Evaluation of the Novel Methotrexate Architect Chemiluminescent Immunoassay: Clinical Impact on Pharmacokinetic Monitoring

被引:7
作者
Aumente, Maria D. [1 ]
Lopez-Santamaria, Julia [1 ]
Concepcion Donoso-Rengifo, Maria [1 ]
Reyes-Torres, Inmaculada [1 ]
Montejano Hervas, Pablo [1 ]
机构
[1] Hosp Univ Reina Sofia, Unidad Farmacocinet, Serv Farm, Cordoba, Spain
关键词
methotrexate; immunoassay; method validation; therapeutic drug monitoring; leucovorin rescue; HIGH-DOSE METHOTREXATE; ACUTE LYMPHOBLASTIC-LEUKEMIA; FLUORESCENCE POLARIZATION IMMUNOASSAY; POPULATION PHARMACOKINETICS; MASS-SPECTROMETRY; CHILDREN; OSTEOSARCOMA; THERAPY;
D O I
10.1097/FTD.0000000000000434
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Fluorescence polarization immunoassay (FPIA) has probably been the most widely used technique for the determination of methotrexate (MTX) concentrations in clinical laboratories. After its replacement by a novel architect chemiluminescent microparticle immunoassay (CMIA), it is essential to verify that there are no differences between the methods that can induce an error in leucovorin rescue with dire consequences for the patient. The objective of our study was to compare plasma/serum MTX measurements between CMIA and FPIA (reference method in this study) in the work conditions of a clinical pharmacokinetics unit to determine whether any difference would affect clinical decisions on the management of this drug. Methods: FPIA on TDx/FLx and CMIA on Architect ci8200 were simultaneously used to evaluate 127 clinical samples. Within-run (20 repetitions on same day) and between-run (20 repetitions on different days) imprecision was evaluated using 6 control samples provided by the manufacturer and diluting 2 of them by 50% for 0.03 and 0.22 mmol/L, respectively. The Passing-Bablok regression method, Bland-Altman plot, and concordance correlation coefficient (CCC) were used in the statistical analysis. Results: Within-run imprecision was <5% (3.6%-4.39%) and between-run imprecision,11% (2.42%-10.65%). Between-assay correlation for the studied concentration range (0.05-250 mmol/L) was CMIA = 20.026 + 1.033 FPIA (n = 127), r = 0.9963, and CCC = 0.9946. For samples <1.5 mmol/L (nondiluted) included in the assay calibration curve, the correlation was CMIA = 20.009 + 0.955 FPIA (n = 54), r = 0.9819, and CCC = 0.9807. No significant difference was observed between the measurements by the 2 assays, given that the 95% confidence interval of the ordinate at the origin included "0" (20.020 to 0.0007), and the 95% confidence interval of the slope included 1 (0.923-1.020). The interchangeability of these assays was confirmed by Bland-Altman plot results, which showed a mean difference insignificant at concentrations <10 mu mol/L. Conclusions: The correlation between methods was excellent, and Passing-Bablok regression analysis detected no virtually difference in their results. Utilization of the CMIA-Architect assay to measure MTX concentrations would therefore not affect clinical decisions on MTX management, supporting its employment in routine MTX monitoring.
引用
收藏
页码:492 / 498
页数:7
相关论文
共 11 条
  • [1] Evaluation of a Methotrexate Chemiluminescent Microparticle Immunoassay Comparison to Fluorescence Polarization Immunoassay and Liquid Chromatography-Tandem Mass Spectrometry
    Bouquie, Regis
    Gregoire, Matthieu
    Hernando, Helene
    Azoulay, Claudine
    Dailly, Eric
    Monteil-Ganiere, Catherine
    Pineau, Alain
    Deslandes, Guillaume
    Jolliet, Pascale
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2016, 146 (01) : 119 - 124
  • [2] Multi-site analytical evaluation of a chemiluminescent magnetic microparticle immunoassay (CMIA) for sirolimus on the Abbott ARCHITECT analyzer
    Schmid, Rainer W.
    Lotz, Johannes
    Schweigert, Rosemarie
    Lackner, Karl
    Aimo, Giuseppe
    Friese, Judith
    Rosiere, Thomas
    Dickson, Diana
    Kenney, Daniel
    Maine, Gregory T.
