The Accuracy of Platelet Counting in Thrombocytopenic Blood Samples Distributed by the UK National External Quality Assessment Scheme for General Haematology

被引:26
作者
De la Salle, Barbara J. [1 ]
McTaggart, Paul N. [1 ]
Briggs, Carol [2 ]
Harrison, Paul [3 ]
Dore, Caroline J. [4 ]
Longair, Ian [2 ]
Machin, Samuel J. [2 ]
Hyde, Keith [1 ]
机构
[1] UK Natl External Qual Assessment Scheme Gen Haema, Watford, England
[2] Univ Coll London Hosp, Dept Haematol, London, England
[3] Churchill Hosp, Oxford Haemophilia & Thrombosis Ctr, Oxford OX3 7LJ, England
[4] MRC, Clin Trials Unit, London W1N 4AL, England
关键词
Proficiency testing; Thrombocytopenia; Platelet count; Laboratory hematology; Transfusion; ANALYZERS;
D O I
10.1309/AJCP86JMBFUCFCXA
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
A knowledge of the limitations of automated platelet counting is essential for the effective care of thrombocytopenic patients and management of platelet stocks for transfusion. For this study, 29 external quality assessment specimen pools with platelet counts between 5 and 64 x 10(9)/L were distributed to more than 1,100 users of 23 different hematology analyzer models. The same specimen pools were analyzed by the international reference method (IRM) for platelet counting at 3 reference centers. The IRM values were on average lower than the all-methods median values returned by the automated analyzers. The majority (similar to 67%) of the automated analyzer results overestimated the platelet count compared with the IRM, with significant differences in 16.5% of cases. Performance differed between analyzer models. The observed differences may depend in part on the nature of the survey material and analyzer technology, but the findings have implications for the interpretation of platelet counts at levels of clinical decision making.
引用
收藏
页码:65 / 74
页数:10
相关论文
共 21 条
[1]  
[Anonymous], INT J LAB HEMATOL
[2]  
[Anonymous], 1996, LAB HAEMATOL
[3]  
[Anonymous], 88 WHO LAB INT COUNC
[4]  
[Anonymous], 2006, DACIE LEWIS PRACTICA
[5]   A PARALLEL EVALUATION OF 4 AUTOMATED HEMATOLOGY ANALYZERS [J].
BENTLEY, SA ;
JOHNSON, A ;
BISHOP, CA .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1993, 100 (06) :626-632
[6]   Discrepancy between impedance and immunofluorescence platelet counting has implications for clinical decision making in patients with idiopathic thrombocytopenia purpura [J].
Bowles, Kristian M. ;
Bloxham, David M. ;
Perry, David J. ;
Baglin, Trevor P. .
BRITISH JOURNAL OF HAEMATOLOGY, 2006, 134 (03) :320-322
[7]  
BRECHER G, 1953, AM J CLIN PATHOL, V23, P15
[8]   Relationships between platelet counts, platelet volumes and reticulated platelets in patients with ITP: evidence for significant platelet count inaccuracies with conventional instrument methods [J].
Diquattro, M. ;
Gagliano, F. ;
Calabro, G. M. ;
Tommasi, M. ;
Scott, C. S. ;
Mancuso, G. ;
Palma, B. ;
Menozzi, I. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2009, 31 (02) :199-206
[9]  
Harrison P, 2000, BRIT J HAEMATOL, V108, P228
[10]  
Harrison P, 2001, AM J CLIN PATHOL, V115, P448