Highly selective direct determination of chlorate ions by using a newly developed potentiometric electrode based on modified smectite

被引:23
|
作者
Topcu, Cihan [1 ]
机构
[1] Erzincan Univ, Fac Arts & Sci, Dept Chem, Erzincan, Turkey
关键词
Chlorate selective electrode; Potentiometric chlorate determination; Modified smectite; Water analysis; FLOW-INJECTION ANALYSIS; CARBON-PASTE ELECTRODE; MEMBRANE-ELECTRODE; SPECTROPHOTOMETRIC DETERMINATION; ACTIVE COMPONENT; DRINKING-WATER; MODIFIED CLINOPTILOLITE; IODOMETRIC METHOD; LEVEL CHLORATE; SURFACTANT;
D O I
10.1016/j.talanta.2016.09.018
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel polyvinyl chloride membrane chlorate (ClO3-) selective electrode based on modified smectite was developed for the direct determination of chlorate ions and the potentiometric performance characteristics of its were examined. The best selectivity and sensitivity for chlorate "ions were obtained for the electrode membrane containing ionophore/polyvinylchloride/o-nitrophenyloctylether in composition of 12/28/60 (w/w%). The proposed electrode showed a Nernstian response toward chlorate ions at pH=7 in the concentration range of 1 x 10(-7)-1 x 10(-1) M and the limit of detection was calculated as 9 x 10(-8) M from the constructed response plot. The linear slope of the electrode was -61 +/- 1 mV decade(-1) for chlorate activity in the mentioned linear working range. The selectivity coefficients were calculated according to both the matched potential method and the separate solution method. The calculated selectivity coefficients showed that the electrode performed excellent selectivity for chlorate ions. The potentiometric response of electrode toward chlorate ions was found to be highly reproducible. The electrode potential was stable between pH=4-10 and it had a dynamic response time of <5 s. The potentiometric behavior of the electrode in partial non-aqueous medium was also investigated and the obtained results (up to 5% (v/v) alcohol) were satisfactory. The proposed electrode was used during 15 weeks without any significant change in its potential response. Additionally, the electrode was very useful in water analysis studies such as dam water, river water, tap water, and swimming pool water where the direct determination of chlorate ions was required. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:623 / 631
页数:9
相关论文
共 50 条