Performance of ion pairing chromatography and hydrophilic interaction liquid chromatography coupled to charged aerosol detection for the analysis of underivatized amino acids

被引:6
作者
Pawellek, Ruben [1 ]
Holzgrabe, Ulrike [1 ]
机构
[1] Univ Wurzburg, Inst Pharm & Food Chem, D-97074 Wurzburg, Germany
关键词
Branched-chain amino acid (BCAA); Charged aerosol detector (CAD); High-performance liquid chromatography (HPLC); Hydrophilic interaction chromatography (HILIC); Ion pair chromatography (IPC); Response surface methodology (RSM); HPLC; CAD;
D O I
10.1016/j.chroma.2021.462613
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The charged aerosol detector (CAD) is frequently employed in liquid chromatography for the analysis of small polar and ionizable compounds such as amino acids and amino sugars, which provide a weak chro-mophore only. Separation of these compounds is achieved by means of ion pair chromatography (IPC), and, more recently, hydrophilic interaction chromatography (HILIC) techniques. However, as the CAD's re-sponse is highly dependent on the mobile phase composition, the substantial differences in the mobile phase composition of IPC and HILIC have a distinct impact on the detector's performance. This study was aimed at systematically comparing the performance of IPC and HILIC when coupled to the CAD. Therefore, the separation techniques characterized by their specific mobile phase compositions were evaluated for their influence on the CAD response and the signal-to-noise ratio (S/N) of the amino acids L-alanine, L-leucine, and L-phenylalanine applying the response surface methodology (RSM). The RSM results derived from flow injection analysis (FIA) indicated that the CAD response and thus the obtainable S/N are sig-nificantly higher in HILIC compared to IPC where the S/N decreased with the chain length of the applied ion-pairing reagent. In addition, an IPC and a HILIC method, respectively, were developed for the impurity profiling of the branched-chain amino acids (BCAAs) L-leucine, L-isoleucine, and L-valine. The beneficial effects of the HILIC conditions on the S/N observed under FIA conditions were partly offset by moderate column bleed effects when using an amide functionalized column, which facilitates the separation in the HILIC method. Satisfactory LOQs (3-10 ng on column) were obtained with both methods; however, the HILIC method was found to be slightly superior in terms of sensitivity and separation efficiency. (c) 2021 Elsevier B.V. All rights reserved.
引用
收藏
页数:15
相关论文
共 44 条
[11]  
Gamache P. H, 2017, Charged aerosol detection for liquid chromatography and related reparation techniques, V1st
[12]  
Gamache PH, 2005, LC GC EUR, V18, P345
[13]   Composition of structural carbohydrates in biomass: Precision of a liquid chromatography method using a neutral detergent extraction and a charged aerosol detector [J].
Godin, Bruno ;
Agneessens, Richard ;
Gerin, Patrick A. ;
Delcarte, Jerome .
TALANTA, 2011, 85 (04) :2014-2026
[14]   Universal response in liquid chromatography using charged aerosol detection [J].
Gorecki, Tadeusz ;
Lynen, Frederic ;
Szucs, Roman ;
Sandra, Pat .
ANALYTICAL CHEMISTRY, 2006, 78 (09) :3186-3192
[15]   Control of impurities in L-aspartic acid and L-alanine by high-performance liquid chromatography coupled with a corona charged aerosol detector [J].
Holzgrabe, Ulrike ;
Nap, Cees-Jan ;
Almeling, Stefan .
JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (03) :294-301
[16]   Determination of inorganic pharmaceutical counterions using hydrophilic interaction chromatography coupled with a Corona® CAD detector [J].
Huang, Z. ;
Richards, M. A. ;
Zha, Y. ;
Francis, R. ;
Lozano, R. ;
Ruan, J. .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2009, 50 (05) :809-814
[17]   Investigation of polar organic solvents compatible with Corona Charged Aerosol Detection and their use for the determination of sugars by hydrophilic interaction liquid chromatography [J].
Hutchinson, Joseph P. ;
Remenyi, Tomas ;
Nesterenko, Pavel ;
Farrell, William ;
Groeber, Elizabeth ;
Szucs, Roman ;
Dicinoski, Greg ;
Haddad, Paul R. .
ANALYTICA CHIMICA ACTA, 2012, 750 :199-206
[18]   Comparison of the response of four aerosol detectors used with ultra high pressure liquid chromatography [J].
Hutchinson, Joseph P. ;
Li, Jianfeng ;
Farrell, William ;
Groeber, Elizabeth ;
Szucs, Roman ;
Dicinoski, Greg ;
Haddad, Paul R. .
JOURNAL OF CHROMATOGRAPHY A, 2011, 1218 (12) :1646-1655
[19]  
International Council for HarmonizationGuideline Q3A (R2), 2006, GUID Q3A R2 IMP NEW
[20]  
International Council of Harmonisation, 2005, GUID Q2 R1 VAL AN PR