A compact time-resolved system for near infrared spectroscopy based on wavelength space multiplexing

被引:35
|
作者
Re, Rebecca [1 ]
Contini, Davide [1 ]
Caffini, Matteo [1 ]
Cubeddu, Rinaldo [1 ,2 ,3 ]
Spinelli, Lorenzo [2 ]
Torricelli, Alessandro [1 ,3 ]
机构
[1] Politecn Milan, Dipartimento Fis, I-20133 Milan, Italy
[2] Ist Foton & Nanotecnol, IFN CNR, I-20133 Milan, Italy
[3] Politecn Milan, Res Unit IIT, I-20133 Milan, Italy
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2010年 / 81卷 / 11期
关键词
OPTICAL-PROPERTIES; PHOTON MIGRATION; BRAIN; LIGHT; MODEL; REFLECTANCE;
D O I
10.1063/1.3495957
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We designed and developed a compact dual-wavelength and dual-channel time-resolved system for near-infrared spectroscopy studies of muscle and brain. The system employs pulsed diode lasers as sources, compact photomultipliers, and time-correlated single photon counting boards for detection. To exploit the full temporal and dynamic range of the acquisition technique, we implemented an approach based on wavelength space multiplexing: laser pulses at the two wavelengths are alternatively injected into the two channels by means of an optical 2 x 2 switch. In each detection line (i.e., in each temporal window), the distribution of photon time-of-flights at one wavelength is acquired. The proposed approach increases the signal-to-noise ratio and avoids wavelength cross-talk with respect to the typical approach based on time multiplexing. The instrument was characterized on tissue phantoms to assess its properties in terms of linearity, stability, noise, and reproducibility. Finally, it was successfully tested in preliminary in vivo measurements on muscle during standard cuff occlusion and on the brain during a motor cortex response due to hand movements. (C) 2010 American Institute of Physics. [doi:10.1063/1.3495957]
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Compact dual-wavelength system for time-resolved diffuse optical spectroscopy
    Renna, Marco
    Buttafava, Mauro
    Zappa, Franco
    Tosi, Alberto
    Martinenghi, Edoardo
    Zanoletti, Marta
    Dalla Mora, Alberto
    Pifferi, Antonio
    Torricelli, Alessandro
    Contini, Davide
    2017 13TH CONFERENCE ON PH.D. RESEARCH IN MICROELECTRONICS AND ELECTRONICS (PRIME), 2017, : 293 - 296
  • [2] Functional brain imaging by multi-wavelength time-resolved near infrared spectroscopy
    Torricelli, A.
    Contini, D.
    Pifferi, A.
    Spinelli, L.
    Cubeddu, R.
    OPTO-ELECTRONICS REVIEW, 2008, 16 (02) : 131 - 135
  • [3] A compact time-resolved system for NIR spectroscopy
    Re, Rebecca
    Contini, Davide
    Caffini, Matteo
    Spinelli, Lorenzo
    Cubeddu, Rinaldo
    Torricelli, Alessandro
    CLINICAL AND BIOMEDICAL SPECTROSCOPY, 2009, 7368
  • [4] Near-infrared, time-resolved optical tomography with a novel compact system
    Tagliabue, S.
    Di Sieno, L.
    Farina, A.
    Martinenghi, E.
    Arridge, S.
    Martelli, F.
    Durduran, T.
    Pifferi, A.
    Torricelli, A.
    Dalla Mora, A.
    JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 2017, 37 : 356 - 357
  • [5] Multi-wavelength, dual-detection channel system for time-resolved near-infrared spectroscopy
    Renna, Marco
    Buttafava, Mauro
    Behera, Anurag
    Zanoletti, Marta
    Di Sieno, Laura
    Dalla Mora, Alberto
    Contini, Davide
    Tosi, Alberto
    OPTICAL TOMOGRAPHY AND SPECTROSCOPY OF TISSUE XIII, 2019, 10874
  • [6] Multi-channel time-resolved system for functional near infrared spectroscopy
    Contini, Davide
    Torricelli, Alessandro
    Pifferi, Antonio
    Spinelli, Lorenzo
    Paglia, Floriano
    Cubeddu, Rinaldo
    OPTICS EXPRESS, 2006, 14 (12) : 5418 - 5432
  • [7] Multi-channel time-resolved functional near infrared spectroscopy system
    Contini, D.
    Re, R.
    Turola, M.
    Spinelli, L.
    Romano, G.
    Cubeddu, R.
    Torricelli, A.
    OPTICAL TOMOGRAPHY AND SPECTROSCOPY OF TISSUE X, 2013, 8578
  • [8] A time-resolved near-infrared spectroscopy based on CMOS image sensor
    Lioe, De Xing
    Hakamata, Masashi
    Yasutomi, Keita
    Kagawa, Keiichiro
    Niwayama, Masatsugu
    Fukushi, Yasuko
    Yamamoto, Seiji
    Kawahito, Shoji
    OPTICAL BIOPSY XVIII: TOWARD REAL-TIME SPECTROSCOPIC IMAGING AND DIAGNOSIS, 2020, 11234
  • [9] Multi-channel medical device for time domain functional near infrared spectroscopy based on wavelength space multiplexing
    Re, Rebecca
    Contini, Davide
    Turola, Massimo
    Spinelli, Lorenzo
    Zucchelli, Lucia
    Caffini, Matteo
    Cubeddu, Rinaldo
    Torricelli, Alessandro
    BIOMEDICAL OPTICS EXPRESS, 2013, 4 (10): : 2231 - 2246
  • [10] Time-Resolved Near Infrared Spectroscopy in Ischemic Stroke Patients
    Zanoletti, Marta
    Giacalone, Giacomo
    Contini, Davide
    Re, Rebecca
    Spinelli, Lorenzo
    Germinario, Bruno
    Roveri, Luisa
    Torricelli, Alessandro
    DIFFUSE OPTICAL SPECTROSCOPY AND IMAGING VII, 2019, 11074