Accelerator beam data commissioning equipment and procedures: Report of the TG-106 of the Therapy Physics Committee of the AAPM

被引:332
作者
Das, Indra J. [1 ]
Cheng, Chee-Wai [2 ]
Watts, Ronald J. [3 ]
Ahnesjo, Anders [4 ,5 ]
Gibbons, John [6 ]
Li, X. Allen [7 ]
Lowenstein, Jessica [8 ]
Mitra, Raj K. [9 ]
Simon, William E. [10 ]
Zhu, Timothy C. [1 ]
机构
[1] Univ Penn, Dept Radiat Oncol, Philadelphia, PA 19104 USA
[2] Morristown Mem Hosp, Dept Radiat Oncol, Morristown, NJ 07962 USA
[3] Int Med Phys Serv, San Antonio, TX 78232 USA
[4] Uppsala Univ, S-75147 Uppsala, Sweden
[5] Nucletron Scandinavia AB, S-75147 Uppsala, Sweden
[6] Mary Bird Perkins Canc Ctr, Dept Radiat Oncol, Baton Rouge, LA 70809 USA
[7] Med Coll Wisconsin, Dept Radiat Oncol, Milwaukee, WI 53226 USA
[8] Univ Texas MD Anderson Canc Ctr, Radiol Phys Ctr, Houston, TX 77030 USA
[9] Alton Ochsner Med Fdn & Ochsner Clin, Dept Radiat Oncol, New Orleans, LA 70121 USA
[10] Sun Nucl Corp, Melbourne, FL 32940 USA
关键词
accelerator; commissioning; data acquisition;
D O I
10.1118/1.2969070
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
For commissioning a linear accelerator for clinical use, medical physicists are faced with many challenges including the need for precision, a variety of testing methods, data validation, the lack of standards, and time constraints. Since commissioning beam data are treated as a reference and ultimately used by treatment planning systems, it is vitally important that the collected data are of the highest quality to avoid dosimetric and patient treatment errors that may subsequently lead to a poor radiation outcome. Beam data commissioning should be performed with appropriate knowledge and proper tools and should be independent of the person collecting the data. To achieve this goal, Task Group 106 (TG-106) of the Therapy Physics Committee of the American Association of Physicists in Medicine was formed to review the practical aspects as well as the physics of linear accelerator commissioning. The report provides guidelines and recommendations on the proper selection of phantoms and detectors, setting up of a phantom for data acquisition (both scanning and no-scanning data), procedures for acquiring specific photon and electron beam parameters and methods to reduce measurement errors (< 1%), beam data processing and detector size convolution for accurate profiles. The TG-106 also provides a brief discussion on the emerging trend in Monte Carlo simulation techniques in photon and electron beam commissioning. The procedures described in this report should assist a qualified medical physicist in either measuring a complete set of beam data, or in verifying a subset of data before initial use or for periodic quality assurance measurements. By combining practical experience with theoretical discussion, this document sets a new standard for beam data commissioning. (C) 2008 American Association of Physicists in Medicine.
引用
收藏
页码:4186 / 4215
页数:30
相关论文
共 227 条
  • [61] CONDUCTIVITY INDUCED IN INSULATING MATERIALS BY X-RAYS
    FOWLER, JF
    FARMER, FT
    [J]. NATURE, 1954, 173 (4398) : 317 - 318
  • [62] American Association of Physicists in Medicine radiation therapy committee task group 53: Quality assurance for clinical radiotherapy treatment planning
    Fraass, B
    Doppke, K
    Hunt, M
    Kutcher, G
    Starkschall, G
    Stern, R
    Van Dyke, J
    [J]. MEDICAL PHYSICS, 1998, 25 (10) : 1773 - 1829
  • [63] Dose verification of an IMRT treatment planning system with the BEAM EGS4-based Monte Carlo code
    Francescon, P
    Cora, S
    Chiovati, P
    [J]. MEDICAL PHYSICS, 2003, 30 (02) : 144 - 157
  • [64] Total scatter factors of small beams: A multidetector and Monte Carlo study
    Francescon, Paolo
    Cora, Stefania
    Cavedon, Carlo
    [J]. MEDICAL PHYSICS, 2008, 35 (02) : 504 - 513
  • [65] CHARACTERIZATION OF A MULTILEAF COLLIMATOR SYSTEM
    GALVIN, JM
    SMITH, AR
    LALLY, B
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (02): : 181 - 192
  • [66] A comparison of multileaf-collimator and alloy-block field shaping
    Galvin, JM
    Han, K
    Cohen, R
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 40 (03): : 721 - 731
  • [67] Experimental determination of the convolution kernel for the study of the spatial response of a detector
    Garcia-Vicente, F
    Delgado, JM
    Peraza, C
    [J]. MEDICAL PHYSICS, 1998, 25 (02) : 202 - 207
  • [68] Build-up modification of commercial diodes for entrance dose measurements in 'higher energy' photon beams
    Georg, D
    De Ost, B
    Hoornaert, MT
    Pilette, P
    Van Dam, J
    Van Dycke, M
    Huyskens, D
    [J]. RADIOTHERAPY AND ONCOLOGY, 1999, 51 (03) : 249 - 256
  • [69] MEASUREMENT OF DOSE IN THE BUILDUP REGION USING FIXED-SEPARATION PLANE-PARALLEL IONIZATION CHAMBERS
    GERBI, BJ
    KHAN, FM
    [J]. MEDICAL PHYSICS, 1990, 17 (01) : 17 - 26
  • [70] Calculation of enhanced dynamic wedge factors for symmetric and asymmetric photon fields
    Gibbons, JP