First long term in vivo study on subdurally implanted Micro-ECoG electrodes, manufactured with a novel laser technology

被引:79
作者
Henle, C. [1 ]
Raab, M. [2 ]
Cordeiro, J. G. [3 ]
Doostkam, S. [4 ]
Schulze-Bonhage, A. [3 ]
Stieglitz, T. [1 ,5 ]
Rickert, J. [2 ,5 ]
机构
[1] Univ Freiburg, Lab Biomed Microtechnol, Dept Microsyst Engn IMTEK, Freiburg I Breisgau, Germany
[2] Univ Freiburg, Inst Biol 1, Freiburg I Breisgau, Germany
[3] Univ Hosp Freiburg, Epilepsy Ctr Freiburg, Freiburg I Breisgau, Germany
[4] Univ Hosp Freiburg, Dept Neuropathol, Freiburg I Breisgau, Germany
[5] Univ Freiburg, Bernstein Ctr Freiburg, Freiburg I Breisgau, Germany
关键词
Neuroprostheses; Subdural electrodes; Laser technology; Brain-computer interface; Chronic implantation; Impedance spectroscopy; ELECTROCORTICOGRAPHIC SIGNALS; IMPEDANCE; TISSUE; INTERFACE; ARRAYS;
D O I
10.1007/s10544-010-9471-9
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A novel computer aided manufacturing (CAM) method for electrocorticography (ECoG) microelectrodes was developed to be able to manufacture small, high density microelectrode arrays based on laser-structuring medical grade silicone rubber and high purity platinum. With this manufacturing process, we plan to target clinical applications, such as presurgical epilepsy monitoring, functional imaging during cerebral tumor resections and brain-computer interface control in paralysed patients, in the near future. This paper describes the manufacturing, implantation and long-term behaviour of such an electrode array. In detail, we implanted 8-channel electrode arrays subdurally over rat cerebral cortex over a period of up to 25 weeks. Our primary objective was to ascertain the electrode's stability over time, and to analyse the host response in vivo. For this purpose, impedance measurements were carried out at regular intervals over the first 18 weeks of the implantation period. The impedances changed between day 4 and day 7 after implantation, and then remained stable until the end of the implantation period, in accordance with typical behaviour of chronically implanted microelectrodes. A post-mortem histological examination was made to assess the tissue reaction due to the implantation. A mild, chronically granulated inflammation was found in the area of the implant, which was essentially restricted to the leptomeninges. Overall, these findings suggest that the concept of the presented ECoG-electrodes is promising for use in long-term implantations.
引用
收藏
页码:59 / 68
页数:10
相关论文
共 35 条
[1]   A study of intra-cochlear electrodes and tissue interface by electrochemical impedance methods in vivo [J].
Duan, YY ;
Clark, GM ;
Cowan, RSC .
BIOMATERIALS, 2004, 25 (17) :3813-3828
[2]   Impedance characterization and modeling of electrodes for biomedical applications [J].
Franks, W ;
Schenker, I ;
Schmutz, P ;
Hierlemann, A .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2005, 52 (07) :1295-1302
[3]   Spatial spectral analysis of human electrocorticograms including the alpha and gamma bands [J].
Freeman, WJ ;
Rogers, LJ ;
Holmes, MD ;
Silbergeld, DL .
JOURNAL OF NEUROSCIENCE METHODS, 2000, 95 (02) :111-121
[4]   Preservation of function in vestibular schwannoma surgery [J].
Gharabaghi, Alireza ;
Samii, Amir ;
Koerbel, Andrei ;
Rosahl, Steffen K. ;
Tatagiba, Marcos ;
Samii, Madjid .
NEUROSURGERY, 2007, 60 (02) :124-127
[5]   Cytotoxicity of implantable microelectrode arrays produced by laser micromachining [J].
Green, Rylie A. ;
Ordonez, Juan S. ;
Schuettler, Martin ;
Poole-Warren, Laura A. ;
Lovell, Nigel H. ;
Suaning, Gregg J. .
BIOMATERIALS, 2010, 31 (05) :886-893
[6]   ELECTRICAL-PROPERTIES OF IMPLANT ENCAPSULATION TISSUE [J].
GRILL, WM ;
MORTIMER, JT .
ANNALS OF BIOMEDICAL ENGINEERING, 1994, 22 (01) :23-33
[7]   Scaling Limitations of Laser-Fabricated Nerve Electrode Arrays [J].
Henle, Christian ;
Schuettler, Martin ;
Ordonez, Juan S. ;
Stieglitz, Thomas .
2008 30TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-8, 2008, :4208-4211
[8]   A MEMS fabricated flexible electrode array for recording surface field potentials [J].
Hollenberg, Brian A. ;
Richards, Cecilia D. ;
Richards, Robert ;
Bahr, David F. ;
Rector, David M. .
JOURNAL OF NEUROSCIENCE METHODS, 2006, 153 (01) :147-153
[9]  
JESCHKE C, 2008, P IFMBE, P2447
[10]   A brain-computer interface using electrocorticographic signals in humans [J].
Leuthardt, Eric C. ;
Schalk, Gerwin ;
Wolpaw, Jonathan R. ;
Ojemann, Jeffrey G. ;
Moran, Daniel W. .
JOURNAL OF NEURAL ENGINEERING, 2004, 1 (02) :63-71