Assessment of material loss of retrieved magnetically controlled implants for limb lengthening

被引:9
作者
Foong, Bryan [1 ,2 ]
Panagiotopoulou, Vasiliki-Christina [1 ,2 ]
Hothi, Harry S. [1 ,2 ]
Henckel, Johann [2 ]
Calder, Peter R. [2 ]
Goodier, David W. [2 ]
Hart, Alister J. [1 ,2 ]
机构
[1] Univ Coll London, Inst Orthopaed & Musculoskeletal Sci, Res Dept Mat & Tissue, Stanmore, Middx, England
[2] Royal Natl Orthopaed Hosp, Stanmore, Middx, England
关键词
Limb length discrepancy; intramedullary nails; distraction osteogenesis; magnetically controlled intramedullary nails; material loss; SKELETAL KINETIC DISTRACTOR; TOTAL HIP-ARTHROPLASTY; LEG-LENGTH; INTRAMEDULLARY NAIL; INEQUALITY; FAILURE; DISCREPANCY; SYSTEM; RATES; RODS;
D O I
10.1177/0954411918806329
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose: We aimed to understand wear from the telescopic component of PRECICE nails, which are used for distraction osteogenesis of the femur or tibia. We also aimed to identify any correlation between implant performance and patient factors. Methods: This retrieval study involved 11 magnetically controlled intramedullary nails from nine patients who had achieved the targeted leg length. All the nails were assessed macroscopically and microscopically for wear. All implants were radiographed to assess the internal mechanism. A Talyrond 365 (Taylor Hobson, Leicester, UK) roundness measuring machine was used to generate three-dimensional surface maps of the telescopic components to allow for measurement of wear. Results: Visual assessment of all the nails showed evidence of wear from the telescopic component. The radiographs revealed that all the nails had intact internal mechanism and no evidence of fractured pins. The roundness measuring machine showed that the quantity of wear was lowest in the latest design of the PRECICE nail. There was no significant correlation between wear and the two patient factors (duration of the lengthening phase, the time of implantation) included in this study. Conclusion: This study is the first to investigate the performance of the three different designs of the PRECICE system with a focus on wear. We found that the latest design had the best implant performance. We are confident of the continued success of the PRECICE system and reassure surgeons and patients that they are unlikely to encounter problems with the implant related to wear.
引用
收藏
页码:1129 / 1136
页数:8
相关论文
共 36 条
  • [1] Inequality in leg length is important for the understanding of the pathophysiology of lumbar disc herniation
    Balik, Mehmet Sabri
    Kanat, Ayhan
    Erkut, Adem
    Ozdemir, Bulent
    Batcik, Osman Ersagun
    [J]. JOURNAL OF CRANIOVERTEBRAL JUNCTION AND SPINE, 2016, 7 (02) : 87 - 90
  • [2] The effect on mechanical axis deviation of femoral lengthening with an intramedullary telescopic nail
    Burghardt, R. D.
    Paley, D.
    Specht, S. C.
    Herzenberg, J. E.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2012, 94B (09): : 1241 - 1245
  • [3] Mechanical failure of the Intramedullary Skeletal Kinetic Distractor in limb lengthening
    Burghardt, R. D.
    Herzenberg, J. E.
    Specht, S. C.
    Paley, D.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (05): : 639 - 643
  • [4] Pseudotumors Associated with Total Hip Arthroplasty
    Daniel, Joseph
    Holland, James
    Quigley, Laura
    Sprague, Sheila
    Bhandari, Mohit
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2012, 94A (01) : 86 - 93
  • [5] Symptoms of the knee and hip in individuals with and without limb length inequality
    Golightly, Y. M.
    Allen, K. D.
    Helmick, C. G.
    Renner, J. B.
    Jordan, J. M.
    [J]. OSTEOARTHRITIS AND CARTILAGE, 2009, 17 (05) : 596 - 600
  • [6] GUICHET JM, 1991, CLIN ORTHOP RELAT R, P235
  • [7] Gradual femoral lengthening with the Albizzia intramedullary nail
    Guichet, JM
    Deromedis, B
    Donnan, LT
    Peretti, G
    Lascombes, P
    Bado, F
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (05) : 838 - 848
  • [8] Association of Leg-Length Inequality With Knee Osteoarthritis A Cohort Study
    Harvey, William F.
    Yang, Mei
    Cooke, Theodore D. V.
    Segal, Neil A.
    Lane, Nancy
    Lewis, Cora E.
    Felson, David T.
    [J]. ANNALS OF INTERNAL MEDICINE, 2010, 152 (05) : 287 - W92
  • [9] Damage Patterns at the Head-Stem Taper Junction Helps Understand the Mechanisms of Material Loss
    Hothi, Harry S.
    Panagiotopoulos, Andreas C.
    Whittaker, Robert K.
    Bills, Paul J.
    McMillan, Rebecca A.
    Skinner, John A.
    Hart, Alister J.
    [J]. JOURNAL OF ARTHROPLASTY, 2017, 32 (01) : 291 - 295
  • [10] ILIZAROV GA, 1988, B HOSP JOINT DIS ORT, V48, P1