Denoising analysis of compact CCD-based spectrometer

被引:23
作者
Huang, Chan [1 ]
Xia, Guo [1 ]
Jin, Shiqun [1 ]
Hu, Mingyong [1 ]
Wu, Su [1 ]
Xing, Jinyu [2 ]
机构
[1] Hefei Univ Technol, Key Lab Special Display Technol, Minist Educ,Natl Key Lab Adv Display Technol, Natl Engn Lab Special Display Technol,Acad Photoe, Hefei 230009, Anhui, Peoples R China
[2] Hefei Univ Technol, Sch Instrument Sci & Optoelect Engn, Hefei 230009, Anhui, Peoples R China
来源
OPTIK | 2018年 / 157卷
关键词
Spectrometers and spectroscopy; CCD; Spectrum analysis; NOISE;
D O I
10.1016/j.ijleo.2017.11.170
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
CCD spectrometers are widely used in many fields for quick and accurate measurement of spectra. This paper firstly analyzes various noise phenomena by using Ocean Optics USB4000 spectrometer, including readout noise, photoelectron noise, dark noise and fixed pattern noise. Secondly, noise models of CCD-based spectrometers are built based on the theoretical analysis of noise phenomena. A number of noise-limiting techniques are applied to broadband spectra and line spectra with different noise levels. Lastly by comparing the denoising results, the optimal denoising algorithms are proposed for different spectral characteristics and noise levels. (C) 2017 Elsevier GmbH. All rights reserved.
引用
收藏
页码:693 / 706
页数:14
相关论文
共 50 条
  • [1] A CCD-based γ-ray dosimeter
    Harris, EJ
    Royle, GJ
    Mooney, MJ
    Speller, RD
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 458 (1-2) : 227 - 232
  • [2] CCD-based vertex detector for JLC
    Aso, T
    Iwai, G
    Fujiwara, K
    Takayama, H
    Tamura, N
    Sugimoto, Y
    Miyamoto, A
    Abe, K
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 511 (1-2) : 235 - 239
  • [3] CCD-based vertex detector for ILC
    Stefanov, Konstantin D.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 569 (01) : 48 - 52
  • [4] CCD-based vertex detector for GLC
    Sugimoto, Y
    Iwai, G
    Abe, K
    Aso, T
    Kujiwara, K
    Katou, Y
    Miyamoto, A
    Nakayoshi, K
    Takayama, H
    Tamura, N
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 549 (1-3) : 87 - 92
  • [5] Developments in CCD-based beta/X-ray imaging
    Wells, K
    Kokkinou, E
    Petrou, M
    Ranicar, A
    Spinks, T
    Ott, RJ
    PENETRATING RADIATION SYSTEMS AND APPLICATIONS III, 2001, 4508 : 99 - 110
  • [6] Accuracy control and influence analysis of wavelength calibration for compact CCD spectrometer
    Liu L.
    Li Y.
    Zheng F.
    Zhang G.
    Xu Y.
    Yi Qi Yi Biao Xue Bao/Chinese Journal of Scientific Instrument, 2019, 40 (05): : 19 - 27
  • [7] Feasibility study of CCD-based gamma camera
    Lee, Hakjae
    Jung, Young-Jun
    Yoon, Joochul
    Kang, Jungwon
    Lee, Sangjoon
    Shin, Hyungsup
    Lee, Kisung
    HARD X-RAY, GAMMA-RAY, AND NEUTRON DETECTOR PHYSICS XII, 2010, 7805
  • [8] A Fast Double Shutter for CCD-Based Metrology
    Geisler, R.
    SELECTED PAPERS FROM THE 31ST INTERNATIONAL CONGRESS ON HIGH-SPEED IMAGING AND PHOTONICS, 2017, 10328
  • [9] Novel CCD-based pH imaging sensor
    Sawada, K
    Mimura, S
    Tomita, K
    Nakanishi, T
    Tanabe, H
    Ishida, M
    Ando, T
    IEEE TRANSACTIONS ON ELECTRON DEVICES, 1999, 46 (09) : 1846 - 1849
  • [10] Evaluation of a CCD-based film digitizer for digital mammography
    Kallergi, M
    Gavrielides, MA
    Gross, WW
    Clarke, LP
    PHYSICS OF MEDICAL IMAGING - MEDICAL IMAGING 1997, 1997, 3032 : 282 - 291