Preliminary analysis of ultrasound elastography imaging-based thermometry on non-perfused ex vivo swine liver

被引:8
作者
Giurazza, Francesco [1 ]
Massaroni, Carlo [2 ]
Silvestri, Sergio [2 ]
Zobel, Bruno Beomonte [3 ]
Schena, Emiliano [2 ]
机构
[1] Cardarelli Hosp, Intervent Radiol Dept, Via Cardarelli 9, I-80100 Naples, Italy
[2] Univ Campus Biomed Roma, Measurement & Biomed Instrumentat Lab, Via A Del Portillo 200, I-00198 Rome, Italy
[3] Univ Campus Biomed Roma, Radiol Dept, Via A Del Portillo 200, I-00198 Rome, Italy
关键词
Thermometry; Ablation; Liver; Elastography; Laser; CT-BASED THERMOMETRY; RADIOFREQUENCY ABLATION; LASER-ABLATION; SHEAR MODULUS; THERMOTHERAPY; FEASIBILITY;
D O I
10.1007/s40477-019-00407-z
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Aims Real-time monitoring of tissue temperature during percutaneous tumor ablation improves treatment efficacy, leading clinicians in adjustment of treatment settings. This study aims at assessing feasibility of ultrasound thermometry during laser ablation of biological tissue using a specific ultrasound imaging techniques based on elastography acoustic radiation force impulse (ARFI). Methods ARFI uses high-intensity focused ultrasound pulses to generate 'radiation force' in tissue; this provokes tissue displacements trackable using correlation-based ultrasound methods: the sensitivity of shear waves velocity is able to detect temperature changes. Experiments were carried out using a Nd:YAG laser (power: 5 W) in three non-perfused ex vivo pig livers. In each organ, a thermocouple was placed close to the applicator tip (distance range 1.5-2.5 cm) used to record a reference temperature. Positioning of laser applicator and thermocouple was eco-guided. The organ was scanned by an echography system equipped with ARFI; propagation velocity was measured in a region of interest of 1 x 0.5 cm located close to thermocouple, to investigate influence of tissue temperature on shear waves velocity. Results Shear wave velocity has a very low sensitivity to temperature up to 55-60 degrees C, and in all cases, velocity is < 5 m s(-1); for temperature > 55-60 degrees C, velocity shows a steep increment. The system measures a value "over limit", meaning a velocity > 5 m s(-1). Conclusions Ultrasound thermometry during laser ablation of biological tissue based on elastography shows an abrupt output change at temperatures > 55-60 degrees C. This issue can have a relevant clinical impact, considering tumor necrosis when temperature crosses 55 degrees C to define the boundary of damaged volume.
引用
收藏
页码:69 / 75
页数:7
相关论文
共 28 条
[1]   Principles of and Advances in Percutaneous Ablation [J].
Ahmed, Muneeb ;
Brace, Christopher L. ;
Lee, Fred T., Jr. ;
Goldberg, S. Nahum .
RADIOLOGY, 2011, 258 (02) :351-369
[2]  
[Anonymous], 2008, EVALUATION MEASUREME, DOI DOI 10.1016/0263-2241(95)00017-F
[3]   Monitoring of Thermal Therapy Based on Shear Modulus Changes: II. Shear Wave Imaging of Thermal Lesions [J].
Arnal, Bastien ;
Pernot, Mathieu ;
Tanter, Mickael .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2011, 58 (08) :1603-1611
[4]   FEASIBILITY STUDY OF ULTRASONIC COMPUTED TOMOGRAPHY-GUIDED HIGH-INTENSITY FOCUSED ULTRASOUND [J].
Azhari, Haim .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2012, 38 (04) :619-625
[5]   Temporal analysis of tissue displacement induced by a transient ultrasound radiation force [J].
Callé, S ;
Remenieras, JP ;
Matar, OB ;
Hachemi, ME ;
Patat, F .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2005, 118 (05) :2829-2840
[6]   Endoscopic ultrasound-guided Nd:YAG laser ablation of recurrent pancreatic neuroendocrine tumor: a promising revolution? [J].
Di Matteo, Francesco ;
Picconi, Francesca ;
Martino, Margareth ;
Pandolfi, Monica ;
Pacella, Claudio Maurizio ;
Schena, Emiliano ;
Costamagna, Guido .
ENDOSCOPY, 2014, 46 :E380-E381
[7]   CT-based thermometry: An overview [J].
Fani, F. ;
Schena, E. ;
Saccomandi, P. ;
Silvestri, S. .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2014, 30 (04) :219-227
[8]   Point quantification elastography in the evaluation of liver elasticity in healthy volunteers: a reliability study based on operator expertise [J].
Felicani C. ;
De Molo C. ;
Stefanescu H. ;
Conti F. ;
Mazzotta E. ;
Gabusi V. ;
Nardi E. ;
Morselli-Labate A.M. ;
Andreone P. ;
Serra C. .
Journal of Ultrasound, 2018, 21 (2) :89-98
[9]   Non-invasive thermometry for monitoring hepatic radiofrequency ablation [J].
Frich, L .
MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2006, 15 (01) :18-25
[10]   Renal cell carcinoma: Clinical experience and technical success with radio-frequency ablation of 42 tumors [J].
Gervais, DA ;
McGovern, F ;
Arellano, RS ;
McDougal, WS ;
Mueller, PR .
RADIOLOGY, 2003, 226 (02) :417-424