Repeatability and reproducibility of anterior chamber volume measurements using 3-dimensional corneal and anterior segment optical coherence tomography

被引:25
作者
Fukuda, Shinichi [1 ,2 ]
Kawana, Keisuke [1 ,2 ]
Yasuno, Yoshiaki [2 ,3 ]
Oshika, Tetsuro [1 ]
机构
[1] Univ Tsukuba, Inst Clin Med, Dept Ophthalmol, Tsukuba, Ibaraki 3058575, Japan
[2] Univ Tsukuba, Computat Opt & Ophthalmol Grp, Tsukuba, Ibaraki 3058575, Japan
[3] Univ Tsukuba, Computat Opt Grp, Tsukuba, Ibaraki 3058575, Japan
关键词
ANGLE-CLOSURE GLAUCOMA; HIGH-SPEED; SWEPT-SOURCE; ULTRASOUND BIOMICROSCOPY; EYE; IRIDOTOMY; DEPTH;
D O I
10.1016/j.jcrs.2010.08.053
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To evaluate the repeatability and reproducibility of anterior chamber volume (ACV) measurements using swept-source 3-dimensional corneal and anterior segment optical coherence tomography (CAS-OCT) and dual Scheimpflug imaging. SETTING: Department of Ophthalmology, Institute of Clinical Medicine, University of Tsukuba, Ibaraki, Japan. DESIGN: Nonrandomized clinical trial. METHODS: Measurements were taken in normal eyes (subject group) and in eyes with primary angle closure (PAC) (patient group). In the subject group, the entire ACV and the central 8.0 mm diameter ACV were measured using CAS-OCT and dual Scheimpflug imaging. In the patient group, the entire ACV and 8.0 mm ACV were measured using CAS-OCT. The coefficient of variation and intraclass correlation coefficient (ICC) were calculated to evaluate repeatability and reproducibility, and the correlation between the 2 devices was assessed. RESULTS: In the subject group, the mean 8.0 mm ACV was 110.14 nnm(3) +/- 12.57 (SD) using CAS-OCT and 114.51 +/- 14.69 mm(3) using Scheimpflug imaging; there was a significant linear correlation (r = 0.878, P < .0001). The mean entire ACV on CAS-OCT was 165.15 +/- 29.29 nnm(3). The ICCs of the 8.0 mm and entire ACV measurements were greater than 0.94. The coefficients of repeatability and reproducibility of the 8.0 mm ACV and entire ACV measurements were less than 5%. In the patient group, the 8.0 mm and entire ACV measurements showed good reproducibility and repeatability. CONCLUSIONS: The CAS-OCT method allowed noninvasive measurement of the entire ACV with sufficient repeatability and reproducibility. The 8.0 mm ACV measurements with CAS-OCT and Scheimpflug imaging were comparable.
引用
收藏
页码:461 / 468
页数:8
相关论文
共 27 条
[1]  
[Anonymous], 1994, 57251 ISO
[2]   Determination of the aqueous humor volume by three-dimensional mapping of the anterior chamber [J].
Behndig, A ;
Markström, K .
OPHTHALMIC RESEARCH, 2005, 37 (01) :13-16
[3]   COMPARING METHODS OF MEASUREMENT - WHY PLOTTING DIFFERENCE AGAINST STANDARD METHOD IS MISLEADING [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1995, 346 (8982) :1085-1087
[4]   Sensitivity advantage of swept source and Fourier domain optical coherence tomography [J].
Choma, MA ;
Sarunic, MV ;
Yang, CH ;
Izatt, JA .
OPTICS EXPRESS, 2003, 11 (18) :2183-2189
[5]   Comparison of anterior segment optical coherence tomography and ultrasound biomicroscopy for assessment of the anterior segment [J].
Dada, Tanuj ;
Sihota, Ramanjit ;
Gadia, Ritu ;
Aggarwal, Anand ;
Mandal, Subrata ;
Gupta, Viney .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (05) :837-840
[6]   Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography [J].
de Boer, JF ;
Cense, B ;
Park, BH ;
Pierce, MC ;
Tearney, GJ ;
Bouma, BE .
OPTICS LETTERS, 2003, 28 (21) :2067-2069
[7]  
FONTANA ST, 1980, ARCH OPHTHALMOL-CHIC, V98, P1803
[8]   Anterior Ocular Biometry Using 3-Dimensional Optical Coherence Tomography [J].
Fukuda, Shinichi ;
Kawana, Keisuke ;
Yasuno, Yoshiaki ;
Oshika, Tetsuro .
OPHTHALMOLOGY, 2009, 116 (05) :882-889
[9]   A prospective ultrasound biomicroscopy evaluation of changes in anterior segment morphology after laser iridotomy in Asian eyes [J].
Gazzard, G ;
Friedman, DS ;
Devereux, JG ;
Chew, P ;
Seah, SKL .
OPHTHALMOLOGY, 2003, 110 (03) :630-638
[10]  
*INT ORG STAND, 1994, 57252 ISO