Immunologic Adverse Reaction Associated with Low-carbide Metal-on-metal Bearings in Total Hip Arthroplasty

被引:39
作者
Aroukatos, Panagiotis [2 ]
Repanti, Maria [2 ]
Repantis, Thomas [1 ]
Bravou, Vassiliki [2 ]
Korovessis, Panagiotis [1 ]
机构
[1] Gen Hosp Agios Andreas, Dept Pathol, Patras 26224, Achaia, Greece
[2] Gen Hosp Agios Andreas, Dept Orthopaed, Patras 26224, Achaia, Greece
关键词
DELAYED-TYPE HYPERSENSITIVITY; RED-BLOOD-CELLS; IMMUNE-RESPONSE; FOLLOW-UP; COBALT; OSTEOLYSIS; PROSTHESES; CHROMIUM; REPLACEMENTS; WEAR;
D O I
10.1007/s11999-009-1187-x
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background An increased incidence of periprosthetic osteolysis, resulting in loss of biologic fixation, has been reported in contemporary THAs with low-carbide metal-on-metal compared with metal-on-polyethylene couple bearings. Although a hypersensitivity reaction attributable to Co and Cr debris is reportedly a potential cause for failure of THAs with high-carbide bearings, there are no evidence-based data for this reaction in low-carbide metal-on-metal bearings, although such hypersensitivity might be related to osteolysis. Questions/purposes We investigated whether there were differences in immunologic hypersensitivity reactions in retrievals from revised THAs with ceramic-on-polyethylene versus metal-on-metal bearing couples. Patients and Methods We compared newly formed capsule and periprosthetic interface membranes from revision surgery for aseptic failure from 20 patients with low-carbide bearings and 13 patients with ceramic-on-polyethylene bearings. For control tissue, we obtained samples from the hip capsule during the primary THA implantation in 13 patients with low-carbide bearings and seven with ceramic-on-polyethylene bearings. We examined the tissues with conventional histologic and immunohistochemical methods. Results Compared with tissue from the control subjects and patients with ceramic-on-polyethylene bearings, the tissues from patients with low-carbide metal-on-metal bearings were associated with (1) extensive necrosis and fibrin exudation in the newly formed hip capsule and (2) diffuse and perivascular lymphocytic infiltration of a higher degree than in the hips with ceramic-on-polyethylene bearings in conventional histologic examination, and (3) more T than B cells. Conclusions The conventional histologic and immunohistochemical findings in tissues retrieved from failed THAs with low-carbide metal-on-metal bearings are consistent with a link between hypersensitivity and osteolysis with low-carbide bearing couples.
引用
收藏
页码:2135 / 2142
页数:8
相关论文
共 38 条
[1]  
Boss JH, 1996, J LONG-TERM EFF MED, V6, P73
[2]   Serum cobalt levels after metal-on-metal total hip arthroplasty [J].
Brodner, W ;
Bitzan, P ;
Meisinger, V ;
Kaider, A ;
Gottsauner-Wolf, F ;
Kotz, R .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (11) :2168-2173
[3]   Metal-on-metal hip replacements: Wear performance and cellular response to wear particles [J].
Campbell, P ;
McKellop, H ;
Alim, R ;
Mirra, J ;
Nutt, S ;
Dorr, L ;
Amstutz, HC .
COBALT-BASE ALLOYS FOR BIOMEDICAL APPLICATIONS, 1999, 1365 :193-209
[4]   An unusual lymphocytic perivascular infiltration in tissues around contemporary metal-on-metal joint replacements [J].
Davies, AP ;
Willert, HG ;
Campbell, PA ;
Learmonth, ID ;
Case, CP .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (01) :18-27
[5]   Metal-on-metal bearings in cementless primary total hip arthroplasty [J].
Delaunay, CP .
JOURNAL OF ARTHROPLASTY, 2004, 19 (08) :35-40
[6]  
Doorn PF, 1996, CLIN ORTHOP RELAT R, pS187
[7]   Fixation and osteolysis with Metasul metal-on-metal articulation [J].
Dorr, LD ;
Wan, ZN ;
Sirianni, LE ;
Boutary, M ;
Chandran, S .
JOURNAL OF ARTHROPLASTY, 2004, 19 (08) :951-955
[8]  
Firkins PJ, 2001, BIO-MED MATER ENG, V11, P143
[9]   Metal sensitivity in patients with orthopaedic implants [J].
Hallab, N ;
Merritt, K ;
Jacobs, JJ .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2001, 83A (03) :428-436
[10]  
Hallab NJ., 2008, J Orthop Surg Res, V3, P6, DOI [10.1186/1749-799X-3-6, DOI 10.1186/1749-799X-3-6]