Determination of cisapride in human plasma by high-performance liquid chromatography

被引:12
|
作者
Campanero, MA [1 ]
Calahorra, B
Garcia-Quetglas, E
Honorato, J
Carballal, JJ
机构
[1] Univ Navarra, Clin Univ, Dept Farmacol Clin, Pamplona, Spain
[2] Labs CEPA SL, Madrid, Spain
关键词
column liquid chromatography; cisapride in plasma; pharmacokinetics;
D O I
10.1007/BF02467491
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An HPLC method has been developed for the quantification of cisapride in plasma for pharmacokinetic studies. Clebopride was used as internal standard. Plasma samples were extracted at alkaline pH with tert-butyl methyl ether. The organic phase was then extracted with sulphuric acid to eliminate endogenous interferences, and cisapride and the internal standard were then extracted at alkaline pH into tert-butyl methyl ether. After evaporation of tert-butyl methyl ether, the residue was analysed by HPLC, Chromatography was performed at 20 degrees C on a 250 mm x 4 mm i.d. reversed-phase column selective for basic compounds. The isocratic mobile phase was 48:52 (v/v) acetonitrile-water containing 0.05 M potassium dihydrogen phosphate and 0.04 M triethylamine, adjusted to pH 5.5; the flow rate was 1 mL min(-1). Cisapride and the internal standard were detected by ultraviolet monitoring at 276 nm. The calibration graph was linear for quantities of cisapride from 1 to 200 ng mL(-1). Intra- and inter-day precision (CV) did not exceed 13.98%. The limit of quantitation (LOQ) was 0.68 ng mL(-1) for human plasma. The applicability of the method has been demonstrated in a pharmacokinetic study of normal volunteers who received 10 mg cisapride orally.
引用
收藏
页码:537 / 541
页数:5
相关论文
共 50 条
  • [1] Determination of cisapride in human plasma by high-performance liquid chromatography
    M. A. Campanero
    B. Calahorra
    E. García-Quetglás
    J. Honorato
    J. J. Carballal
    Chromatographia, 1998, 47 : 537 - 541
  • [2] Determination of cisapride in human plasma by high-performance liquid chromatography with fluorescence detection
    B. Dalmadi Kiss
    K. Balogh Nemes
    I. Klebovich
    Chromatographia, 2003, 57 : 47 - 50
  • [3] Determination of cisapride in human plasma by high-performance liquid chromatography with fluorescence detection
    Kiss, BD
    Nemes, KB
    Klebovich, I
    CHROMATOGRAPHIA, 2003, 57 (1-2) : 47 - 50
  • [4] Rapid determination of ranitidine in human plasma by high-performance liquid chromatography
    M. A. Campanero
    A. Lopez-Ocariz
    E. García-Quetglás
    B. Sádaba
    A. de la Maza
    Chromatographia, 1998, 47 : 391 - 395
  • [5] Rapid determination of ranitidine in human plasma by high-performance liquid chromatography
    Campanero, MA
    Lopez-Ocariz, A
    Garcia-Quetglas, E
    Sadaba, B
    de la Maza, A
    CHROMATOGRAPHIA, 1998, 47 (7-8) : 391 - 395
  • [6] DETERMINATION OF IVERMECTIN IN HUMAN PLASMA BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    KRISHNA, DR
    KLOTZ, U
    ARZNEIMITTEL-FORSCHUNG/DRUG RESEARCH, 1993, 43-1 (05): : 609 - 611
  • [7] An improved high-performance liquid chromatography method for the determination of homocysteine in human plasma
    Tröbs, M
    Renner, T
    Scherer, G
    CHROMATOGRAPHIA, 1998, 48 (7-8) : 506 - 510
  • [8] An improved high-performance liquid chromatography method for the determination of homocysteine in human plasma
    M. Tröbs
    T. Renner
    G. Scherer
    Chromatographia, 1998, 48 : 506 - 510
  • [9] DETERMINATION OF TRICHLORMETHIAZIDE IN HUMAN PLASMA AND URINE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    SAKATA, T
    MATSUURA, A
    KITAGAWA, T
    CHROMATOGRAPHIA, 1992, 33 (7-8) : 339 - 343
  • [10] Determination of Linezolid in Human Plasma by High-Performance Liquid Chromatography With Ultraviolet Detection
    Cattaneo, Dario
    Baldelli, Sara
    Conti, Francesca
    Cozzi, Valeria
    Clementi, Emilio
    THERAPEUTIC DRUG MONITORING, 2010, 32 (04) : 520 - 524