Determination of tobramycin in serum using liquid chromatography-tandem mass spectrometry and comparison with a fluorescence polarisation assay

被引:59
作者
Keevil, BG [1 ]
Lockhart, SJ [1 ]
Cooper, DR [1 ]
机构
[1] Wythenshawe Hosp, S Manchester Univ Hosp NHS Trust, Dept Clin Biochem, Manchester M23 9LT, Lancs, England
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2003年 / 794卷 / 02期
关键词
tobramycin;
D O I
10.1016/S1570-0232(03)00492-6
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We have developed a tandem mass spectrometry (LC-MS-MS) method for measuring tobramycin concentrations in serum samples and have compared it with a fluorescence polarisation immunoassay. After protein precipitation with acetonitrile supernatant was injected into the LC-MS-MS system. A C-18 cartridge (4 x 2 mm) was eluted with a step gradient of 20-100% methanol containing HFBA. The retention times were, tobramycin 1.05 min and sisomycin 1.05 min. The MRM transitions were: m/z 467.8 > 163 (tobramycin) and m/z 447.8 > 160 (sisomycin). The limit of quantification was 0.15 mg/l and the assay was linear up to 50 mg/l. Assay precision was <6% within and between batch. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:329 / 335
页数:7
相关论文
共 19 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]  
Cherlet M, 2000, J MASS SPECTROM, V35, P1342, DOI 10.1002/1096-9888(200011)35:11<1342::AID-JMS71>3.0.CO
[3]  
2-Y
[4]   Effect of variability of plasma interferences on the accuracy of drug immunoassays [J].
Ezan, E ;
Emmanuel, A ;
Valente, D ;
Grognet, JM .
THERAPEUTIC DRUG MONITORING, 1997, 19 (02) :212-218
[5]   Trace analysis of tobramycin in human plasma by derivatization and high-performance liquid chromatography with ultraviolet detection [J].
Feng, CH ;
Lin, SJ ;
Wu, HL ;
Chen, SH .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2002, 780 (02) :349-354
[6]  
Heller DN, 2000, J MASS SPECTROM, V35, P39, DOI 10.1002/(SICI)1096-9888(200001)35:1<39::AID-JMS911>3.0.CO
[7]  
2-Y
[8]  
Isoherranen N, 2000, CLIN CHEM, V46, P837
[9]   ENHANCEMENT OF DETECTION SENSITIVITY AND CLEANUP SELECTIVITY FOR TOBRAMYCIN THROUGH PRECOLUMN DERIVATIZATION [J].
LAI, F ;
SHEEHAN, T .
JOURNAL OF CHROMATOGRAPHY, 1992, 609 (1-2) :173-179
[10]  
MAITRA SK, 1979, CLIN CHEM, V25, P1361