Simultaneous Quantification of 11 Antiseizure Medications and Metabolites in Serum for Therapeutic Drug Monitoring Using High-Performance Liquid Chromatography with Ultraviolet Detection: A Short Communication

被引:1
|
作者
Lipecki, Yannick [1 ,2 ]
Haen, Ekkehard [1 ,2 ]
机构
[1] Univ Regensburg, Dept Psychiat & Psychotherapy, Clin Pharmacol, Univ Str 31, D-93053 Regensburg, Germany
[2] Inst AGATE gGmbH, Clin Pharmacol, Pentling, Germany
关键词
HPLC-UV; therapeutic drug monitoring; antiseizure medication; AED; antiepileptic drugs; ANTIEPILEPTIC DRUGS;
D O I
10.1097/FTD.0000000000000908
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Therapeutic drug monitoring (TDM) of antiseizure medications (ASMs) is widely used to guide therapy, avoid toxicity, and assess patient compliance. Commercial immunologic quantification methods are common practice; however, as they are only applicable to one specific drug and prone to cross-reacting metabolites, their practical applicability is limited. In this article, the authors proposed a high-performance liquid chromatography method using ultraviolet detection (HPLC-UV) for simultaneous quantification of 11 ASMs and active metabolites (carbamazepine, felbamate, lacosamide, lamotrigine, levetiracetam, phenobarbital, phenytoin, primidone, zonisamide, carbamazepine-10,11-epoxide, and licarbazepine) in serum. Methods: Chromatographic separation was performed on a Phenomenex Luna PFP(2) (3-mu m particle size; 150 x 4.6 mm i.d.) analytical column. The mobile phase comprised phosphate buffer (20 mM; pH 3), acetonitrile (ACN), and methanol using gradient elution. Analyses were conducted at 35 degrees C and a 1.3-mL/min flow rate. The detection wavelength for all analytes was 210 nm. The samples were prepared by protein precipitation using ACN. Results: The HPLC-UV method was validated according to the FDA guidelines and applied to measure patient samples in TDM. Calibration curves showed excellent linearity (r(2) > 0.99) and covered the entire reference range for each analyte. Intraday and interday imprecisions and inaccuracies were <10% for all samples. Extensive stability testing showed no significant degradation (<15%), and interference measurements additionally ensured clinical applicability. Furthermore, the sensitivity was comparable with that of previously published HPLC methods using mass spectrometry. Conclusions: The authors developed an HPLC-UV method for the simultaneous quantification of 11 ASMs in the human serum and demonstrated its practical applicability in TDM. The method requires only standard laboratory equipment and simple sample preparation, making TDM available in less specialized laboratories.
引用
收藏
页码:345 / 350
页数:6
相关论文
共 50 条
  • [21] Simultaneous determination of codeine and it seven metabolites in plasma and urine by high-performance liquid chromatography with ultraviolet and electrochemical detection
    He, HB
    Shay, SD
    Caraco, Y
    Wood, M
    Wood, AJJ
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 1998, 708 (1-2): : 185 - 193
  • [22] Simultaneous quantification of levofloxacin, pefloxacin, ciprofloxacin and moxifloxacin in microvolumes of human plasma using high-performance liquid chromatography with ultraviolet detection
    Zheng, Yi
    Wang, Ziqing
    Lui, Gabrielle
    Hirt, Deborah
    Treluyer, Jean-Marc
    Benaboud, Sihem
    Aboura, Radia
    Gana, Ines
    BIOMEDICAL CHROMATOGRAPHY, 2019, 33 (05)
  • [23] SIMULTANEOUS QUANTIFICATION OF ZIDOVUDINE AND ITS METABOLITES IN SERUM AND URINE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY USING A COLUMN-SWITCHING TECHNIQUE
    LACROIX, C
    HOANG, TP
    WOJCIECHOWSKI, F
    DUWOOS, H
    NOUVEAU, J
    DIQUET, B
    JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1990, 525 (01): : 240 - 245
  • [24] Quantification of fluoxetine and norfluoxetine serum levels by reversed-phase high-performance liquid chromatography with ultraviolet detection
    Holladay, JW
    Dewey, MJ
    Yoo, SD
    JOURNAL OF CHROMATOGRAPHY B, 1997, 704 (1-2): : 259 - 263
  • [25] Determination of several estrogens in serum by high-performance liquid chromatography with ultraviolet detection
    Cai, YQ
    Wang, L
    Liang, LN
    Liu, JS
    Shi, YL
    Mou, SF
    Jiang, GB
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2005, 33 (07) : 913 - 916
  • [26] DETERMINATION OF AMIKACIN IN SERUM BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH ULTRAVIOLET DETECTION
    BARENDS, DM
    BLAUW, JS
    SMITS, MH
    HULSHOFF, A
    JOURNAL OF CHROMATOGRAPHY, 1983, 276 (02): : 385 - 394
  • [27] SIMULTANEOUS DETERMINATION OF TRAPIDIL AND ITS METABOLITES IN SERUM BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    NAKAGAWA, H
    MAEYAMA, S
    HIGASHITANI, T
    NAKAHATA, K
    HOSHINO, F
    FUKUI, Y
    YAKUGAKU ZASSHI-JOURNAL OF THE PHARMACEUTICAL SOCIETY OF JAPAN, 1984, 104 (06): : 635 - 639
  • [28] Quantification of letermovir in human serum using high-performance liquid chromatography with diode array detection
    Isberner, Nora
    Aghai, Fatemeh
    Schirmer, Diana
    Kraus, Sabrina
    Grigoleit, Goetz Ulrich
    Klinker, Hartwig
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2020, 1159
  • [29] Therapeutic isoniazid monitoring using a simple high-performance liquid chromatographic method with ultraviolet detection
    Moussa, LA
    Khassouani, CE
    Soulaymani, R
    Jana, M
    Cassanas, G
    Alric, R
    Hüe, B
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2002, 766 (01): : 181 - 187
  • [30] SIMULTANEOUS QUANTIFICATION OF ZIDOVUDINE AND ITS GLUCURONIDE IN SERUM BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    GOOD, SS
    REYNOLDS, DJ
    DEMIRANDA, P
    JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1988, 431 (01): : 123 - 133