Amperometric determination of hydrogen peroxide residue in beverages using a Nafion modified palladium electrode

被引:12
|
作者
Hsu, Chuan-Liang [1 ]
Jang, Hung-Der [1 ]
Su, Ming-Sheng [1 ]
Chang, Ku-Shang [1 ]
Huang, Yung-Sheng [1 ]
机构
[1] Yuanpei Univ, Dept Food Sci, Hsinchu 300, Taiwan
关键词
hydrogen peroxide; aseptic packages; amperometric detection; Nafion; ascorbic acid;
D O I
10.1007/s00217-007-0598-8
中图分类号
TS2 [食品工业];
学科分类号
0832 ;
摘要
Hydrogen peroxide (H2O2) residues are not permitted in food due to health concerns. An amperometric sensor based on a Nafion modified palladium (Pd) electrode was developed for the rapid determination of the H2O2 residue in aseptically packaged beverages. A prepared Pd electrode shows a slowly increasing reaction upon repeating the procedure of adding the same concentration of H2O2 into the reaction cell unless pretreating the electrode with cyclic voltammetry to obtain reproducible surfaces, whereas a Nafion modified Pd electrode can reach stable reaction to H2O2 without any pretreatment. The interference level of ascorbic acid (the ration of electrochemical response to ascorbic acid to the electrochemical response of H2O2) was calculated as 0.0004 for the Nafion modified Pd electrode. Interference (99.5%)from ascorbic acid was diminished by Nafion membrane. The response of the sensor to H2O2 in distilled water is linear over the range of 0.25-15 mu M (r(2)= 0.9956). The sensor developed in this study can efficiently determine H2O2 residues in commercial beverages with good linear correlation between spiked H2O2 concentration and detected H2O2 concentration, for winter melon-flavored drink, r(2)= 0.9904 in the range of 0.25-15 mu M and for lemon-flavored tea, r(2)= 0.9927 in the range of 0.25-25 mu M.
引用
收藏
页码:809 / 815
页数:7
相关论文
共 50 条
  • [31] Amperometric Determination of Hydrogen Peroxide in Whitening Gels Using Boron-doped Diamond Electrode
    Gustavo C. Azevedo
    Roberto H. S. Castro
    Maria Auxiliadora C. Matos
    Renato Camargo Matos
    Analytical Sciences, 2018, 34 : 913 - 917
  • [32] Amperometric Determination of Nicotine using a Composite of Palladium hexacyanoferrate incorporated Graphene oxide modified Electrode
    Yang, Cheng Yu
    Chang, Min-Hui
    Chen, Shen-Ming
    Devadas, Balamurugan
    INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE, 2016, 11 (04): : 2650 - 2658
  • [33] An amperometric immunosensor based on Nafion-modified electrode for the determination of Schistosoma japonicum antibody
    Zhou, YM
    Wu, ZY
    Shen, GL
    Yu, RQ
    SENSORS AND ACTUATORS B-CHEMICAL, 2003, 89 (03) : 292 - 298
  • [34] Amperometric hydrogen peroxide biosensor based on a modified gold electrode with silver nanowires
    Song, Min-Jung
    Hwang, Sung Woo
    Whang, Dongmok
    JOURNAL OF APPLIED ELECTROCHEMISTRY, 2010, 40 (12) : 2099 - 2105
  • [35] A poly(o-aminophenol) modified electrode as an amperometric hydrogen peroxide biosensor
    Garcia, MAV
    Blanco, PT
    Ivaska, A
    ELECTROCHIMICA ACTA, 1998, 43 (23) : 3533 - 3539
  • [36] Amperometric hydrogen peroxide biosensor based on a modified gold electrode with silver nanowires
    Min-Jung Song
    Sung Woo Hwang
    Dongmok Whang
    Journal of Applied Electrochemistry, 2010, 40 : 2099 - 2105
  • [37] Amperometric detection of hydrogen peroxide at a Prussian Blue-modified FTO electrode
    Tseng, KS
    Chen, LC
    Ho, KC
    SENSORS AND ACTUATORS B-CHEMICAL, 2005, 108 (1-2) : 738 - 745
  • [38] Flow injection amperometric determination of hydrogen peroxide in household commercial products with a ruthenium oxide hexacyanoferrate modified electrode
    Roselyn C. Peña
    Juan C. M. Gamboa
    Thiago R. L. C. Paixão
    Mauro Bertotti
    Microchimica Acta, 2009, 166 : 277 - 281
  • [39] Flow injection amperometric determination of hydrogen peroxide in household commercial products with a ruthenium oxide hexacyanoferrate modified electrode
    Pena, Roselyn C.
    Gamboa, Juan C. M.
    Paixao, Thiago R. L. C.
    Bertotti, Mauro
    MICROCHIMICA ACTA, 2009, 166 (3-4) : 277 - 281
  • [40] Determination of hydrogen peroxide using a Prussian Blue modified macroporous gold electrode
    Jiao Yang
    Meng Lin
    MiSuk Cho
    Youngkwan Lee
    Microchimica Acta, 2015, 182 : 1089 - 1094