Predicting knee replacement damage in a simulator machine using a computational model with a consistent wear factor

被引:48
作者
Zhao, Dong
Sakoda, Hideyuki
Sawyer, W. Gregory
Banks, Scott A. [1 ]
Fregly, Benjamin J. [1 ]
机构
[1] Univ Florida, Dept Mech & Aerosp Engn, Dept Biomed Engn, Dept Orthopaed & Rehabil, Gainesville, FL 32611 USA
来源
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME | 2008年 / 130卷 / 01期
关键词
dynamic contact simulation; computational wear prediction; knee simulator machine; total knee replacement; biomechanics;
D O I
10.1115/1.2838030
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Wear of ultrahigh molecular weight polyethylene remains a primary factor limiting the longevity of total knee replacements (TKRs). However wear testing on a simulator machine is time consuming and expensive, making it impractical for iterative design purposes. The objectives of this paper were first, to evaluate whether a computational model using a wear factor consistent with the TKR material pair can predict accurate TKR damage measured in a simulator machine, and second, to investigate how choice of surface evolution method (fixed or variable step) and material model (linear or nonlinear)affect the prediction. An iterative computational damage model was constructed for a commercial knee implant in an AMTI simulator machine. The damage model combined a dynamic contact model with a surface evolution model to predict how wear plus creep progressively alter tibial insert geometry over multiple simulations. The computational frame work was validated by predicting wear in a cylinder-on-plate system for which an analytical solution was derived. The implant damage model was evaluated for 5 million cycles of simulated gait using damage measurements made on the same implant in an AMTI machine. Using a pin-on-plate wear factor for the same material pair as the implant, the model predicted tibial insert wear volume to within 2% error and damage depths and areas to within 18% and 10% error respectively. Choice of material model had little influence, while inclusion of surface evolution affected damage depth and area but not wear volume predictions. Surface evolution method was important only during the initial cycles, where variable step was needed to capture rapid geometry changes due to the creep. Overall, our results indicate that accurate TKR damage predictions can be made with a computational model using a constant wear factor obtained from pin-on-plate tests for the same material pair and furthermore, that surface evolution method matters only during the initial "break in" period of the simulation.
引用
收藏
页数:10
相关论文
共 31 条
[1]   PRESSURE DISTRIBUTION ON ARTICULAR SURFACES - APPLICATION TO JOINT STABILITY EVALUATION [J].
AN, KN ;
HIMENO, S ;
TSUMURA, H ;
KAWAI, T ;
CHAO, EYS .
JOURNAL OF BIOMECHANICS, 1990, 23 (10) :1013-1020
[2]   THE WEAR OF METALS UNDER UNLUBRICATED CONDITIONS [J].
ARCHARD, JF ;
HIRST, W .
PROCEEDINGS OF THE ROYAL SOCIETY OF LONDON SERIES A-MATHEMATICAL AND PHYSICAL SCIENCES, 1956, 236 (1206) :397-&
[3]   The influence of stress conditions on the wear of UHMWPE for total joint replacements [J].
Barbour, PSM ;
Barton, DC ;
Fisher, J .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1997, 8 (10) :603-611
[4]   Investigation of wear of knee prostheses in a new displacement/force-controlled simulator [J].
Barnett, PI ;
McEwen, HMJ ;
Auger, DD ;
Stone, MH ;
Ingham, E ;
Fisher, J .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2002, 216 (H1) :51-61
[5]  
BARTEL DL, 1995, CLIN ORTHOP RELAT R, P76
[6]   Multibody dynamic simulation of knee contact mechanics [J].
Bei, YH ;
Fregly, BJ .
MEDICAL ENGINEERING & PHYSICS, 2004, 26 (09) :777-789
[7]   ARTICULAR CONTACT IN A 3-DIMENSIONAL MODEL OF THE KNEE [J].
BLANKEVOORT, L ;
KUIPER, JH ;
HUISKES, R ;
GROOTENBOER, HJ .
JOURNAL OF BIOMECHANICS, 1991, 24 (11) :1019-1031
[8]   Development of a six station knee wear simulator and preliminary wear results [J].
Burgess, IC ;
Kolar, M ;
Cunningham, JL ;
Unsworth, A .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1997, 211 (01) :37-47
[9]   The use of a force-controlled dynamic knee simulator to quantify the mechanical performance of total knee replacement designs during functional activity [J].
DesJardins, JD ;
Walker, PS ;
Haider, H ;
Perry, J .
JOURNAL OF BIOMECHANICS, 2000, 33 (10) :1231-1242
[10]  
Endo MM, 1999, BIO-MED MATER ENG, V9, P113