Measurement of pH and dissolved oxygen within cell culture media using a hydrogel microarray sensor

被引:69
作者
Lee, Seungjoon [2 ,3 ,4 ]
They, Bennett L. [2 ,3 ,4 ]
Cote, Gerard L. [2 ,3 ,4 ]
Pishko, Michael V. [1 ]
机构
[1] Texas A&M Univ, Dept Biomed Engn, College Stn, TX 77843 USA
[2] Penn State Univ, Dept Chem Engn, University Pk, PA 16802 USA
[3] Penn State Univ, Dept Chem, University Pk, PA 16802 USA
[4] Penn State Univ, Dept Mat Sci & Technol, University Pk, PA 16802 USA
基金
美国国家航空航天局;
关键词
poly(ethylene glycol); fluorescent sensing; cell culture;
D O I
10.1016/j.snb.2007.06.027
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The sensing of analytes in cell culture media without the introduction of exogenous reagents proves difficult in today's cell culturing systems. In this paper, a completely embeddable microarray sensor that has been developed along with a compact fluorescent imaging system to sense oxygen and pH non-invasively is described. The compact detection system consists of an ultra-bright blue diode source, coupling optics, interference filters and a compact moisture resistant CCD camera. The microarray sensor was created by photoreaction injection molding and contains two separate sensing elements. The system was constructed and the array tested using simple spiked buffer solutions and complex cell media. In addition, the system was coupled to a bioreactor and tested over 2 weeks. The standard error of prediction during a bioprocess for both oxygen and pH detection was 0.75% and 0.092, respectively. The time response of the sensor in the designed flow-through arrangement was less than 90s for both pH and oxygen, well within that needed for cell culture monitoring. Overall, this sensing approach has promise as a lab-on-a-chip type sensor for complete sensing of the culture environment and eventually may be used in a feedback system that allows monitoring and control without human intervention. (C) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:388 / 398
页数:11
相关论文
共 39 条
  • [11] Lakowicz J. R., 1999, PRINCIPLES FLUORESCE
  • [12] OPTICAL CHARACTERISTICS OF A RUTHENIUM(II) COMPLEX IMMOBILIZED IN A SILICONE-RUBBER FILM FOR OXYGEN MEASUREMENT
    LI, XM
    RUAN, FC
    WONG, KY
    [J]. ANALYST, 1993, 118 (03) : 289 - 292
  • [13] LED-BASED FIBER OPTIC OXYGEN SENSOR USING SOL-GEL COATING
    MACCRAITH, BD
    OKEEFFE, G
    MCDONAGH, C
    MCEVOY, AK
    [J]. ELECTRONICS LETTERS, 1994, 30 (11) : 888 - 889
  • [14] Multi-analyte optical chemical sensor employing a plastic substrate
    Malins, C
    Niggemann, M
    MacCraith, BD
    [J]. MEASUREMENT SCIENCE AND TECHNOLOGY, 2000, 11 (08) : 1105 - 1110
  • [15] Fiber-optic oxygen sensor based on the fluorescence quenching of tris (5-acrylamido, 1,10 phenanthroline) ruthenium chloride
    McNamara, KP
    Li, XP
    Stull, AD
    Rosenzweig, Z
    [J]. ANALYTICA CHIMICA ACTA, 1998, 361 (1-2) : 73 - 83
  • [16] Release of protein from highly cross-linked hydrogels of poly(ethylene glycol) diacrylate fabricated by UV polymerization
    Mellott, MB
    Searcy, K
    Pishko, MV
    [J]. BIOMATERIALS, 2001, 22 (09) : 929 - 941
  • [17] Fabrication of polymer organized thin films containing ruthenium complexes
    Miyashita, T
    Chen, JF
    Yuasa, M
    Mitsuishi, M
    [J]. POLYMER JOURNAL, 1999, 31 (11) : 1121 - 1126
  • [18] Fluorescence ratiometric pH sensor prepared from covalently immobilized porphyrin and benzothioxanthene
    Niu, CG
    Gui, XQ
    Zeng, GM
    Guan, AL
    Gao, PF
    Qin, PZ
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2005, 383 (02) : 349 - 357
  • [19] Oxygen sensor based on the fluorescence quenching of a ruthenium complex immobilized in a biocompatible poly(ethylene glycol) hydrogel
    O'Neal, DP
    Meledeo, MA
    Davis, JR
    Ibey, BL
    Gant, VA
    Pishko, MV
    Coté, GL
    [J]. IEEE SENSORS JOURNAL, 2004, 4 (06) : 728 - 734
  • [20] OPTICAL-FIBER SENSOR FOR BIOLOGICAL OXYGEN-DEMAND
    PREININGER, C
    KLIMANT, I
    WOLFBEIS, OS