Ex Vivo Evaluation of Mouse Brain Elasticity Using High-Frequency Ultrasound Elastography

被引:19
作者
Lay, Fang-Yi [1 ]
Chen, Pei-Yu [1 ]
Cheng, Hsiang-Fan [2 ]
Kuo, Yu-Min [2 ,3 ]
Huang, Chih-Chung [1 ,4 ]
机构
[1] Natl Cheng Kung Univ, Coll Engn, Dept Biomed Engn, Tainan 701, Taiwan
[2] Natl Cheng Kung Univ, Coll Med, Inst Basic Med Sci, Tainan, Taiwan
[3] Natl Cheng Kung Univ, Coll Med, Dept Cell Biol & Anat, Tainan, Taiwan
[4] Natl Cheng Kung Univ, Med Device Innovat Ctr, Tainan 701, Taiwan
关键词
Brain elasticity; high-frequency ultrasound; elastography; shear wave imaging; small-animal imaging; MAGNETIC-RESONANCE ELASTOGRAPHY; IN-VIVO; TRANSIENT ELASTOGRAPHY; VISCOELASTIC PROPERTIES; TISSUE ELASTICITY; MR ELASTOGRAPHY; STIFFNESS; PARAMETERS; INJURY;
D O I
10.1109/TBME.2019.2905551
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: Most neurodegenerative diseases are highly linked with aging. The mechanical properties of the brain should be determined for predicting and diagnosing age-related brain diseases. A preclinical animal study is crucial for neurological disease research. However, estimation of the elasticity properties of different regions of mouse brains remains difficult because of the size of the brain. In this paper, high-frequency ultrasound elastography (HFUSE) based on shear wave imaging was proposed for mapping the stiffness of the mouse brain at different ages ex vivo. Methods: For HFUSE, a 40-MHz ultrasound array transducer with an ultrafast ultrasound imaging system was used in this paper. The accuracy and resolution during HFUSE were determined through a mechanical testing system and by conducting phantom experiments. Results: In the experiments, the error in the elastic modulus measurement was approximately 10% on average, and the axial resolution was 248 mu m. Animal testing was conducted using mice that were 4 (young aged) and 11 (middle aged) months old. The elasticity distributions of the cortex and hippocampus in the mouse brains were obtained through HFUSE. Conclusion: The average shear moduli of the cortex and hippocampus were 3.84 and 2.33 kPa for the 4-month-old mice and 3.77 and 1.94 kPa for the 11-month-old mice, respectively. No statistical difference was observed in the cortex stiffness of mice of different ages. However, the hippocampus significantly softened with aging.
引用
收藏
页码:3426 / 3435
页数:10
相关论文
共 50 条
[1]   Ex Vivo Evaluation of Mechanical Anisotropic Tissues with High-Frequency Ultrasound Shear Wave Elastography [J].
Lee, Seungyeop ;
Eun, Lucy Youngmin ;
Hwang, Jae Youn ;
Eun, Yongsoon .
SENSORS, 2022, 22 (03)
[2]   Quantitative Evaluation of Neonatal Brain Elasticity Using Shear Wave Elastography [J].
Garces Inigo, Enrique ;
Llorens Salvador, Roberto ;
Escrig, Raquel ;
Hervas, David ;
Vento, Maximo ;
Marti-Bonmati, Luis .
JOURNAL OF ULTRASOUND IN MEDICINE, 2021, 40 (04) :795-804
[3]   Characterization of the extensor digitorum communis tendon using high-frequency ultrasound shear wave elastography [J].
Hsiao, Yan-Yi ;
Yang, Tai-Hua ;
Chen, Pei-Yu ;
Hsu, Hsiu-Yun ;
Kuo, Li-Chieh ;
Su, Fong-Chin ;
Huang, Chih-Chung .
MEDICAL PHYSICS, 2020, 47 (04) :1609-1618
[4]   In vivo and ex vivo elastic properties of brain tissues measured with ultrasound elastography [J].
Liu, Yan-Lin ;
Liu, Dong ;
Xu, Liang ;
Su, Chen ;
Li, Guo-Yang ;
Qian, Lin-Xue ;
Cao, Yanping .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2018, 83 :120-125
[5]   Regional Biomechanical Weakening in Keratoconus Corneas Detected by In Vivo High-Frequency Ultrasound Elastography [J].
Kwok, Sunny ;
Pan, Xueliang ;
Pan, Manqi ;
Chen, Zihao ;
Ammon, Madison ;
Hendershot, Andrew ;
Liu, Jun .
TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2025, 14 (03)
[6]   Shear-Wave Elasticity Imaging of a Liver Fibrosis Mouse Model Using High-Frequency Ultrasound [J].
Yeh, Chia-Lun ;
Chen, Bo-Rong ;
Tseng, Ling-Yi ;
Jao, Ping ;
Su, Tung-Hung ;
Li, Pai-Chi .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2015, 62 (07) :1295-1307
[7]   Evaluation of Hand Tendon Elastic Properties During Rehabilitation Through High-Frequency Ultrasound Shear Elastography [J].
Chen, Pei-Yu ;
Yang, Tai-Hua ;
Kuo, Li-Chieh ;
Hsu, Hsiu-Yun ;
Su, Fong-Chin ;
Huang, Chih-Chung .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2021, 68 (08) :2716-2726
[8]   Characterization of Hand Tendons Through High-Frequency Ultrasound Elastography [J].
Chen, Pei-Yu ;
Yang, Tai-Hua ;
Kuo, Li-Chieh ;
Shih, Cho-Chiang ;
Huang, Chih-Chung .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2020, 67 (01) :37-48
[9]   Intraocular Pressure-dependent Corneal Elasticity Measurement Using High-frequency Ultrasound [J].
Osapoetra, Laurentius O. ;
Watson, Dan M. ;
McAleavey, Stephen A. .
ULTRASONIC IMAGING, 2019, 41 (05) :251-270
[10]   Three-Dimensional Reconstruction and Visualization of Human Enamel Ex Vivo Using High-Frequency Ultrasound [J].
Du, Juan ;
Mao, Xue-Li ;
Ye, Peng-Fei ;
Huang, Qing-Hua .
JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2017, 37 (01) :112-122