Comparison of HASTE with multiple signal averaging versus conventional turbo spin echo sequence: a new option for T2-weighted MRI of the female pelvis

被引:13
|
作者
Tsuboyama, Takahiro [1 ]
Takei, Oki [1 ]
Okada, Atsuhiko [1 ]
Honda, Toru [1 ]
Kuriyama, Keiko [1 ]
机构
[1] Natl Hosp Org Osaka Natl Hosp, Dept Radiol, Chuo Ku, 2-1-14 Hoenzaka, Osaka 5400006, Japan
关键词
Magnetic resonance imaging; Spin echo imaging; Pelvis; Female; IMAGE QUALITY; INCREASED SPEED; ACQUISITION; BLADE; 3D;
D O I
10.1007/s00330-020-06686-z
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives This study was conducted in order to evaluate the feasibility of a newly developed half-Fourier acquisition single-shot turbo spin echo (HASTE) sequence for T2-weighted MRI of the female pelvis, in which scanning and post-processing filtering techniques were optimized to minimize blurring and the resultant reduction in signal-to-noise ratio (SNR) was compensated by signal averaging of multiple excitations (multi-NEX), and to compare its image quality with that of a conventional turbo spin echo (TSE) sequence. Methods Fifty females with gynecological disease were prospectively enrolled. They underwent pelvic MRI including sagittal T2-weighted imaging obtained with multi-NEX HASTE (mHASTE) and TSE with acquisition times of 1 m 9 s and 3 m 26 s, respectively. At the prospective reading, mHASTE was reviewed first, and any changes in the image interpretation detected during the subsequent TSE reading were recorded. The SNR and tissue contrast were quantitatively measured. Two radiologists independently assessed tissue conspicuity and overall image quality considering noise, artifacts, and blurring, and visual grading characteristics (VGC) analysis was performed to compare the two sequences. Results Adding TSE to mHASTE did not change image interpretations in any case. mHASTE produced an equivalent SNR and significantly higher contrast between the myometrium and junctional zone (p < 0.05) compared with TSE. The area under the VGC curve (AUC(VGC)) showed that the conspicuity of the uterus, ovaries, and lesions and the overall image quality were significantly higher on mHASTE than on TSE for both readers (AUC(VGC), 0.740-0.880; p < 0.01). Conclusion mHASTE is a feasible option for T2-weighted MRI of the female pelvis, as it allows faster and higher-quality image acquisition than conventional TSE.
引用
收藏
页码:3245 / 3253
页数:9
相关论文
共 50 条
  • [21] MRI of the female pelvis at 3T compared to 1.5T: Evaluation on high-resolution T2-weighted and HASTE images
    Kataoka, Masako
    Kido, Aki
    Koyama, Takashi
    Isoda, Hiroyoshi
    Umeoka, Shigeaki
    Tamai, Ken
    Nakamoto, Yuji
    Maetani, Yoji
    Morisawa, Nobuko
    Saga, Tsuneo
    Togashi, Kaori
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2007, 25 (03) : 527 - 534
  • [22] REDUCTION OF MRT TIMING FOR EXAMINATION OF THE FEMALE PELVIS USING T2-WEIGHTED FAST SPIN-ECHO (FSE) TECHNIQUES - QUANTITATIVE AND QUALITATIVE COMPARISON OF CONVENTIONAL SPIN-ECHO (SE) AND FAST SPIN-ECHO (FSE)
    HALDEMANN, RC
    KRESTIN, GP
    DUEWELL, S
    MARINCEK, B
    VONSCHULTHESS, GK
    SAADIELMANDJRA, M
    FUCHS, WA
    FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN, 1993, 158 (02): : 127 - 132
  • [23] Comparison of single-echo T2-weighted fast spin-echo and conventional spin-echo MR sequences - In the detection of brain lesions of multiple sclerosis
    Sparacia, G
    Angileri, T
    Banco, A
    Miraglia, R
    Nicastro, T
    Lagalla, R
    RIVISTA DI NEURORADIOLOGIA, 1998, 11 (04): : 471 - 476
  • [24] Assessment of the distal anastomosis of coronary artery bypass grafts with a 2D T2-weighted turbo spin echo sequence and correlation to conventional coronary angiography
    Wittlinger, T
    Martinovic, I
    Moritz, A
    Kalden, P
    Kreitner, KF
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2006, 109 (02) : 219 - 225
  • [25] Implementation of a sagittal T2-weighted DIXON turbo spin-echo sequence may shorten MRI acquisitions in the emergency setting of suspected spinal bleeding
    Sollmann, Nico
    Ruether, Charlotte
    Schoen, Simon
    Zimmer, Claus
    Baum, Thomas
    Kirschke, Jan S.
    EUROPEAN RADIOLOGY EXPERIMENTAL, 2021, 5 (01)
  • [26] Implementation of a sagittal T2-weighted DIXON turbo spin-echo sequence may shorten MRI acquisitions in the emergency setting of suspected spinal bleeding
    Nico Sollmann
    Charlotte Rüther
    Simon Schön
    Claus Zimmer
    Thomas Baum
    Jan S. Kirschke
    European Radiology Experimental, 5
  • [27] MRI of acute cerebral infarction: A comparison of FLAIR and T2-weighted fast spin-echo imaging
    Noguchi, K
    Ogawa, T
    Inugami, A
    Fujita, H
    Hatazawa, J
    Shimosegawa, E
    Okudera, T
    Uemura, K
    Seto, H
    NEURORADIOLOGY, 1997, 39 (06) : 406 - 410
  • [28] Comparison of T2-weighted fast spin-echo and spin-echo MR sequences in the detection of brain lesions of multiple sclerosis
    Sparacia, G
    Angileri, T
    Accardi, M
    Pappalardo, S
    Sarno, C
    DeMaria, M
    RIVISTA DI NEURORADIOLOGIA, 1997, 10 : 233 - 235
  • [29] Comparative study of image quality between axial T2-weighted BLADE and turbo spin-echo MRI of the upper abdomen on 3.0 T
    Zhang, Lin
    Tian, ChunMei
    Wang, PeiYuan
    Chen, Liang
    Mao, Xijin
    Wang, ShanShan
    Wang, Xu
    Dong, JingMin
    Wang, Bin
    JAPANESE JOURNAL OF RADIOLOGY, 2015, 33 (09) : 585 - 590
  • [30] Comparative study of image quality between axial T2-weighted BLADE and turbo spin-echo MRI of the upper abdomen on 3.0 T
    Lin Zhang
    ChunMei Tian
    PeiYuan Wang
    Liang Chen
    XiJin Mao
    ShanShan Wang
    Xu Wang
    JingMin Dong
    Bin Wang
    Japanese Journal of Radiology, 2015, 33 : 585 - 590