,2 ]
Debrigode, Charles [1 ]
Barrau, Corinne [2 ]
Bons, Francoise [1 ]
Greffier, Joel [2 ]
机构
[1] Nimes Univ Hosp, Dept Radiotherapy, Nimes, France
[2] Nimes Univ Hosp, Med Phys Unit, Nimes, France
关键词
ArcCheck; Gamma index; PDIP; VMAT pretreatment quality assurance; IMRT QA; TOMOTHERAPY; SENSITIVITY; VMAT;
D O I
10.1002/acm2.12460
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To assess the efficiency of combined use of ArcCheck((R)) detector (AC) and portal dosimetry (PDIP) for delivery quality assurance of head and neck and prostate volumetric-modulated arc therapy. Materials and methods Measurement processes were studied with the Gamma index method according to three analysis protocols. The detection sensitivity to technical errors of each individual or combined measurement processes was studied by inserting collimator, dose and MLC opening error into five head and neck and five prostate initial treatment plans. A total of 220 plans were created and 660 analyses were conducted by comparing measurements to error free planned dose matrix. Results For head and neck localization, collimator errors could be detected from 2 degrees for AC and 3 degrees for PDIP. Dose and MLC errors could be detected from 2% and 0.5 mm for AC and PDIP. Depending on the analysis protocol, the detection sensitivity of total simulated errors ranged from 54% to 88% for AC vs 40% to 74% for PDIP and 58% to 92% for the combined process. For the prostate localization, collimator errors could be detected from 4 degrees for AC while they could not be detected by PDIP. Dose and MLC errors could be detected from 3% and 0.5 mm for AC and PDIP. The detection sensitivity of total simulated errors ranged from 30% to 56% for AC vs 16% to 38% for PDIP and 30% to 58% for combined process. Conclusion The combined use of the two measurement processes did not statistically improve the detectability of technical errors compared to use of single process.
引用
收藏
页码:133 / 139
页数:7
相关论文
共 16 条
[1]   Tomotherapy treatment plan quality assurance: The impact of applied criteria on passing rate in gamma index method [J].
Bresciani, Sara ;
Di Dia, Amalia ;
Maggio, Angelo ;
Cutaia, Claudia ;
Miranti, Anna ;
Infusino, Erminia ;
Stasi, Michele .
MEDICAL PHYSICS, 2013, 40 (12)
[2]   How the detector resolution affects the clinical significance of SBRT pre-treatment quality assurance results [J].
Bruschi, A. ;
Esposito, M. ;
Pini, S. ;
Ghirelli, A. ;
Zatelli, G. ;
Russo, S. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 49 :129-134
[3]   Patient QA systems for rotational radiation therapy: A comparative experimental study with intentional errors [J].
Fredh, Anna ;
Scherman, Jonas Bengtsson ;
Fog, Lotte S. ;
Af Rosenschold, Per Munck .
MEDICAL PHYSICS, 2013, 40 (03)
[4]   On the sensitivity of common gamma-index evaluation methods to MLC misalignments in Rapidarc quality assurance [J].
Heilemann, G. ;
Poppe, B. ;
Laub, W. .
MEDICAL PHYSICS, 2013, 40 (03)
[5]   Institutional Patient-specific IMRT QA Does Not Predict Unacceptable Plan Delivery [J].
Kry, Stephen F. ;
Molineu, Andrea ;
Kerns, James R. ;
Faught, Austin M. ;
Huang, Jessie Y. ;
Pulliam, Kiley B. ;
Tonigan, Jackie ;
Alvarez, Paola ;
Stingo, Francesco ;
Followill, David S. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 (05) :1195-1201
[6]   A technique for the quantitative evaluation of dose distributions [J].
Low, DA ;
Harms, WB ;
Mutic, S ;
Purdy, JA .
MEDICAL PHYSICS, 1998, 25 (05) :656-661
[7]   Volumetric modulated arc therapy for nasopharyngeal carcinoma: A dosimetric comparison with TomoTherapy and step-and-shoot IMRT [J].
Lu, Szu-Huai ;
Cheng, Jason Chia-Hsien ;
Kuo, Sung-Hsin ;
Lee, Jason Jeun-Shenn ;
Chen, Liang-Hsin ;
Wu, Jian-Kuen ;
Chen, Yu-Lisuan ;
Chen, Wan-Yu ;
Wen, Shu-Yu ;
Chong, Fok-Ching ;
Wu, Chien-Jang ;
Wang, Chun-Wei .
RADIOTHERAPY AND ONCOLOGY, 2012, 104 (03) :324-330
[8]   Tolerance limits and methodologies for IMRT measurement-based verification QA: Recommendations of AAPM Task Group No. 218 [J].
Miften, Moyed ;
Olch, Arthur ;
Mihailidis, Dimitris ;
Moran, Jean ;
Pawlicki, Todd ;
Molineu, Andrea ;
Li, Harold ;
Wijesooriya, Krishni ;
Shi, Jie ;
Xia, Ping ;
Papanikolaou, Nikos ;
Low, Daniel A. .
MEDICAL PHYSICS, 2018, 45 (04) :E53-E83
[9]   Gamma-index method sensitivity for gauging plan delivery accuracy of volumetric modulated arc therapy [J].
Park, Jong In ;
Park, Jong Min ;
Kim, Jung-in ;
Park, So-Yeon ;
Ye, Sung-Joon .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (08) :1118-1122
[10]   IMRT QA: Selecting gamma criteria based on error detection sensitivity [J].
Steers, Jennifer M. ;
Fraass, Benedick A. .
MEDICAL PHYSICS, 2016, 43 (04) :1982-1994