Simultaneous Quantification of 11 Antiseizure Medications and Metabolites in Serum for Therapeutic Drug Monitoring Using High-Performance Liquid Chromatography with Ultraviolet Detection: A Short Communication

被引:1
|
作者
Lipecki, Yannick [1 ,2 ]
Haen, Ekkehard [1 ,2 ]
机构
[1] Univ Regensburg, Dept Psychiat & Psychotherapy, Clin Pharmacol, Univ Str 31, D-93053 Regensburg, Germany
[2] Inst AGATE gGmbH, Clin Pharmacol, Pentling, Germany
关键词
HPLC-UV; therapeutic drug monitoring; antiseizure medication; AED; antiepileptic drugs; ANTIEPILEPTIC DRUGS;
D O I
10.1097/FTD.0000000000000908
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Therapeutic drug monitoring (TDM) of antiseizure medications (ASMs) is widely used to guide therapy, avoid toxicity, and assess patient compliance. Commercial immunologic quantification methods are common practice; however, as they are only applicable to one specific drug and prone to cross-reacting metabolites, their practical applicability is limited. In this article, the authors proposed a high-performance liquid chromatography method using ultraviolet detection (HPLC-UV) for simultaneous quantification of 11 ASMs and active metabolites (carbamazepine, felbamate, lacosamide, lamotrigine, levetiracetam, phenobarbital, phenytoin, primidone, zonisamide, carbamazepine-10,11-epoxide, and licarbazepine) in serum. Methods: Chromatographic separation was performed on a Phenomenex Luna PFP(2) (3-mu m particle size; 150 x 4.6 mm i.d.) analytical column. The mobile phase comprised phosphate buffer (20 mM; pH 3), acetonitrile (ACN), and methanol using gradient elution. Analyses were conducted at 35 degrees C and a 1.3-mL/min flow rate. The detection wavelength for all analytes was 210 nm. The samples were prepared by protein precipitation using ACN. Results: The HPLC-UV method was validated according to the FDA guidelines and applied to measure patient samples in TDM. Calibration curves showed excellent linearity (r(2) > 0.99) and covered the entire reference range for each analyte. Intraday and interday imprecisions and inaccuracies were <10% for all samples. Extensive stability testing showed no significant degradation (<15%), and interference measurements additionally ensured clinical applicability. Furthermore, the sensitivity was comparable with that of previously published HPLC methods using mass spectrometry. Conclusions: The authors developed an HPLC-UV method for the simultaneous quantification of 11 ASMs in the human serum and demonstrated its practical applicability in TDM. The method requires only standard laboratory equipment and simple sample preparation, making TDM available in less specialized laboratories.
引用
收藏
页码:345 / 350
页数:6
相关论文
共 50 条
  • [31] SIMULTANEOUS DETERMINATION OF DIHYDROCODEINE AND ITS METABOLITES IN DOG PLASMA BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH ELECTROCHEMICAL AND ULTRAVIOLET DETECTION
    OHNO, M
    SHIONO, Y
    KONISHI, M
    JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1994, 654 (02): : 213 - 219
  • [32] Therapeutic drug monitoring of three antiepileptic drugs using High-Performance Liquid Chromatography with UV-Visible detection
    Gaies, E.
    Jebabli, N.
    Ben Sassi, M. B. S.
    Charfi, R.
    El Jebari, H. E. J.
    Issam, I. S.
    Riadh, R. D.
    Sameh, S. T.
    FUNDAMENTAL & CLINICAL PHARMACOLOGY, 2021, 35 : 148 - 149
  • [33] A simple method for quantification of voriconazole by high-performance liquid chromatography with ultraviolet-detection
    Yamada, T.
    Suzuki, K.
    Urashima, Y.
    Kato, R.
    Ijiri, Y.
    Nishihara, M.
    Tanaka, K.
    Takiuchi, H.
    Tamai, H.
    THERAPEUTIC DRUG MONITORING, 2011, 33 (04) : 519 - 519
  • [34] Clozapine and norclozapine quantification in human plasma by high-performance liquid chromatography with ultraviolet detection
    Illamola, S. M.
    Perez, A. A.
    Morillas, M. G.
    Fresquet, X. S.
    Barreda, F. C.
    THERAPEUTIC DRUG MONITORING, 2011, 33 (04) : 525 - 525
  • [35] Rapid quantification of indinavir in human plasma by high-performance liquid chromatography with ultraviolet detection
    Foisy, ML
    Sommadossi, JP
    JOURNAL OF CHROMATOGRAPHY B, 1999, 721 (02): : 239 - 247
  • [36] RAPID SIMULTANEOUS DETERMINATION OF NITRATE AND NITRITE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY USING ULTRAVIOLET DETECTION
    GERRITSE, RG
    JOURNAL OF CHROMATOGRAPHY, 1979, 171 (APR): : 527 - 529
  • [37] Therapeutic Drug Monitoring of Voriconazole: Comparison of Bioassay with High-Performance Liquid Chromatography
    Badiee, Parisa
    Hashemizadeh, Zahra
    Montaseri, Hashem
    JUNDISHAPUR JOURNAL OF MICROBIOLOGY, 2017, 10 (04)
  • [38] Simultaneous determination of verapamil and norverapamil in biological samples by high-performance liquid chromatography using ultraviolet detection
    Garcia, MA
    Aramayona, JJ
    Bregante, MA
    Fraile, LJ
    Solans, C
    JOURNAL OF CHROMATOGRAPHY B, 1997, 693 (02): : 377 - 382
  • [39] Simultaneous Determination of 12 β-Lactam Antibiotics in Human Plasma by High-Performance Liquid Chromatography with UV Detection: Application to Therapeutic Drug Monitoring
    Verdier, Marie-Clemence
    Tribut, Olivier
    Tattevin, Pierre
    Le Tulzo, Yves
    Michelet, Christian
    Bentue-Ferrer, Daniele
    ANTIMICROBIAL AGENTS AND CHEMOTHERAPY, 2011, 55 (10) : 4873 - 4879
  • [40] Simultaneous quantification of amoxycillin and metronidazole in plasma using high-performance liquid chromatography with photodiode array detection
    Menelaou, A
    Somogyi, AA
    Barclay, ML
    Bochner, F
    JOURNAL OF CHROMATOGRAPHY B, 1999, 731 (02): : 261 - 266