Improving mass measurement of coronary artery calcification using threshold correction and thin collimation in multi-detector row computed tomography: In vitro experiment

被引:9
作者
Hong, C [1 ]
Pilgram, TK [1 ]
Zhu, F [1 ]
Joe, BN [1 ]
Towler, DA [1 ]
Bae, KT [1 ]
机构
[1] Washington Univ, Sch Med, Mallinckrodt Inst Radiol, St Louis, MO 63110 USA
关键词
coronary vessels; calcification; computed tomography (CT); screening; ELECTRON-BEAM CT; HELICAL CT; SENSITIVITY PROFILES; RADIATION EXPOSURE; SPIRAL CT; CALCIUM; REPRODUCIBILITY; QUANTIFICATION; NOISE; HEART;
D O I
10.1016/S1076-6332(03)00111-9
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives. This in vitro study was designed to improve the accuracy of coronary calcium mass measurement from computed tomography (CT) images by developing threshold-based calcium CT number correction and thin-slice spiral techniques. Materials and Methods. A cardiac CT phantom containing simulated calcified cylinders of known calcium density was scanned with sequential 4 X 2.5-, spiral 16 X 1.5-, and spiral 16 X 0.75-mm collimation on 4- and 16-detector row CT scanners. The images obtained from the spiral scans were reconstructed in various slice widths. The calcified cylinders were imaged and their mean CT number and size were measured at thresholds ranging from 50 to 390 Hounsfield unit. The calcium mass measured was compared with the actual value to determine errors, and threshold-based correction factors were derived to minimize the errors. Results. The minimum amount of measurable calcium in 1-mm cylinder was 0.3 mg at the 16 X 1.5-mm protocol and 0.2 mg at the 16 X 0.75-mm protocol. Compared with 2.5-mm sequential protocol, thin-slice spiral protocols yielded a higher radiation exposure and lower or similar image noise levels. The error in calcium mass after correction was significantly smaller than that in measured mass (P<.0001) and was consistent between the imaging protocols (P=.49). The accuracy of mass measurements was clearly improved by using thin-slice imaging protocols especially in 200-mg/cm(3) calcium density (P<.0001). Conclusion. The accuracy of calcium mass CT measurement can be improved by threshold-based calcium CT number correction and thin collimation spiral techniques.
引用
收藏
页码:969 / 977
页数:9
相关论文
共 21 条
  • [1] QUANTIFICATION OF CORONARY-ARTERY CALCIUM USING ULTRAFAST COMPUTED-TOMOGRAPHY
    AGATSTON, AS
    JANOWITZ, WR
    HILDNER, FJ
    ZUSMER, NR
    VIAMONTE, M
    DETRANO, R
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 15 (04) : 827 - 832
  • [2] Predictive value of electron beam computed tomography of the coronary arteries - 19-month follow-up of 1173 asymptomatic subjects
    Arad, Y
    Spadaro, LA
    Goodman, K
    LledoPerez, A
    Sherman, S
    Lerner, C
    Guerci, AD
    [J]. CIRCULATION, 1996, 93 (11) : 1951 - 1953
  • [3] SMALL LESIONS IN THE HEART IDENTIFIED AT ELECTRON-BEAM CT - CALCIFICATION OR NOISE
    BIELAK, LF
    KAUFMANN, RB
    MOLL, PP
    MCCOLLOUGH, CH
    SCHWARTZ, RS
    SHEEDY, PF
    [J]. RADIOLOGY, 1994, 192 (03) : 631 - 636
  • [4] Measurement of coronary artery calcium with dual-slice helical CT compared with coronary angiography: Evaluation of CT scoring methods, interobserver variations, and reproducibility
    Broderick, LS
    Shemesh, J
    Wilensky, RL
    Eckert, GJ
    Zhou, XH
    Torres, WE
    Balk, MA
    Rogers, WJ
    Conces, DJ
    Kopecky, KK
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1996, 167 (02) : 439 - 444
  • [5] BUCK RC, 1951, AMA ARCH PATHOL, V51, P319
  • [6] Effect of HMG-Coa reductase inhibitors on coronary artery disease as assessed by electron-beam computed tomography
    Callister, TQ
    Raggi, P
    Cooil, B
    Lippolis, NJ
    Russo, DJ
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1998, 339 (27) : 1972 - 1978
  • [7] Coronary calcium - The case for helical computed tomography
    Carr, JJ
    [J]. JOURNAL OF THORACIC IMAGING, 2001, 16 (01) : 16 - 24
  • [8] REPRODUCIBILITY OF THE MEASUREMENT OF CORONARY CALCIUM WITH ULTRAFAST COMPUTED-TOMOGRAPHY
    DEVRIES, S
    WOLFKIEL, C
    SHAH, V
    CHOMKA, E
    RICH, S
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1995, 75 (14) : 973 - 975
  • [9] Spiral interpolation algorithms for multislice spiral CT - Part II: Measurement and evaluation of slice sensitivity profiles and noise at a clinical multislice system
    Fuchs, T
    Krause, J
    Schaller, S
    Flohr, T
    Kalender, WA
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2000, 19 (09) : 835 - 847
  • [10] Coronary artery calcium: Accuracy and reproducibility of measurements with multi-detector row CT - Assessment of effects of different thresholds and quantification methods
    Hong, C
    Bae, KT
    Pilgram, TK
    [J]. RADIOLOGY, 2003, 227 (03) : 795 - 801