Scatter correction for cone-beam computed tomography using moving blocker strips

被引:0
作者
Wang, Jing [1 ]
Mao, Weihua [1 ]
Solberg, Timothy [1 ]
机构
[1] Univ Texas SW Med Ctr Dallas, Dept Radiat Oncol, Dallas, TX 75390 USA
来源
MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING | 2011年 / 7961卷
关键词
scatter correction; computed tomography; cone-beam CT; moving blocker; imaging artifacts; X-RAY SCATTER; FLAT-PANEL IMAGER; RADIATION-THERAPY; AIR GAPS; DIAGNOSTIC-RADIOLOGY; DIGITAL RADIOGRAPHY; MICRO-CT; RECONSTRUCTION; REDUCTION; NOISE;
D O I
10.1117/12.877718
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
One well-recognized challenge of cone-beam computed tomography (CBCT) is the presence of scatter contamination within the projection images. Scatter degrades the CBCT image quality by decreasing the contrast, introducing shading artifacts and leading to inaccuracies in the reconstructed CT number. We propose a blocker-based approach to simultaneously estimate scatter signal and reconstruct the complete volume within the field of view (FOV) from a single CBCT scan. A physical strip attenuator (i.e., "blocker"), consists of lead strips, is inserted between the x-ray source and the patient. The blocker moves back and forth along z-axis during the gantry rotation. The two-dimensional (2D) scatter fluence is estimated by interpolating the signal from the blocked regions. A modified Feldkamp-Davis-Kress (FDK) algorithm and an iterative reconstruction based on the constraint optimization are used to reconstruct CBCT images from un-blocked projection data after the scatter signal is subtracted. An experimental study is performed to evaluate the performance of the proposed scatter correction scheme. The scatter-induced shading/cupping artifacts are substantially reduced in CBCT using the proposed strategy. In the experimental study using a CatPhan (R) 600 phantom, CT number errors in the selected regions of interest are reduced from 256 to less than 20. The proposed method allows us to simultaneously estimate the scatter signal in projection data, reduce the imaging dose and obtain complete volumetric information within the FOV.
引用
收藏
页数:12
相关论文
共 36 条
[1]   MONTE-CARLO SIMULATION OF THE SCATTERED RADIATION DISTRIBUTION IN DIAGNOSTIC-RADIOLOGY [J].
BOONE, JM ;
SEIBERT, JA .
MEDICAL PHYSICS, 1988, 15 (05) :713-720
[2]   AN ANALYTICAL MODEL OF THE SCATTERED RADIATION DISTRIBUTION IN DIAGNOSTIC-RADIOLOGY [J].
BOONE, JM ;
SEIBERT, JA .
MEDICAL PHYSICS, 1988, 15 (05) :721-725
[3]   Robust uncertainty principles:: Exact signal reconstruction from highly incomplete frequency information [J].
Candès, EJ ;
Romberg, J ;
Tao, T .
IEEE TRANSACTIONS ON INFORMATION THEORY, 2006, 52 (02) :489-509
[4]   Prior image constrained compressed sensing (PICCS): A method to accurately reconstruct dynamic CT images from highly undersampled projection data sets [J].
Chen, Guang-Hong ;
Tang, Jie ;
Leng, Shuai .
MEDICAL PHYSICS, 2008, 35 (02) :660-663
[5]   Effect of scattered radiation on image noise in cone beam CT [J].
Endo, M ;
Tsunoo, T ;
Nakamori, N ;
Yoshida, K .
MEDICAL PHYSICS, 2001, 28 (04) :469-474
[6]   Flat-panel cone-beam computed tomography for image-guided radiation therapy [J].
Jaffray, DA ;
Siewerdsen, JH ;
Wong, JW ;
Martinez, AA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (05) :1337-1349
[7]   Characterization of scattered radiation in kV CBCT images using Monte Carlo simulations [J].
Jarry, Genevieve ;
Graham, Sean A. ;
Moseley, Douglas J. ;
Jaffray, David J. ;
Siewerdsen, Jeffrey H. ;
Verhaegen, Frank .
MEDICAL PHYSICS, 2006, 33 (11) :4320-4329
[8]   Penalized-likelihood sinogram smoothing for low-dose CT [J].
La Rivière, PJ .
MEDICAL PHYSICS, 2005, 32 (06) :1676-1683
[9]   Scatter kernel estimation with an edge-spread function method for cone-beam computed tomography imaging [J].
Li, Heng ;
Mohan, Radhe ;
Zhu, X. Ronald .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (23) :6729-6748
[10]  
Liu X, 2006, P SOC PHOTO-OPT INST, V6142, P6142341