Adjusting the Connection Length of Additively Manufactured Electrodes Changes the Electrochemical and Electroanalytical Performance

被引:26
作者
Crapnell, Robert D. [1 ]
Ferrari, Alejandro Garcia-Miranda [1 ]
Whittingham, Matthew J. [1 ]
Sigley, Evelyn [1 ]
Hurst, Nicholas J. [1 ]
Keefe, Edmund M. [1 ]
Banks, Craig E. [1 ]
机构
[1] Manchester Metropolitan Univ, Fac Sci & Engn, Chester St, Manchester M1 5GD, England
关键词
additive manufacturing; electrochemistry; electrodes; 3D-printing; electroanalysis; 3D-PRINTED GRAPHENE ELECTRODES; FABRICATION; CAPACITANCE;
D O I
10.3390/s22239521
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Changing the connection length of an additively manufactured electrode (AME) has a significant impact on the electrochemical and electroanalytical response of the system. In the literature, many electrochemical platforms have been produced using additive manufacturing with great variations in how the AME itself is described. It is seen that when measuring the near-ideal outer-sphere redox probe hexaamineruthenium (III) chloride (RuHex), decreasing the AME connection length enhances the heterogeneous electrochemical transfer (HET) rate constant (k0) for the system. At slow scan rates, there is a clear change in the peak-to-peak separation (delta Ep) observed in the RuHex voltammograms, with the delta Ep shifting from 118 +/- 5 mV to 291 +/- 27 mV for the 10 and 100 mm electrodes, respectively. For the electroanalytical determination of dopamine, no significant difference is noticed at low concentrations between 10- and 100-mm connection length AMEs. However, at concentrations of 1 mM dopamine, the peak oxidation is shifted to significantly higher potentials as the AME connection length is increased, with a shift of 150 mV measured. It is recommended that in future work, all AME dimensions, not just the working electrode head size, is reported along with the resistance measured through electrochemical impedance spectroscopy to allow for appropriate comparisons with other reports in the literature. To produce the best additively manufactured electrochemical systems in the future, researchers should endeavor to use the shortest AME connection lengths that are viable for their designs.
引用
收藏
页数:12
相关论文
共 56 条
[51]   New conductive filament ready-to-use for 3D-printing electrochemical (bio)sensors: Towards the detection of SARS-CoV-2 [J].
Stefano, Jessica Santos ;
Guterres e Silva, Luiz Ricardo ;
Rocha, Raquel Gomes ;
Brazaca, Lais Canniatti ;
Richter, Eduardo Mathias ;
Abarza Munoz, Rodrigo Alejandro ;
Janegitz, Bruno Campos .
ANALYTICA CHIMICA ACTA, 2022, 1191
[52]   Conductive 3D printed PLA composites: On the interplay of mechanical, electrical and thermal behaviours [J].
Tirado-Garcia, I ;
Garcia-Gonzalez, D. ;
Garzon-Hernandez, S. ;
Rusinek, A. ;
Robles, G. ;
Martinez-Tarifa, J. M. ;
Arias, A. .
COMPOSITE STRUCTURES, 2021, 265
[53]   A Scientist's Guide to Buying a 3D Printer: How to Choose the Right Printer for Your Laboratory [J].
Tully, Joshua J. ;
Meloni, Gabriel N. .
ANALYTICAL CHEMISTRY, 2020, 92 (22) :14853-14860
[54]   Additive manufacturing for electrochemical labs: An overview and tutorial note on the production of cells, electrodes and accessories [J].
Whittingham, Matthew J. ;
Crapnell, Robert D. ;
Rothwell, Emma J. ;
Hurst, Nicholas J. ;
Banks, Craig E. .
TALANTA OPEN, 2021, 4
[55]   Electrochemical Improvements Can Be Realized via Shortening the Length of Screen-Printed Electrochemical Platforms [J].
Whittingham, Matthew J. ;
Hurst, Nicholas J. ;
Crapnell, Robert D. ;
Ferrari, Alejandro Garcia-Miranda ;
Blanco, Elias ;
Davies, Trevor J. ;
Banks, Craig E. .
ANALYTICAL CHEMISTRY, 2021, 93 (49) :16481-16488
[56]   The effect of water ingress on additively manufactured electrodes [J].
Williams, Rhys J. ;
Brine, Tom ;
Crapnell, Robert D. ;
Ferrari, Alejandro Garcia-Miranda ;
Banks, Craig E. .
MATERIALS ADVANCES, 2022, 3 (20) :7632-7639