In vivo MRI visualization of mesh shrinkage using surgical implants loaded with superparamagnetic iron oxides

被引:44
作者
Kuehnert, Nicolas [1 ]
Kraemer, Nils A. [2 ]
Otto, Jens [1 ]
Donker, Hank C. W. [2 ]
Slabu, Ioana [3 ]
Baumann, Martin [3 ]
Kuhl, Christiane K. [2 ]
Klinge, Uwe [1 ,3 ]
机构
[1] Rhein Westfal TH Aachen, Dept Surg, Univ Hosp, D-52074 Aachen, Germany
[2] Rhein Westfal TH Aachen, Dept Diagnost Radiol, Univ Hosp, D-52074 Aachen, Germany
[3] Rhein Westfal TH Aachen, Helmholtz Inst Appl Med Engn, D-52074 Aachen, Germany
来源
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES | 2012年 / 26卷 / 05期
关键词
Magnetic resonance imaging; SPIO; Mesh implant; Shrinkage; INCISIONAL HERNIA REPAIR; FUNCTIONAL CINE MRI; ADHESION FORMATION; COMPLICATION RATES; SYNTHETIC MESH; ULTRASONOGRAPHY; POLYPROPYLENE; EXCISION;
D O I
10.1007/s00464-011-2057-7
中图分类号
R61 [外科手术学];
学科分类号
摘要
Prosthetic mesh implants are widely used in hernia surgery. To show long-term mesh-related complications such as shrinkage or adhesions, a precise visualization of meshes and their vicinity in vivo is important. By supplementing mesh fibers with ferro particles, magnetic resonance imaging (MRI) can help to delineate the mesh itself. This study aimed to demonstrate and quantify time-dependent mesh shrinkage in vivo by MRI. Polyvinylidenfluoride (PVDF) meshes with incorporated superparamagnetic iron oxides (SPIOs) were implanted as an abdominal wall replacement in 30 rats. On days 1, 7, 14, or 21, MRI was performed using a gradient echo sequence with repetition time (TR)/echo time (TE) of 50/4.6 and a flip angle of 20A degrees. The length, width, and area of the device were measured on axial, coronal, and sagittal images, and geometric deformations were assessed by surgical explantation. In all cases, the meshes were visualized and their area estimated by measuring the length and width of the mesh. The MRI presented a mean area shrinkage in vivo of 13% on day 7, 23% on day 14, and 23% on day 21. Postmortem measurements differed statistically from MRI, with a mean area shrinkage of 23% on day 7, 28% on day 14, and 30% on day 21. Ex vivo measurements of shrinkage showed in vivo measurements to be overestimated approximately 8%. Delineation of the mesh helped to show folding or adhesions close to the intestine. Loading of surgical meshes with SPIOs allows their precise visualization during MRI and guarantees an accurate in vivo assessment of their shrinkage. The authors' observation clearly indicates that shrinkage in vivo is remarkably less than that shown by illustrated explantation measurements. The use of MRI with such meshes could be a reliable technique for checking on proper operation of implanted meshes and showing related complications, obviating the need for exploratory open surgical revision.
引用
收藏
页码:1468 / 1475
页数:8
相关论文
共 33 条
  • [1] Long-term recurrence and complication rates after incisional hernia repair with the open onlay technique
    Andersen, Lars Peter Holst
    Klein, Mads
    Gogenur, Ismail
    Rosenberg, Jacob
    [J]. BMC SURGERY, 2009, 9
  • [2] Differences in polypropylene shrinkage depending on mesh position in an experimental study
    Angel Garcia-Urena, Miguel
    Vega Ruiz, Vicente
    Diaz Godoy, Antonio
    Baez Perea, Jose Maria
    Marin Gomez, Luis Miguel
    Carnero Hernandez, Francisco Javier
    Velasco Garcia, Miguel Angel
    [J]. AMERICAN JOURNAL OF SURGERY, 2007, 193 (04) : 538 - 542
  • [3] Polyvinylidene fluoride: a suitable mesh material for laparoscopic incisional and parastomal hernia repair!
    Berger, D.
    Bientzle, M.
    [J]. HERNIA, 2009, 13 (02) : 167 - 172
  • [4] Effect of tissue shrinkage on histological tumour-free margin after excision of basal cell carcinoma
    Blasdale, C.
    Charlton, F. G.
    Weatherhead, S. C.
    Ormond, P.
    Lawrence, C. M.
    [J]. BRITISH JOURNAL OF DERMATOLOGY, 2010, 162 (03) : 607 - 610
  • [5] Evaluation of new prosthetic meshes for ventral hernia repair
    Burger, J. W. A.
    Halm, J. A.
    Wijsmuller, A. R.
    ten Raa, S.
    Jeekel, J.
    [J]. SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2006, 20 (08): : 1320 - 1325
  • [6] New Polymer for Intra-Abdominal Meshes-PVDF Copolymer
    Conze, Joachim
    Junge, Karsten
    Weiss, Claudia
    Anurov, Michael
    Oettinger, Alexander
    Klinge, Uwe
    Schumpelick, V.
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2008, 87B (02) : 321 - 328
  • [7] Incisional hernia: challenge of re-operations after mesh repair
    Conze, Joachim
    Krones, Carsten J.
    Schumpelick, Volker
    Klinge, Uwe
    [J]. LANGENBECKS ARCHIVES OF SURGERY, 2007, 392 (04) : 453 - 457
  • [8] Shrinkage of skin excision specimens: formalin fixation is not the culprit
    Dauendorffer, J. N.
    Bastuji-Garin, S.
    Guero, S.
    Brousse, N.
    Fraitag, S.
    [J]. BRITISH JOURNAL OF DERMATOLOGY, 2009, 160 (04) : 810 - 814
  • [9] Laparoscopic repair of incisional hernia: Outcomes of 100 consecutive cases comprising 25 wall defects larger than 15 cm
    Ferrari, Giovanni Carlo
    Miranda, Angelo
    Di Lernia, Stefano
    Sansonna, Fabio
    Magistro, Carmelo
    Maggioni, Dario
    Scandroglio, Ildo
    Costanzi, Andrea
    Franzetti, Maurizio
    Pugliese, Raffaele
    [J]. SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2008, 22 (05): : 1173 - 1179
  • [10] Functional cine MRI of the abdomen for the assessment of implanted synthetic mesh in patients after incisional hernia repair: initial results
    Fischer, Tanja
    Ladurner, Roland
    Gangkofer, Alexander
    Mussack, Thomas
    Reiser, Maximilian
    Lienemann, Andreas
    [J]. EUROPEAN RADIOLOGY, 2007, 17 (12) : 3123 - 3129