Evaluation of cobalt-carbon and palladium-carbon eutectic point cells for thermocouple calibration

被引:37
作者
Ogura, H. [1 ]
Izuchi, M. [1 ]
Arai, M. [1 ]
机构
[1] AIST, NMIJ, Tsukuba, Ibaraki, Japan
关键词
calibration; cobalt-carbon eutectic point; palladium-carbon eutectic point; Pt; Pd thermocouple;
D O I
10.1007/s10765-007-0320-x
中图分类号
O414.1 [热力学];
学科分类号
摘要
Two cobalt-carbon (Co-C) eutectic point (1,324 degrees C) cells and one palladium-carbon (Pd-C) eutectic point (1,492 degrees C) cell were constructed for thermocouple calibration. The lengths of the Co-C and Pd-C cells were 297 mm, 140 mm, and 140 mm, respectively. The melting and freezing plateaux at the Co-C and Pd-C eutectic points were observed using Pt/Pd thermocouples. The repeatability of the plateau, the effect of the surrounding temperature, and the temperature profile in the cell were measured, and the heat flux effect along the thermometer well was evaluated. When the plateaux of Co-C (297 mm height), Co-C (140 mm height), and Pd-C cells, were measured three times, seven times, and six times, respectively, the standard deviations of the melting points were 0.1 mu V, 0.1 mu V, and 0.4 mu V, respectively. According to the temperature profiles along the thermometer well during the melting plateaux, it was found that the Pt/Pd thermocouple should be inserted at least 9.5 cm, 5 cm, and 6 cm below the surface of the eutectic alloys in the Co-C (297 mm height), Co-C (140 mm height), and Pd-C cells with the furnace set-point 16 degrees C above the melting point.
引用
收藏
页码:210 / 221
页数:12
相关论文
共 13 条
[1]  
BATTUELLO M, 1998, P EUROMET WORKSH TEM, P107
[2]  
Burns GW, 2002, TEMPMEKO 2001: 8TH INTERNATIONAL SYMPOSIUM ON TEMPERATURE AND THERMAL MEASUREMENT IN INDUSTRY AND SCIENCE, VOL 1 & 2, PROCEEDINGS, P61
[3]   Platinum versus palladium thermocouples:: an emf-temperature reference function for the range 0 °C to 1500 °C [J].
Burns, GW ;
Ripple, DC ;
Battuello, M .
METROLOGIA, 1998, 35 (05) :761-780
[4]   A cobalt-carbon eutectic fixed point for the calibration of contact thermometers at temperatures above 1100°C [J].
Edler, F ;
Baratto, AC .
METROLOGIA, 2005, 42 (04) :201-207
[5]  
EDLER F, 1998, P EUROMET WORKSH TEM, P85
[6]   An investigation of palladium oxidation in the platinum/palladium thermocouple system [J].
Hill, KD .
METROLOGIA, 2002, 39 (01) :51-58
[7]   Features of Co/C and Ni/C eutectic transitions for use in thermocouple thermometry [J].
Kim, YG ;
Yang, I ;
Kwon, SY ;
Gam, KS .
METROLOGIA, 2006, 43 (01) :67-70
[8]  
MORICE R, 2004, P TEMPMEKO 2004 9 IN, P847
[9]  
Ogura H, 2003, AIP CONF PROC, V684, P485, DOI 10.1063/1.1627173
[10]  
OGURA H, 2004, P TEMPMEKO 2004 9 IN, P459