Modulation transfer function measurement of scanning reflectance microscopes

被引:15
|
作者
Wang, Zhao [1 ]
Glazowski, Christopher E. [2 ]
Zavislan, James M. [2 ]
机构
[1] Univ Rochester, Dept Biomed Engn, Rochester, NY 14627 USA
[2] Univ Rochester, Inst Opt, Rochester, NY 14627 USA
关键词
modulation transfer functions; resolution; scanning microscopy;
D O I
10.1117/1.2779352
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Real-time medical imaging systems such as reflectance confocal microscopes and optical coherence microscopes are being tested in multiple-patient and multiple-center clinical trials. The modulation transfer function (MTF) of these systems at any given time influences the image information content and can affect the interpretation of the images. MTF is difficult to measure in real-time scanning systems when imaging at the Nyquist limit. We describe a measurement technique similar to the electronic imaging resolution standards ISO-12233 (electronic cameras) that can be applied to scanned spot imaging systems with asynchronous pixel clocks. This technique requires the acquisition of a single image of a reflective stripe object. An asynchronous pixel clock induces subpixel jitter in the edge location. The jitter is removed using a Fourier method, and an oversampled edge response function is calculated using algorithms developed in MATLAB. This technique provides fast, simple to use, and repeatable full-width at half maximum lateral resolution and MTF measurements based on only one test image. We present the results for reflectance confocal microscopes operating at 0.9 numerical aperature. (c) 2007 Society of Photo-Optical Instrumentation Engineers.
引用
收藏
页数:6
相关论文
共 50 条
  • [41] New method to improve measurement accuracy of modulation transfer function
    Song, Zongxi
    Gao, Wei
    Zhou, Yan
    Liu, Feng
    5TH INTERNATIONAL SYMPOSIUM ON ADVANCED OPTICAL MANUFACTURING AND TESTING TECHNOLOGIES: OPTICAL TEST AND MEASUREMENT TECHNOLOGY AND EQUIPMENT, 2010, 7656
  • [42] Distance-weighted Modulation Transfer Function Measurement Method
    Zhang Zheng
    Feng Huajun
    Xu Zhihai
    Li Qi
    Chen Yueting
    FIFTH CONFERENCE ON FRONTIERS IN OPTICAL IMAGING TECHNOLOGY AND APPLICATIONS (FOI 2018), 2018, 10832
  • [43] Measurement method of the modulation transfer function of focal plane arrays
    Chen, B
    Qian, YS
    Chang, BK
    Cai, Y
    MULTISPECTRAL AND HYPERSPECTRAL IMAGE ACQUISITION AND PROCESSING, 2001, 4548 : 363 - 368
  • [44] Measurement and analysis of modulation transfer function of digital image sensors
    Fang, Yi-Chin
    Tzeng, Yih-Fong
    Wu, Kuo-Ying
    Tsay, Ho-Lin
    Lin, Po-Ming
    MICROSYSTEM TECHNOLOGIES-MICRO-AND NANOSYSTEMS-INFORMATION STORAGE AND PROCESSING SYSTEMS, 2022, 28 (01): : 137 - 142
  • [45] Modulation transfer function measurement of' IR optics at extreme temperatures
    Lindgren, M
    Ljungberg, P
    Nilsson, K
    Kihlén, RG
    INFRARED TECHNOLOGY AND APPLICATIONS XXV111, PTS 1 AND 2, 2003, 4820 : 60 - 68
  • [46] Modulation transfer function measurement on QWIP focal plane array
    Rafol, S. B.
    Cho, E.
    INFRARED IMAGING SYSTEMS: DESIGN, ANALYSIS, MODELING, AND TESTING XIX, 2008, 6941
  • [47] Measurement and analysis of modulation transfer function of digital image sensors
    Yi-chin Fang
    Yih-fong Tzeng
    Kuo-Ying Wu
    Ho-Lin Tsay
    Po-Ming Lin
    Microsystem Technologies, 2022, 28 : 137 - 142
  • [48] DEVICE FOR MEASUREMENT OF MODULATION TRANSFER-FUNCTION OF PHOTOGRAPHIC MATERIALS
    POSPISIL, J
    BUMBA, V
    CZECHOSLOVAK JOURNAL OF PHYSICS, 1973, B 23 (04) : 407 - 415
  • [49] Modulation transfer function and noise measurement using neural networks
    Delvit, JM
    Léger, D
    Roques, S
    Valorge, C
    2003 IEEE XIII WORKSHOP ON NEURAL NETWORKS FOR SIGNAL PROCESSING - NNSP'03, 2003, : 131 - 140
  • [50] ULTRAMICROSCOPY IN SCANNING MICROSCOPES
    BOUWHUIS, G
    DEKKERS, NH
    OPTIK, 1980, 56 (03): : 233 - 242