PORTABLE, REAL-TIME, 12-LEAD ECG MONITORING SYSTEM

被引:2
作者
Tseng, Yi-Li [1 ]
Shi, Yo-Zong [1 ]
Jaw, Fu-Shan [1 ]
机构
[1] Natl Taiwan Univ, Inst Biomed Engn, Taipei 10617, Taiwan
关键词
ECG; electrocardiogram; 12-lead; multiplexing; real-time; wireless; TRANSMISSION; ISCHEMIA;
D O I
10.1080/10739149.2010.508331
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This article introduces a wireless 12-lead electrocardiogram (ECG) recording system to monitor high-risk cardiovascular diseases. The design overcomes the limitation of wireless transmission by analog multiplexing. The proposed technique provides three leads of ECG at a time and requires less than 15sec to display the total 12 leads ECG signals on a remote computer. This increases the chances to discover certain acute symptoms earlier. In addition, the synchronous display of three-lead ECG allows the clinician to survey the ECG from the same plane of the heart. The system is designed with low power consumption and is capable of long-term monitoring.
引用
收藏
页码:305 / 312
页数:8
相关论文
共 15 条
[1]   ZigBee and bluetooth - Strengths and weaknesses for industrial applications [J].
Baker, N .
COMPUTING AND CONTROL ENGINEERING, 2005, 16 (02) :20-25
[2]   Continuous ECG monitoring in the management of pre-hospital health emergencies [J].
Chiarugi, F ;
Trypakis, D ;
Kontogiannis, V ;
Lees, PJ ;
Chronaki, CE ;
Zeaki, M ;
Giannakoudakis, N ;
Vourvahakis, D ;
Tsiknakis, M ;
Orphanoudakis, SC .
COMPUTERS IN CARDIOLOGY 2003, VOL 30, 2003, 30 :205-208
[3]   Identifying and reducing noise in psychophysiological recordings [J].
Cutmore, TRH ;
James, DA .
INTERNATIONAL JOURNAL OF PSYCHOPHYSIOLOGY, 1999, 32 (02) :129-150
[4]   Pre-hospital synthesized 12-lead ECG ischemia monitoring with trans-telephonic transmission in acute coronary syndromes -: Pilot study results of the ST SMART trial [J].
Drew, BJ ;
Dempsey, ED ;
Joo, TH ;
Sommargren, CE ;
Glancy, JP ;
Benedict, K ;
Krucoff, MW .
JOURNAL OF ELECTROCARDIOLOGY, 2004, 37 :214-+
[5]   Experience with an ambulatory 12-lead Holter recording system for evaluation of pediatric dysrhythmias [J].
Emmel, M ;
Sreeram, N ;
Schickendantz, S ;
Brockmeier, K .
JOURNAL OF ELECTROCARDIOLOGY, 2006, 39 (02) :188-193
[6]   Design of a low-power microcontroller-based wireless ECG monitoring system [J].
Fariborzi, H. ;
Moghavvemi, M. ;
Mehrkanoon, S. .
2007 5TH STUDENT CONFERENCE ON RESEARCH AND DEVELOPMENT, 2007, :191-194
[7]  
FENSLI R, 2004, P 26 ANN INT C IEEE
[8]   Wireless transmission of ECG signal [J].
Güler N.F. ;
Fidan U. .
Journal of Medical Systems, 2006, 30 (3) :231-235
[9]  
ISTEPANIAN RH, 2006, M HLTH EMERGING MOBI, P389
[10]  
Klootwijk P, 1997, EUR HEART J, V18, P931