Method for improving the accuracy of continuous X-ray fluorescence analysis of iron ore mixtures

被引:12
作者
Volkov, A. I. [1 ]
Alov, N. V. [1 ]
机构
[1] Moscow MV Lomonosov State Univ, Fac Chem, Moscow 119992, Russia
关键词
LINE INTENSITY; DIVERGENCE; BEAM;
D O I
10.1134/S1061934810070129
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The dependence of the intensity of X-ray fluorescence on the distance between the spectrometer and the test sample is studied. Changes in the intensity of primary X-radiation, absorption of X-rays by the air, and surface area of the analyzed material are calculated. An efficient method is proposed for improving the accuracy of continuous X-ray fluorescence analysis of iron ore mixtures on the conveyor belt through the exclusion of incorrect results by the dead time of the detection unit. The proposed method is universal and can be used for various loose materials.