A swept-field aspiration condenser as an ion-mobility spectrometer

被引:20
作者
Sacristán, E [1 ]
Solis, AA [1 ]
机构
[1] UAM Iztapalapa, Dept Elect Engn, Mexico City, DF, Mexico
关键词
analytical chemistry; anesthetic agents; aspiration condenser; gas monitoring; ionization; ion mobility; spectrometry; Tammet transform;
D O I
10.1109/19.744345
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A method is presented for using a small, low-cost, extremely simple device as an ion-mobility spectrometer, The device is a planar first-order differential aspiration condenser as described by Puumalainen, but in this case the electric field is not kept constant but is swept in discrete nonuniform steps. The result is a characteristic I(V) curve that is a function of the ion-mobility density function of the gas sample flowing through the condenser. A discrete transform is introduced, based on the equations of Tammet describing the operation of aspiration condensers, that converts the I(V) curve into the ion-mobility density spectrum and vice versa, Both the device and the method are described in theoretical detail. The method is illustrated with computer simulations and experimental data using both a prototype aspiration condenser and a conventional drift tube ion-mobility spectrometer, The results suggest this method is an attractive alternative for ion-mobility gas analysis applications.
引用
收藏
页码:769 / 775
页数:7
相关论文
共 20 条
  • [1] ALLMAN SL, 1993, Patent No. 5184015
  • [2] BLYTH DA, 1983, Patent No. 4368388
  • [3] BROKENSHIRE J, 1989, ION MOBILITY SPECTRO
  • [4] MONITORING OF AIRBORNE ORGANIC VAPORS USING ION MOBILITY SPECTROMETRY
    EICEMAN, GA
    SNYDER, AP
    BLYTH, DA
    [J]. INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY, 1990, 38 (03) : 415 - 425
  • [5] ION MOBILITY SPECTROMETRY OF HALOTHANE, ENFLURANE, AND ISOFLURANE ANESTHETICS IN AIR AND RESPIRED GASES
    EICEMAN, GA
    SHOFF, DB
    HARDEN, CS
    SNYDER, AP
    MARTINEZ, PM
    FLEISCHER, ME
    WATKINS, ML
    [J]. ANALYTICAL CHEMISTRY, 1989, 61 (10) : 1093 - 1099
  • [6] EICEMAN GA, 1991, CRIT REV ANAL CHEM, V22, P17
  • [7] EICEMAN GA, 1994, ION MOBILITY SPECTRO
  • [8] GIESE W, 1882, ANN PHYSIK, V17, P236
  • [9] ION MOBILITY SPECTROMETRY
    HILL, HH
    SIEMS, WF
    STLOUIS, RH
    MCMINN, DG
    [J]. ANALYTICAL CHEMISTRY, 1990, 62 (23) : A1201 - A1209
  • [10] JENKINS A, 1989, Patent No. 4831254