A hybrid (iron-fat-water) phantom for liver iron overload quantification in the presence of contaminating fat using magnetic resonance imaging

被引:10
作者
Mobini, Nazanin [1 ,2 ]
Malekzadeh, Malakeh [2 ,3 ]
Haghighatkhah, Hamidreza [4 ]
Rad, Hamidreza Saligheh [1 ,2 ]
机构
[1] Univ Tehran Med Sci, Sch Med, Dept Med Phys & Biomed Engn, Tehran, Iran
[2] Univ Tehran Med Sci, Imam Khomeini Hosp Complex, Inst Adv Med Technol, Quantitat MR Imaging & Spect Grp QMISG,Res Ctr Ce, Keshavarz Blvd, Tehran 1419733141, Iran
[3] Iran Univ Med Sci IUMS, Sch Med, Med Phys Dept, Tehran, Iran
[4] Shahid Beheshti Univ Med Sci SBMU, Shohada E Tajrish Hosp, Dept Radiol, Tehran, Iran
关键词
Iron-fat-water phantom; Magnetic resonance imaging; Iron overload; Fatty liver; MRI; IMAGES;
D O I
10.1007/s10334-019-00795-7
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective Assessment of iron content in the liver is crucial for diagnosis/treatment of iron-overload diseases. Nonetheless, T-2*-based methods become challenging when fat and iron are simultaneously present. This study proposes a phantom design concomitantly containing various concentrations of iron and fat suitable for devising accurate simultaneous T-2* and fat quantification technique. Materials and methods A 46-vial iron-fat-water phantom with various iron concentrations covering clinically relevant T-2* relaxation time values, from healthy to severely overloaded liver and wide fat percentages ranges from 0 to 100% was prepared. The phantom was constructed using insoluble iron (II, III) oxide powder containing microscale particles. T-2*-weighted imaging using multi-gradient-echo (mGRE) sequence, and chemical shift imaging spin-echo (CSI-SE) Magnetic Resonance Spectroscopy (MRS) data were considered for the analysis. T-2* relaxation times and fat fractions were extracted from the MR signals to explore the effects of fat and iron overload. Results Size distribution of iron oxide particles for Magnetite fits with a lognormal function with a mean size of about 1.17 mu m. Comparison of FF color maps, estimated from bi- and mono-exponential model indicated that single-T-2* fitting model resulted in lower NRMSD. Therefore, T-2* values from the mono-exponential signal equation were used and expressed the relationship between relaxation time value across all iron (Fe) and fat concentration as Fe=-28.02+302.84T2*-0.045FF with R-squared = 0.89. Discussion The proposed phantom design with microsphere iron particles closely simulated the single-T-2* behavior of fatty iron-overloaded liver in vivo.
引用
收藏
页码:385 / 392
页数:8
相关论文
共 28 条
  • [1] Medical progress: Disorders of iron metabolism
    Andrews, NC
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1999, 341 (26) : 1986 - 1995
  • [2] Iron overload syndromes and the liver
    Batts, Kenneth P.
    [J]. MODERN PATHOLOGY, 2007, 20 : S31 - S39
  • [3] Comparison of fat quantification methods: A phantom study at 3.0T
    Bernard, Clare P.
    Liney, Gary P.
    Manton, David J.
    Turnbull, Lindsay W.
    Langton, Christian M.
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 27 (01) : 192 - 197
  • [4] Application of the static dephasing regime theory to superparamagnetic iron-oxide loaded cells
    Bowen, CV
    Zhang, XW
    Saab, G
    Gareau, PJ
    Rutt, BK
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2002, 48 (01) : 52 - 61
  • [5] Noninvasive measurement of iron: report of an NIDDK workshop
    Brittenham, GM
    Badman, DG
    [J]. BLOOD, 2003, 101 (01) : 15 - 19
  • [6] Sources of systematic error in proton density fat fraction (PDFF) quantification in the liver evaluated from magnitude images with different numbers of echoes
    Bydder, Mark
    Hamilton, Gavin
    de Rochefort, Ludovic
    Desai, Ajinkya
    Heba, Elhamy R.
    Loomba, Rohit
    Schwimmer, Jeffrey B.
    Szeverenyi, Nikolaus M.
    Sirlin, Claude B.
    [J]. NMR IN BIOMEDICINE, 2018, 31 (01)
  • [7] Bi-exponential proton transverse relaxation rate (R2) image analysis using RF field intensity-weighted spin density projection:: potential for R2 measurement of iron-loaded liver
    Clark, PR
    Chua-anusorn, W
    St Pierre, TG
    [J]. MAGNETIC RESONANCE IMAGING, 2003, 21 (05) : 519 - 530
  • [8] Ultra-short echo time images quantify high liver iron
    Doyle, Eamon K.
    Toy, Kristin
    Valdez, Bertin
    Chia, Jonathan M.
    Coates, Thomas
    Wood, John C.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2018, 79 (03) : 1579 - 1585
  • [9] Evaluation of six-point modified dixon and magnetic resonance spectroscopy for fat quantification: a fat–water–iron phantom study
    Fukuzawa K.
    Hayashi T.
    Takahashi J.
    Yoshihara C.
    Tano M.
    Kotoku J.
    Saitoh S.
    [J]. Radiological Physics and Technology, 2017, 10 (3) : 349 - 358
  • [10] Gupta C. P., 2014, IOSR J APPL CHEM, V7, P38, DOI DOI 10.9790/5736-071123846