    CLINICAL BIOCHEMISTRY, 2009, 42 (15) : 1543 - 1548
  • [3] Evaluation of a Novel Immunoassay for Lacosamide Therapeutic Drug Monitoring: Comparison With a Liquid Chromatography-Mass Spectrometry Assay
    D'Urso, Annachiara
    Ricotta, Tiziano
    de Grazia, Ugo
    THERAPEUTIC DRUG MONITORING, 2017, 39 (06) : 663 - 668
  • [4] Comparison of four immunoassays to an HPLC method for the therapeutic drug monitoring of methotrexate: Influence of the hydroxylated metabolite levels and impact on clinical threshold
    Descoeur, Juliette
    Dupuy, Anne-Marie
    Bargnoux, Anne-Sophie
    Cristol, Jean-Paul
    Mathieu, Olivier
    JOURNAL OF ONCOLOGY PHARMACY PRACTICE, 2022, 28 (01) : 55 - 63
  • [5] Analytical and clinical evaluation of a new immunoassay for therapeutic drug monitoring of etanercept
    Llinares-Tello, Francisca
    Rosas-Gomez de Salazar, Jose
    Miguel Senabre-Gallego, Jose
    Santos-Soler, Gregorio
    Santos-Ramirez, Carlos
    Salas-Heredia, Esteban
    Molina-Garcia, Juan
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2015, 53 (10) : E279 - E282
  • [6] Therapeutic drug monitoring of perampanel: Clinical utility and impact of co-medication on pharmacokinetic variability
    Lin, Chih-Yin
    Chang, Chun-Wei
    Tseng, Wei-En Johnny
    Wu, Tony
    Cheng, Mei-Yun
    Lee, Chih-Hong
    Chiang, Hsing-, I
    Lin, Wey-Ran
    Lin, Chia-Ni
    Liu, Chun-Jing
    Chen, Po-Ru
    Cheng, Hui-Fen
    Lim, Siew-Na
    HELIYON, 2024, 10 (01)
  • [7] Therapeutic drug monitoring of fenfluramine in clinical practice: Pharmacokinetic variability and impact of concomitant antiseizure medications
    Schoonjans, An-Sofie
    Roosens, Laurence
    Dewals, Wendy
    Paelinck, Bernard P.
    Ceulemans, Berten
    EPILEPSIA, 2022, 63 (03) : 686 - 696
  • [8] Clinical Evaluation of a Novel Stool Antigen Test Using Bioluminescent Enzyme Immunoassay for Detecting Helicobacter pylori
    Kakiuchi, Toshihiko
    Matsuo, Muneaki
    Sakata, Yasuhisa
    Fujimoto, Kazuma
    CANADIAN JOURNAL OF GASTROENTEROLOGY AND HEPATOLOGY, 2022, 2022
  • [9] Novel pharmacokinetic behavior of intravenous busulfan in children with thalassemia undergoing hematopoietic stem cell transplantation: a prospective evaluation of pharmacokinetic and pharmacodynamic profile with therapeutic drug monitoring
    Gaziev, Javid
    Nguyen, Laurent
    Puozzo, Christian
    Mozzi, Alessia Francesca
    Casella, Marialuisa
    Donnorso, Michela Perrone
    Gravina, Paolo
    Sodani, Pietro
    Marziali, Marco
    Isgro, Antonella
    Simone, Maria Domenica
    Andreani, Marco
    Formosa, Amanda
    Testi, Manuela
    Federici, Giorgio
    Bernardini, Sergio
    Lucarelli, Guido
    BLOOD, 2010, 115 (22) : 4597 - 4604
  • [10] Clinical evaluation of a novel and highly sensitive immunoassay for anti-hepatitis B core antigen using a fully automated immunochemical analyzer
    Izumida, Kyo
    Kaneko, Atsushi
    Takahashi, Kazuya
    Kusumoto, Shigeru
    Narita, Tomoko
    Takami, Akiyoshi
    Iida, Shinsuke
    Aoyagi, Katsumi
    Tanaka, Yasuhito
    HEPATOLOGY RESEARCH, 2018, 48 (13) : 1081 - 1091