Blood Flow Measurement using Digital Subtraction Angiography for Assessing Hemodialysis Access Function

被引:4
作者
Koirala, Nischal [1 ,2 ]
Setser, Randolph M. [3 ]
Bullen, Jennifer [4 ]
McLennan, Gordon [2 ]
机构
[1] Cleveland State Univ, Dept Chem & Biomed Engn, Cleveland, OH 44115 USA
[2] Cleveland Clin, Dept Biomed Engn, Cleveland, OH 44106 USA
[3] Siemens Med Solut USA Inc, Cleveland, OH USA
[4] Cleveland Clin, Dept Quantitat Hlth Sci, Cleveland, OH 44106 USA
来源
MEDICAL IMAGING 2017: BIOMEDICAL APPLICATIONS IN MOLECULAR, STRUCTURAL, AND FUNCTIONAL IMAGING | 2017年 / 10137卷
关键词
Blood flow; bolus tracking algorithm; cross correlation; digital subtraction angiography; flow phantom; hemodialysis access; indicator-dilution principle; thermodilution; DOPPLER ULTRASOUND; VELOCITY; QUANTIFICATION; CONTRAST; VALIDATION; ARTERIES;
D O I
10.1117/12.2253929
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
Blood flow rate is a critical parameter for diagnosing dialysis access function during fistulography where a flow rate of 600 ml/min in arteriovenous graft or 400-500 ml/min in arteriovenous fistula is considered the clinical threshold for fully functioning access. In this study, a flow rate computational model for calculating intra-access flow to evaluate dialysis access patency was developed and validated in an in vitro set up using digital subtraction angiography. Flow rates were computed by tracking the bolus through two regions of interest using cross correlation (XCOR) and mean arrival time (MAT) algorithms, and correlated versus an in-line transonic flow meter measurement. The mean difference (mean. standard deviation) between XCOR and in-line flow measurements for in vitro setup at 3, 6, 7.5 and 10 frames/s was 118 +/- 63; 37 +/- 59; 31 +/- 31; and 46.+/- 57 ml/min respectively while for MAT method it was 86 +/- 56; 57 +/- 72; 35 +/- 85; and 19 +/- 129 ml/min respectively. The result of this investigation will be helpful for selecting candidate algorithms while blood flow computational tool is developed for clinical application.
引用
收藏
页数:15
相关论文
共 50 条
[41]   Normal anatomy and variations in the confluence of sinuses using digital subtraction angiography [J].
Cheng, Yaqin ;
Li, William A. ;
Fan, Xingjuan ;
Li, Xiaohua ;
Chen, Jian ;
Wu, Yan ;
Meng, Ran ;
Ji, Xunming .
NEUROLOGICAL RESEARCH, 2017, 39 (06) :509-515
[42]   Intra-Aneurysmal Flow Patterns: Illustrative Comparison among Digital Subtraction Angiography, Optical Flow, and Computational Fluid Dynamics [J].
Brina, O. ;
Ouared, R. ;
Bonnefous, O. ;
van Nijnatten, F. ;
Bouillot, P. ;
Bijlenga, P. ;
Schaller, K. ;
Lovblad, K. -O ;
Gruenhagen, T. ;
Ruijters, D. ;
Pereira, V. Mendes .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2014, 35 (12) :2348-2353
[43]   In vivo imaging of rat coronary arteries using bi-plane digital subtraction angiography [J].
Badea, Cristian T. ;
Hedlund, Laurence W. ;
Qi, Yi ;
Berridge, Brian ;
Johnson, G. Allan .
JOURNAL OF PHARMACOLOGICAL AND TOXICOLOGICAL METHODS, 2011, 64 (02) :151-157
[44]   Peri-Therapeutic Quantitative Flow Analysis of Endovascular Revascularization for Ischemic Stroke Patients on Digital Subtraction Angiography [J].
Shih-Neng Yang ;
Yu-Yi Lin ;
Tsung-Jung Ho ;
Shi-Yann Cheng ;
Yu-Rou Chiou ;
Chun-Han Liao ;
Geoffrey G. Zhang ;
Tzung-Chi Huang .
Journal of Medical and Biological Engineering, 2018, 38 :387-395
[45]   Parametric Digital Subtraction Angiography Imaging for the Objective Grading of Collateral Flow in Acute Middle Cerebral Artery Occlusion [J].
Wen, Wan-ling ;
Fang, Yi-bin ;
Yang, Peng-fei ;
Zhang, Yong-wei ;
Wu, Yi-na ;
Shen, Hui ;
Ge, Jia-jia ;
Xu, Yi ;
Hong, Bo ;
Huang, Qing-hai ;
Liu, Jian-min .
WORLD NEUROSURGERY, 2016, 88 :119-125
[46]   Digital subtraction and conventional blood flow studies in porcine experimental aneurysms [J].
Asakura, F ;
Tenjin, H ;
Matsumoto, K ;
Ueda, S ;
Urano, M .
INTERVENTIONAL NEURORADIOLOGY, 1999, 5 (03) :207-212
[47]   Carbon dioxide digital subtraction angiography using a new gas management system [J].
SchmitzRode, T ;
Alzen, G ;
Gunther, RW .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1997, 167 (01) :71-78
[48]   Practical utility of thermodilution versus doppler ultrasound to measure hemodialysis blood access flow [J].
Fontsere, Nestor ;
Mestres, Gaspar ;
Barrufet, Marta ;
Burrel, Marta ;
Vera, Manel ;
Arias, Marta ;
Masso, Elisabeth ;
Cases, Aleix ;
Maduell, Francisco ;
Campistol, Josep M. .
NEFROLOGIA, 2013, 33 (03) :325-332
[49]   Hemodialysis access flow measurement: Comparison of ultrasound dilution and ultrafiltration method on NIKKISO DBB-EXA™ dialysis machine [J].
Kosmadakis, Georgios ;
Enache, Ioana ;
Gueret, Claudine ;
Haskour, Abraham ;
Necoara, Aura ;
Deville, Clemence ;
Baudenon, Julien ;
Rance, Nicole .
JOURNAL OF VASCULAR ACCESS, 2024, 25 (05) :1501-1507
[50]   Evaluation of intracranial aneurysms with high-resolution MR angiography using single-artery highlighting technique: correlation with digital subtraction angiography [J].
Li, Hui ;
Yan, Lei ;
Li, Ming-Hua ;
Li, Yong-Dong ;
Tan, Hua-Qiao ;
Gu, Bin-Xian ;
Wang, Wu .
RADIOLOGIA MEDICA, 2013, 118 (08) :1379-1387