Highly sensitive determination of iodide by ion chromatography with amperometric detection at a silver-based carbon paste electrode

被引:79
作者
Malongo, Tresor Kimbeni [1 ,2 ]
Patris, Stephanie [1 ]
Macours, Pascale [3 ]
Cotton, Frederic [1 ,3 ]
Nsangu, Jean [2 ]
Kauffmann, Jean-Michel [1 ]
机构
[1] Univ Libre Bruxelles, Inst Pharm, Campus Plaine,CP 205-06, B-1050 Brussels, Belgium
[2] Univ Kinshasa, Fac Pharmaceut Sci, Kinshasa, DEM REP CONGO
[3] Erasme Univ Hosp, Dept Clin Chem, Brussels, Belgium
关键词
iodide; silver electrode; carbon paste; ion chromatography; amperometry;
D O I
10.1016/j.talanta.2008.03.053
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A silver-based solid carbon paste electrode was developed for use as a detector in ion chromatography (IC) for the sensitive determination of iodide in real samples. Micro- and nano-particles of silver were investigated for the fabrication of different electrodes. The iodide assay was based on IC with amperometric detection (IC-AD) at a silver composite electrode polarized at +0.080 V versus Ag/AgCl. Free iodide and organoiodide compounds were studied. The detection process was characterized by studying the redox behavior of iodide ions at both silver and silver composite electrodes by cyclic voltammetry (CV). The presence of iodide ions in solution was found to considerably facilitate metallic silver oxidation, with response currents directly related to iodide concentration. The calibration curve at the selected silver carbon paste electrode was linear in the concentration range comprised between 0.635 mu g/L and 63.5 mu g/L iodide. The relative standard deviation (R.S.D.) for successive injections was below 3% for all iodide standard solutions investigated. The limit of detection (LOD) was 0.47 mu g/L (3.7 nmol/L) for an injection volume of 20 mu L, i.e. 74 fmol injected. The IC-AD method was successfully applied to the determination of iodide in complex real samples such as table salts, sea products and iodide bound drug compounds. The analytical accuracy was verified by the assay of iodide in milk powder from an iodide certified reference material (CRM) Community Bureau of Reference (BCR) 150. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:540 / 547
页数:8
相关论文
共 50 条
[1]   Stripping voltammetry of bromide and iodide ions at a renewal silver electrode [J].
Aleksandrova, TP ;
Kletenik, YB .
JOURNAL OF ANALYTICAL CHEMISTRY, 2000, 55 (06) :590-593
[2]   Ion chromatographic determination of iodide in urine and serum using a tubular ion-selective electrode based on a homogeneous crystalline membrane [J].
Almeida, AA ;
Jun, X ;
Lima, JLFC .
MIKROCHIMICA ACTA, 1997, 127 (1-2) :55-60
[3]   Reappraisal of the risk of iodine-induced hyperthyroidism: An epidemiological population survey [J].
Azizi, F ;
Hedayati, M ;
Rahmani, M ;
Sheikholeslam, R ;
Allahverdian, S ;
Salarkia, N .
JOURNAL OF ENDOCRINOLOGICAL INVESTIGATION, 2005, 28 (01) :23-29
[4]   Determination of iodide in urine by ion-pair chromatography with electrochemical detection [J].
Below, H ;
Kahlert, H .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2001, 371 (04) :431-436
[5]   Iodide analysis by anion-exchange chromatography and pulsed amperometric detection in surface water and adsorbable organic iodide [J].
Bruggink, Cees ;
van Rossum, Willem J. M. ;
Spijkerman, Ed ;
van Beelen, Eric S. E. .
JOURNAL OF CHROMATOGRAPHY A, 2007, 1144 (02) :170-174
[6]  
BUCHBERGER W, 1985, MIKROCHIM ACTA, V3, P347, DOI 10.1007/BF01196833
[7]   Sensitive quantification of iodide by ion-exchange chromatography with electrochemical detection at a modified platinum electrode [J].
Cataldi, TRI ;
Rubino, A ;
Ciriello, R .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2005, 382 (01) :134-141
[8]   Comparison of silver, gold and modified platinum electrodes for the electrochemical detection of iodide in urine samples following ion chromatography [J].
Cataldi, TRI ;
Rubino, A ;
Laviola, MC ;
Ciriello, R .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2005, 827 (02) :224-231
[9]   Simultaneous amperometric and potentiometric detection of inorganic anions in flow systems using platinum and silver silver chloride electrodes [J].
Chen, Z ;
Hibbert, DB .
ANALYTICA CHIMICA ACTA, 1997, 350 (1-2) :1-6
[10]   Pulsed amperometric detection of sulfide, cyanide, iodide, thiosulfate, bromide and thiocyanate with microfabricated disposable silver working electrodes in ion chromatography [J].
Cheng, J ;
Jandik, P ;
Avdalovic, N .
ANALYTICA CHIMICA ACTA, 2005, 536 (1-2) :267-274