k-space undersampling in PROPELLER imaging

被引:44
作者
Arfanakis, K
Tamhane, AA
Pipe, JG
Anastasio, MA
机构
[1] IIT, Dept Biomed Engn, Chicago, IL 60616 USA
[2] Univ Chicago, Brain Res Imaging Ctr, Chicago, IL 60637 USA
[3] Barrow Neurol Inst, MRI Dept, Phoenix, AZ 85013 USA
关键词
PROPELLER; undersampling; k-space; artifacts; acceleration;
D O I
10.1002/mrm.20380
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PROPELLER MRI (periodically rotated overlapping parallel lines with enhanced reconstruction) provides images with significantly fewer B-o-related artifacts than echo-planar imaging (EPI), as well as reduced sensitivity to motion compared to conventional multiple-shot fast spin-echo (FSE). However, the minimum imaging time in PROPELLER is markedly longer than in EPI and 50% longer than in conventional multiple-shot FSE. Often in MRI, imaging time is reduced by undersampling k-space. In the present study, the effects of undersampling on PROPELLER images were evaluated using simulated and in vivo data sets. Undersampling using PROPELLER patterns with reduced number of samples per line, number of lines per blade, or number of blades per acquisition, while maintaining the same k-space field of view (FOVk) and uniform sampling at the edges of FoV(k), reduced imaging time but led to severe image artifacts. In contrast, undersampling by means of removing whole blades from a PROPELLER sampling pattern that sufficiently samples k-space produced only minimal image artifacts, mainly manifested as blurring in directions parallel to the blades removed, even when reducing imaging time by as much as 50%. Finally, undersampling using asymmetric blades and taking advantage of Hermitian symmetries to fill-in the missing data significantly reduced imaging time without causing image artifacts.
引用
收藏
页码:675 / 683
页数:9
相关论文
共 30 条
  • [1] Diffusion-weighted interleaved echo-planar imaging with a pair of orthogonal navigator echoes
    Butts, K
    deCrespigny, A
    Pauly, JM
    Moseley, M
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (05) : 763 - 770
  • [2] Improved image quality and detection of acute cerebral infarction with PROPELLER diffusion-weighted MR imaging
    Forbes, KP
    Pipe, JG
    Karis, JP
    Heiserman, JE
    [J]. RADIOLOGY, 2002, 225 (02) : 551 - 555
  • [3] GUI M, 2004, P 12 ANN M ISMRM KYO, P1283
  • [4] SELECTION OF A CONVOLUTION FUNCTION FOR FOURIER INVERSION USING GRIDDING
    JACKSON, JI
    MEYER, CH
    NISHIMURA, DG
    MACOVSKI, A
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1991, 10 (03) : 473 - 478
  • [5] QUANTIFICATION OF RELATIVE CEREBRAL BLOOD-FLOW CHANGE BY FLOW-SENSITIVE ALTERNATING INVERSION-RECOVERY (FAIR) TECHNIQUE - APPLICATION TO FUNCTIONAL MAPPING
    KIM, SG
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1995, 34 (03) : 293 - 301
  • [6] Reduction of motion artifacts in cine MRI using variable-density spiral trajectories
    Liao, JR
    Pauly, JM
    Brosnan, TJ
    Pelc, NJ
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1997, 37 (04) : 569 - 575
  • [7] QUANTITATIVE-EVALUATION OF SEVERAL PARTIAL FOURIER RECONSTRUCTION ALGORITHMS USED IN MRI
    MCGIBNEY, G
    SMITH, MR
    NICHOLS, ST
    CRAWLEY, A
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1993, 30 (01) : 51 - 59
  • [8] CORRECTION OF MOTIONAL ARTIFACTS IN DIFFUSION-WEIGHTED MR-IMAGES USING NAVIGATOR ECHOES
    ORDIDGE, RJ
    HELPERN, JA
    QING, ZX
    KNIGHT, RA
    NAGESH, V
    [J]. MAGNETIC RESONANCE IMAGING, 1994, 12 (03) : 455 - 460
  • [9] GRASE (GRADIENT-ECHO AND SPIN-ECHO) IMAGING - A NOVEL FAST MRI TECHNIQUE
    OSHIO, K
    FEINBERG, DA
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1991, 20 (02) : 344 - 349
  • [10] Peters DC, 2000, MAGNET RESON MED, V43, P91, DOI 10.1002/(SICI)1522-2594(200001)43:1<91::AID-MRM11>3.0.CO