Optic Nerve Sheath Diameter Ultrasound: Optic Nerve Growth Curve and Its Application to Detect Intracranial Hypertension in Children

被引:13
作者
Fontanel, Liviana [1 ]
Pensiero, Stefano [2 ]
Ronfani, Luca [3 ]
Rosolen, Valentina [3 ]
Barbi, Egidio [1 ,4 ]
机构
[1] Univ Trieste, Dept Med Surg & Hlth Sci, Trieste, Italy
[2] Ist Ricovero & Cura Carattere Sci Burlo Garofolo, Inst Maternal & Child Hlth, Dept Ophthalmol, Trieste, Italy
[3] Ist Ricovero & Cura Carattere Sci Burlo Garofolo, Inst Maternal & Child Hlth, Dept Clin Epidemiol & Publ Hlth, Trieste, Italy
[4] Ist Ricovero & Cura Carattere Sci Burlo Garofolo, Inst Maternal & Child Hlth, Dept Pediat, Trieste, Italy
关键词
PRESSURE; ULTRASONOGRAPHY; SONOGRAPHY; DIAGNOSIS;
D O I
10.1016/j.ajo.2019.07.014
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: First, to create an optic nerve growth curve from normal optic nerve sheath diameter (ONSD) values measured by using B-scan ultrasonography in subjects 0-18 years of age. Second, to identify age-appropriate cutoff values of ONSD to be used in the diagnosis of intracranial hypertension (IHT). DESIGN: Prospective cross-sectional study. METHODS: B-scan ocular ultrasonography was performed on both eyes of 215 subjects 0.18 years of age, divided into 3 groups: 165 healthy children, 29 children with IHT (all > 4 years of age), and 21 children with optic disc drusen (ODD). RESULTS: There were no statistically significant differences in between the ONSDs of healthy children and those in subjects with ODD. An optic nerve growth curve was created by using ONSDs measured in healthy subjects 0-18 years of age, using the equation: ONSD = In [33.15] (-0.18 x In [children's age]). The curve showed a progressive increase of ONSD up to 10 years of age, and it remained constant until the age of 18. For this reason, 2 different cutoff values were calculated for age groups 4-10 and 11-18. Values were 4.10 mm and 4.4 mm, respectively, with a 100.0% sensitivity and a specificity ranging from 83.9% to 98.8%. A sensitivity of 28.6% was reached for the population of subjects 4-18 years of age with a threshold value of 5 mm, as used in published reports. CONCLUSIONS: The ONSD continued to enlarge gradually until the age of 10. Therefore, 2 different cutoff values for the age groups 4-10 and 11-18 were calculated, considering the ONSDs of subjects 11-18 years of age overlapping with those of adults. No patients with IHT <4 years old were found. Further studies are needed to evaluate the correct cutoff values for these ages. (C) 2019 Elsevier Inc. All rights reserved.
引用
收藏
页码:421 / 428
页数:8
相关论文
共 26 条
[1]   CSF opening pressure in children with optic nerve head edema [J].
Avery, R. A. ;
Licht, D. J. ;
Shah, S. S. ;
Huh, J. W. ;
Seiden, J. A. ;
Boswinkel, J. ;
Ruppe, M. D. ;
Mistry, R. D. ;
Liu, G. T. .
NEUROLOGY, 2011, 76 (19) :1658-1661
[2]   Pediatric Intracranial Hypertension: a Current Literature Review [J].
Aylward, Shawn C. ;
Way, Amanda L. .
CURRENT PAIN AND HEADACHE REPORTS, 2018, 22 (02)
[3]   Reproducibility and accuracy of optic nerve sheath diameter assessment using ultrasound compared to magnetic resonance imaging [J].
Baeuerle, Jochen ;
Schuchardt, Florian ;
Schroeder, Laure ;
Egger, Karl ;
Weigel, Matthias ;
Harloff, Andreas .
BMC NEUROLOGY, 2013, 13
[4]   Transorbital optic nerve sheath ultrasonography in normal children [J].
Ballantyne, J ;
Hollman, AS ;
Hamilton, R ;
Bradnam, MS ;
Carachi, R ;
Young, DG ;
Dutton, GN .
CLINICAL RADIOLOGY, 1999, 54 (11) :740-742
[5]   Idiopathic Intracranial Hypertension in Children and Adolescents: An Update [J].
Cleves-Bayon, Catalina .
HEADACHE, 2018, 58 (03) :485-493
[6]   Pseudotumor cerebri as an important differential diagnosis of papilledema in children [J].
Distelmaier, F ;
Sengler, U ;
Messing-Juenger, M ;
Assmann, B ;
Mayatepek, E ;
Rosenbaum, T .
BRAIN & DEVELOPMENT, 2006, 28 (03) :190-195
[7]   Individual patient data systematic review and meta-analysis of optic nerve sheath diameter ultrasonography for detecting raised intracranial pressure: protocol of the ONSD research group. [J].
Dubourg J. ;
Messerer M. ;
Karakitsos D. ;
Rajajee V. ;
Antonsen E. ;
Javouhey E. ;
Cammarata A. ;
Cotton M. ;
Daniel R.T. ;
Denaro C. ;
Douzinas E. ;
Dubost C. ;
Berhouma M. ;
Kassai B. ;
Rabilloud M. ;
Gullo A. ;
Hamlat A. ;
Kouraklis G. ;
Mannanici G. ;
Marill K. ;
Merceron S. ;
Poularas J. ;
Ristagno G. ;
Noble V. ;
Shah S. ;
Kimberly H. ;
Cammarata G. ;
Moretti R. ;
Geeraerts T. .
Systematic Reviews, 2 (1) :62
[8]   Ultrasound assessment of optic nerve sheath diameter in healthy volunteers [J].
Goeres, Patrick ;
Zeiler, Frederick A. ;
Unger, Bertram ;
Karakitsos, Dimitrios ;
Gillman, Lawrence M. .
JOURNAL OF CRITICAL CARE, 2016, 31 (01) :168-171
[9]   Fundamentals of transorbital sonographic evaluation of optic nerve sheath expansion under intracranial hypertension .2. Patient study [J].
Helmke, K ;
Hansen, HC .
PEDIATRIC RADIOLOGY, 1996, 26 (10) :706-710
[10]   Hierarchical Bayesian modeling of spatioTtemporal patterns of scrub typhus incidence for 2009-2013 in South Korea [J].
Kim, Seungwon ;
Kim, Youngho .
APPLIED GEOGRAPHY, 2018, 100 :1-11