Independent review of 4DCT scans used for SABR treatment planning

被引:14
作者
Antony, Rachitha [1 ]
Lonski, Peta [1 ]
Ungureanu, Elena [1 ]
Hardcastle, Nicholas [1 ,2 ]
Yeo, Adam [1 ]
Siva, Shankar [3 ,4 ]
Kron, Tomas [1 ,4 ]
机构
[1] Peter MacCallum Canc Ctr, Dept Phys Sci, Melbourne, Vic, Australia
[2] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW, Australia
[3] Peter MacCallum Canc Ctr, Dept Radiat Oncol, Melbourne, Vic, Australia
[4] Univ Melbourne, Sir Peter MacCallum Dept Oncol, Melbourne, Vic, Australia
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2020年 / 21卷 / 03期
关键词
4DCT; motion management; radiotherapy; SABR; BODY RADIATION-THERAPY; STEREOTACTIC ABLATIVE RADIOTHERAPY; LUNG; CANCER; OPTION; IMAGES; MOTION;
D O I
10.1002/acm2.12825
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Four-dimensional computerized tomography (4DCT) is required for stereotactic ablative body radiotherapy (SABR) of mobile targets to account for tumor motion during treatment planning and delivery. In this study, we report on the impact of an image review quality assurance process performed prior to treatment planning by medical physicists for 4DCT scans used for SABR treatment. Reviews were performed of 211 4DCT scans (193 patients) over a 3-yr period (October 2014 to October 2017). Treatment sites included lung (n = 168), kidney/adrenal/adrenal gland (n = 12), rib (n = 4), mediastinum (n = 10), liver (n = 2), T-spine (n = 1), and other abdominal sites (n = 14). It was found that in 23% (n = 49) of cases patient management was altered due to the review process. The most frequent intervention involved patient-specific contouring advice (n = 35 cases, 17%) including adjustment of internal target volume (ITV) margins. In 13 cases (6%) a rescan was requested due to extensive motion artifact rendering the scan inadequate for SABR treatment planning. 4DCT review by medical physicists was found to be an effective method to improve plan quality for SABR.
引用
收藏
页码:62 / 67
页数:6
相关论文
共 25 条
[1]   Review and Uses of Stereotactic Body Radiation Therapy for Oligometastases [J].
Alongi, Filippo ;
Arcangeli, Stefano ;
Filippi, Andrea Riccardo ;
Ricardi, Umberto ;
Scorsetti, Marta .
ONCOLOGIST, 2012, 17 (08) :1100-1107
[2]   Stereotactic ablative radiotherapy versus standard radiotherapy in stage 1 non-small-cell lung cancer (TROG 09.02 CHISEL): a phase 3, open-label, randomised controlled trial [J].
Ball, David ;
Mai, G. Tao ;
Vinod, Shalini ;
Babington, Scott ;
Ruben, Jeremy ;
Kron, Tomas ;
Chesson, Brent ;
Herschtal, Alan ;
Vanevski, Marijana ;
Rezo, Angela ;
Elder, Christine ;
Skala, Marketa ;
Wirth, Andrew ;
Wheeler, Greg ;
Lim, Adeline ;
Shaw, Mark ;
Schofield, Penelope ;
Irving, Louis ;
Solomon, Benjamin .
LANCET ONCOLOGY, 2019, 20 (04) :494-503
[3]   Stereotactic body radiation therapy - A comprehensive review [J].
Chang, Brian K. ;
Timmerman, Robert D. .
AMERICAN JOURNAL OF CLINICAL ONCOLOGY-CANCER CLINICAL TRIALS, 2007, 30 (06) :637-644
[4]   Stereotactic Ablative Body Radiotherapy for the Treatment of Spinal Oligometastases [J].
Chang, J. H. ;
Gandhidasan, S. ;
Finnigan, R. ;
Whalley, D. ;
Nair, R. ;
Herschtal, A. ;
Eade, T. ;
Kneebone, A. ;
Ruben, J. ;
Foote, M. ;
Siva, S. .
CLINICAL ONCOLOGY, 2017, 29 (07) :E119-E125
[5]   The effect of irregular breathing patterns on internal target volumes in four-dimensional CT and cone-beam CT images in the context of stereotactic lung radiotherapy [J].
Clements, N. ;
Kron, T. ;
Franich, R. ;
Dunn, L. ;
Roxby, P. ;
Aarons, Y. ;
Chesson, B. ;
Siva, S. ;
Duplan, D. ;
Ball, D. .
MEDICAL PHYSICS, 2013, 40 (02)
[6]   Stereotactic body radiation therapy: An ablative treatment option for primary and secondary liver tumors [J].
Fuss, M ;
Thomas, CR .
ANNALS OF SURGICAL ONCOLOGY, 2004, 11 (02) :130-138
[7]   Audio-visual biofeedback for respiratory-gated radiotherapy: Impact of audio instruction and audio-visual biofeedback on respiratory-gated radiotherapy [J].
George, Rohini ;
Chung, Theodore D. ;
Vedam, Sastry S. ;
Ramakrishnan, Viswanathan ;
Mohan, Radhe ;
Weiss, Elisabeth ;
Keall, Paul J. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 65 (03) :924-933
[8]   Evaluation of two synchronized external surrogates for 4D CT sorting [J].
Glide-Hurst, Carri K. ;
Smith, Megan Schwenker ;
Ajlouni, Munther ;
Chetty, Indrin J. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (06) :117-132
[9]   Stereotactic Body Radiotherapy for Oligometastatic Disease [J].
Hanna, G. G. ;
Landau, D. .
CLINICAL ONCOLOGY, 2015, 27 (05) :290-297
[10]   Results of patient specific quality assurance for patients undergoing stereotactic ablative radiotherapy for lung lesions [J].
Hardcastle, Nicholas ;
Clements, Natalie ;
Chesson, Brent ;
Aarons, Yolanda ;
Cramb, Jim ;
Siva, Shankar ;
Wanigaratne, Derrick M. ;
Ball, David ;
Kron, Tomas .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2014, 37 (01) :45-52