Occlusion of Fontan fenestrations using Amplatzer septal occluder

被引:8
作者
Al-Hay, Amira A. A. H. [1 ]
Abu Shaban, Lulu [1 ]
Al-Qbandi, Mustafa A. [1 ]
Al Anbaei, Muath [2 ]
机构
[1] Chest Dis Hosp, Dept Pediat Cardiol, Salmia 22043, Kuwait
[2] Chest Dis Hosp, Dept Adult Cardiol, Adult Congenital Heart Dis Unit, Salmia 22043, Kuwait
关键词
Complex congenital heart diseases; Single ventricle; Total cavopulmonary connections (TCPC); Amplatzer septal occluder; Amplatzer duct occluder; Fenestration closure; TRANSCATHETER COIL OCCLUSION; RESIDUAL INTERATRIAL COMMUNICATIONS; EXTRACARDIAC FONTAN; CATHETER CLOSURE; DUCT OCCLUDER; OPERATION;
D O I
10.1007/s10554-010-9694-0
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The objective of the study is to assess efficacy and safety of the Amplatzer septal device for percutaneous occlusion of Fontan fenestration. Fenestration improves postoperative outcomes in children undergoing Fontan repair. Many of these fenestrations close spontaneously; persistent fenestrations ultimately cause desaturation and can be a potential cause of paradoxical embolism. This is a retrospective review of 26 consecutive patients who underwent transcatheter closure of Fontan fenestration in a tertiary cardiac center in Kuwait. After assessment of hemodynamic suitability for fenestration closure, appropriate balloon sizing of fenestration was obtained. The median age was 9.2 years (range = 1.5-18.3 years). Occlusion was accomplished using a 4-7 ml single Amplatzer septal device in 12 patients and 8-13 ml devices in the other 13 patients; 2 devices were deployed in one patient. Median fluoroscopic and procedure times were 22 and 143.5 min, respectively. One immediate complication was the embolization and successful retrieval of the Amplatzer duct occluder. The fenestration was subsequently occluded by Amplatzer septal device. A 100% occlusion rate of fenestration was achieved in both lateral tunnel and extracardiac conduit types of Fontans. Mean (SD) arterial oxygen saturation increased from 84.7% (4.7) to 95.2% (1.7, P < 0.001) 10 min after deployment of the occluder; Fontan mean circuit pressure had a minimal acceptable increment from 12.9 mmHg (3.2) to 14.6 mmHg (2.6, P < 0.001). No complications or device failures were seen during follow up. The Amplatzer septal occluder device is effective in closing both Fontan fenestrations. Longer single or dual anti-platelet therapy may be considered to prevent thrombotic events. Long-term outpatients follow up with transthoracic echocardiographic monitoring for systemic venous congestion or thrombosis is warranted.
引用
收藏
页码:483 / 490
页数:8
相关论文
共 24 条
  • [1] Univentricular repair: Is routine fenestration justified?
    Airan, B
    Sharma, R
    Choudhary, SK
    Mohanty, SR
    Bhan, A
    Chowdhari, UK
    Juneja, R
    Kothari, SS
    Saxena, A
    Venugopal, P
    [J]. ANNALS OF THORACIC SURGERY, 2000, 69 (06) : 1900 - 1906
  • [2] Boudjemline Y, 2005, ARCH MAL COEUR VAISS, V98, P449
  • [3] Clipped tube fenestration after extracardiac Fontan allows for simple transcatheter coil occlusion
    Bradley, TJ
    Human, DG
    Culham, JAG
    Duncan, WJ
    Patterson, MWH
    LeBlanc, JG
    Sett, SS
    [J]. ANNALS OF THORACIC SURGERY, 2003, 76 (06) : 1923 - 1928
  • [4] BRIDGES ND, 1998, SEMIN THORAC CARDIOV, V1, P9
  • [5] CHEUNG YF, 2005, HONG KONG J PAEDIATR, V10, P265
  • [6] Ethical authorship and publishing
    Coats, Andrew J. S.
    [J]. INTERNATIONAL JOURNAL OF CARDIOLOGY, 2009, 131 (02) : 149 - 150
  • [7] Transcatheter coil occlusion of residual interatrial communications after Fontan procedure
    Gamillscheg, A
    Beitzke, A
    Stein, JI
    Rupitz, M
    Zobel, G
    Rigler, B
    [J]. HEART, 1998, 80 (01) : 49 - 53
  • [8] Interventional closure of residual interatrial communications after modified Fontan operations
    Gamillscheg, A
    Beitzke, A
    Stein, JI
    Zobel, G
    Rödl, S
    Zartner, P
    [J]. ZEITSCHRIFT FUR KARDIOLOGIE, 2002, 91 (04): : 304 - 311
  • [9] Clinical outcome of fenestrated Fontan patients after closure - The first 10 years
    Goff, DA
    Blume, ED
    Gauvreau, K
    Mayer, JE
    Lock, JE
    Jenkins, KJ
    [J]. CIRCULATION, 2000, 102 (17) : 2094 - 2099
  • [10] Transesophageal echocardiographic results of catheter closure of atrial septal defect in children and adults using the Amplatzer device
    Hijazi, ZM
    Cao, QL
    Patel, HT
    Rhodes, J
    Hanlon, KM
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2000, 85 (11) : 1387 - +