Investigation of strategies to achieve optimal DQE performance from indirect detection, active matrix flat-panel imagers (AMFPIs) through novel pixel amplification architectures

被引:10
作者
Antonuk, LE [1 ]
Li, YX [1 ]
Du, H [1 ]
El-Mohri, Y [1 ]
Zhao, QH [1 ]
Yamamoto, J [1 ]
Sawant, A [1 ]
Wang, Y [1 ]
Su, Z [1 ]
Lu, JP [1 ]
Street, RA [1 ]
Weisfield, R [1 ]
Yao, B [1 ]
机构
[1] Univ Michigan, Dept Radiat Oncol, Ann Arbor, MI 48109 USA
来源
MEDICAL IMAGING 2005: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2 | 2005年 / 5745卷
关键词
active matrix flat-panel imager; in-pixel amplifiers; polycrystalline silicon TFTs; digital x-ray imaging;
D O I
10.1117/12.596786
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The numerous merits of x-ray imagers based on active matrix, flat-panel array technology have led to their introduction in a wide variety of x-ray imaging applications. However, under certain conditions, the performance of direct and indirect detection AMFPIs is significantly limited by the relatively modest ratio of signal to noise provided by conventional systems. While substantial reduction in the additive noise of such systems is difficult, significant enhancement of signal can be achieved through the incorporation of an amplification circuit in each pixel. In addition, innovative photodiode structures can be incorporated into indirect detection designs to enhance optical signal collection efficiency. In this paper, an investigation of these strategies, involving the design, fabrication and performance evaluation of a variety of novel, small area, indirect detection arrays, is described. Each prototype array incorporates innovative features, such as continuous photodiodes and single-stage and dual-stage in-pixel amplifiers, that are designed to provide insight into promising avenues for achieving significant signal-to-noise enhancement. This information will assist in the realization of a next generation of highly-optimized AMFPI pixel architectures whose DQE performance will be limited only by x-ray noise and x-ray converter properties under a very wide range of conditions. In this paper, the design and operation of the present prototype arrays are described and initial performance results are reported. In addition, the benefits of significant improvements to the signal-to-noise properties of AMFPIs are illustrated through cascaded systems calculations of the DQE performance of hypothetical systems.
引用
收藏
页码:18 / 31
页数:14
相关论文
共 40 条
[21]   Flat panel imagers with pixel level amplifiers based on polycrystalline silicon thin-film transistor technology [J].
Lu, JP ;
Van Schuylenbergh, K ;
Ho, J ;
Wang, Y ;
Boyce, JB ;
Street, RA .
APPLIED PHYSICS LETTERS, 2002, 80 (24) :4656-4658
[22]  
LU JP, 2001, MAT RES SOC S P, V685
[23]   Simulated and measured data-line parasitic capacitance of amorphous silicon large-area image sensor arrays [J].
Mulato, M ;
Lu, JP ;
Street, RA .
JOURNAL OF APPLIED PHYSICS, 2001, 89 (01) :638-647
[24]   Digital indirect detection x-ray imagers with microlens focusing: Effects of Fresnel reflections from the microlens layer [J].
Petrick, N ;
Badano, A ;
Chan, HP ;
Sahiner, B ;
Hadjiiski, LM .
MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 :504-510
[25]   A technique to improve the effective fill factor of digital mammographic imagers [J].
Petrick, N ;
Keshavmurthy, S ;
Chan, HP ;
Goodsitt, MM ;
Sahiner, B .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :210-218
[26]   High resolution, high fill factor a-Si:H sensor arrays for medical imaging [J].
Rahn, JT ;
Lemmi, F ;
Weisfield, RL ;
Lujan, R ;
Mei, P ;
Lu, JP ;
Ho, J ;
Ready, SE ;
Apte, RB ;
Nylen, P ;
Boyce, J ;
Street, RA .
MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 1999, 3659 :510-517
[27]   PbO as direct conversion X-ray detector material [J].
Simon, M ;
Ford, RA ;
Franklin, AR ;
Grabowski, SP ;
Menser, B ;
Much, G ;
Nascetti, A ;
Overdick, M ;
Powell, MJ ;
Wiechert, DU .
MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 :188-199
[28]   X-ray imaging using lead iodide as a semiconductor detector [J].
Street, RA ;
Rahn, JT ;
Ready, SE ;
Shah, K ;
Bennett, P ;
Dmitriyev, Y ;
Mei, P ;
Lu, JP ;
Apte, RB ;
Ho, J ;
van Schuylenbergh, K ;
Lemmi, F ;
Boyce, JB ;
Nylen, P .
MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 1999, 3659 :36-47
[29]   New materials and processes for flat panel X-ray detectors [J].
Street, RA ;
Lu, JP ;
Ready, SR .
IEE PROCEEDINGS-CIRCUITS DEVICES AND SYSTEMS, 2003, 150 (04) :250-257
[30]   Comparison of PbI2 and HgI2 for direct detection active matrix x-ray image sensors [J].
Street, RA ;
Ready, SE ;
Van Schuylenbergh, K ;
Ho, J ;
Boyce, JB ;
Nylen, P ;
Shah, K ;
Melekhov, L ;
Hermon, H .
JOURNAL OF APPLIED PHYSICS, 2002, 91 (05) :3345-3355