A Promising Approach for Improving the Coating Stability and In Vivo Performance of Biodegradable Polymer-Coated Sirolimus-Eluting Stent

被引:16
作者
Bedair, Tarek M. [1 ,2 ,3 ]
Kang, Sung Nam [1 ,2 ]
Joung, Yoon Ki [1 ,2 ]
Han, Dong Keun [1 ,2 ]
机构
[1] Korea Inst Sci & Technol, Biomed Res Inst, Ctr Biomat, Hwarangno 14 Gil 5, Seoul 02792, South Korea
[2] Korea Univ Sci & Technol, Dept Biomed Engn, 113 Gwahangno, Daejeon 34113, South Korea
[3] Minia Univ, Fac Sci, Dept Chem, El Minia 61519, Egypt
基金
新加坡国家研究基金会;
关键词
Drug-Eluting Stent; Polymer Brush; Drug-In-Polymer Coating; Durability; In-Stent Restenosis; SELF-ASSEMBLED MONOLAYERS; CORONARY STENTS; RELEASE KINETICS; DRUG-RELEASE; SURFACE; PROLIFERATION; DURABILITY; RESTENOSIS; COPOLYMERS; MECHANISM;
D O I
10.1166/jbn.2016.2305
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
Several developments are in progress for improving the performance of drug-eluting stents (DESs) including use of biodegradable polymers, polymer-free DES, fully bioabsorbable stents, and so on. The commercially available DESs still suffer from polymer defects that could affect the performance of a DES through a series of adverse events such as coating delamination and/ or peeling-off that lead to non-uniform local drug distribution, restenosis, and thrombosis. The goal of this work was to enhance the stability of drug-in-polymer matrix coating on a stent metal surface through surface modification. The cobalt-chromium (Co-Cr) surface was chemically modified using poly(dopamine) (PDA) nano-coating and poly(L-lactide) (PLLA) nano-brush in order to be applied to a biodegradable polymer-coated DES. The biodegradable polymer loaded with sirolimus was coated using an ultrasonic spray coating instrument. The coating morphology on all samples showed a very smooth and uniform coating. The stability of the coating was evaluated for 2 months under the circulation system in which the drug-in-polymer coating on the PLLA brush-modified stent presented the most stable coating behavior as compared to other samples. The in vitro sirolimus release study from both unmodified and modified stents was studied in phosphate-buffered saline (PBS), and the modified stents showed slower sirolimus release profile as compared to unmodified stents. In vivo study was performed in a porcine coronary artery injury model for 28 days. The percentage of in-stent restenosis area (ISR) for PLLA brush-modified sirolimus-eluting stent (SES) decreased significantly as compared to unmodified SES and bare metal stent (BMS). This study demonstrated that the modification of stent surface using PLLA brushes affects in vitro and in vivo performance effectively to be applied for biodegradable polymercoated DES.
引用
收藏
页码:2015 / 2028
页数:14
相关论文
共 61 条
[1]   Optimization of Cardiovascular Stent against Restenosis: Factorial Design-Based Statistical Analysis of Polymer Coating Conditions [J].
Acharya, Gayathri ;
Lee, Chi H. ;
Lee, Yugyung .
PLOS ONE, 2012, 7 (08)
[2]   Effect of molecular structure of self-assembled monolayers on their tribological behaviors in nano- and microscales [J].
Ahn, HS ;
Cuong, PD ;
Park, S ;
Kim, YW ;
Lim, JC .
WEAR, 2003, 255 :819-825
[3]   Biodegradation and biocompatibility of PLA and PLGA microspheres [J].
Anderson, JM ;
Shive, MS .
ADVANCED DRUG DELIVERY REVIEWS, 1997, 28 (01) :5-24
[4]   Scanning electron microscopic assessment of the biodegradable coating on expanded biolimus-eluting stents [J].
Basalus, Mounir W. Z. ;
van Houwelingen, K. Gert ;
Ankone, Marc ;
de Man, Frits H. A. F. ;
von Birgelen, Clemens .
EUROINTERVENTION, 2009, 5 (04) :505-510
[5]   Coating defects in polymer-coated drug-eluting stents [J].
Bedair, Tarek M. ;
Cho, Youngjin ;
Park, Bang Ju ;
Joung, Yoon Ki ;
Han, Dong Keun .
BIOMATERIALS AND BIOMECHANICS IN BIOENGINEERING, 2014, 1 (03) :131-150
[6]   Effects of interfacial layer wettability and thickness on the coating morphology and sirolimus release for drug-eluting stent [J].
Bedair, Tarek M. ;
Yu, Seung Jung ;
Im, Sung Gap ;
Park, Bang Ju ;
Joung, Yoon Ki ;
Han, Dong Keun .
JOURNAL OF COLLOID AND INTERFACE SCIENCE, 2015, 460 :189-199
[7]   Biodegradable polymer brush as nanocoupled interface for improving the durability of polymer coating on metal surface [J].
Bedair, Tarek M. ;
Cho, Youngjin ;
Joung, Yoon Ki ;
Han, Dong Keun .
COLLOIDS AND SURFACES B-BIOINTERFACES, 2014, 122 :808-817
[8]   Reinforcement of Interfacial Adhesion of a Coated Polymer Layer on a Cobalt-Chromium Surface for Drug-Eluting Stents [J].
Bedair, Tarek M. ;
Cho, Youngjin ;
Kim, Tae Jung ;
Kim, Young Dong ;
Park, Bang Ju ;
Joung, Yoon Ki ;
Han, Dong Keun .
LANGMUIR, 2014, 30 (27) :8020-8028
[9]   Stability of phosphonic self assembled monolayers (SAMs) on cobalt chromium (Co-Cr) alloy under oxidative conditions [J].
Bhure, Rahul ;
Abdel-Fattah, Tarek M. ;
Bonner, Carl ;
Hall, Felicia ;
Mahapatro, Anil .
APPLIED SURFACE SCIENCE, 2011, 257 (13) :5605-5612
[10]   Formation of Nanosized Phosphonic Acid Self Assembled Monolayers on Cobalt-Chromium Alloy for Potential Biomedical Applications [J].
Bhure, Rahul ;
Abdel-Fattah, Tarek M. ;
Bonner, Carl ;
Hall, Joseph C. ;
Mahapatro, Anil .
JOURNAL OF BIOMEDICAL NANOTECHNOLOGY, 2010, 6 (02) :117-128