Reliability and Validity of the Belt-Stabilized Handheld Dynamometer in Hip- and Knee-Strength Tests

被引:93
作者
Martins, Jaqueline [1 ]
da Silva, Janaina Rodrigues [1 ]
Barbosa da Silva, Marcelo Rodrigues [1 ]
Bevilaqua-Grossi, Debora [1 ]
机构
[1] Univ Sao Paulo, Ribeirao Preto Med Sch, Ave Bandeirantes,3900 Monte Alegre, BR-14049900 Ribeirao Preto, SP, Brazil
基金
巴西圣保罗研究基金会;
关键词
muscle strength; lower extremity; reproducibility of results; PATELLOFEMORAL PAIN SYNDROME; EXTREMITY MUSCLE STRENGTH; QUADRICEPS STRENGTH; ISOKINETIC DYNAMOMETER; ABDUCTOR STRENGTH; RISK-FACTORS; OSTEOARTHRITIS; METAANALYSIS; ASSOCIATION; AGREEMENT;
D O I
10.4085/1062-6050-52.6.04
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
Context: The belt-stabilized handheld dynamometer (HHD) has been used to assess the strength of knee-and hip-muscle groups. However, few researchers have examined its reliability and validity for assessing the strength of these muscles. Objective: To evaluate the intra-examiner reliability of the belt-stabilized HHD and its validity and agreement with the isokinetic dynamometer for assessing the strength of knee-and hip-muscle groups. Design: Cross-sectional study. Setting: University laboratory. Patients or Other Participants: We evaluated 26 healthy participants (13 men, 13 women; age = 23.5 +/- 2.8 years, height = 1.7 +/- 0.1 m, mass = 68.6 +/- 12.4 kg) in 2 sessions using the belt-stabilized HHD and an isokinetic dynamometer for maximum strength of the hip adductors, abductors, flexors, extensors, internal rotators, and external rotators and the knee flexors and extensors. Main Outcome Measure(s): We used reliability values provided by the intraclass correlation coefficient (2,3), standard error of measurement (SEM and percentage SEM), and minimal detectable change; correlation values comparing the belt-stabilized HHD and the isokinetic instrument using the Pearson correlation coefficient (r); and the mean difference in values comparing the 2 instruments using the Bland-Altman method. Results: The intrarater HHD reliability was excellent for most measurements (range = 0.80-0.96; SEM = 1.3-5.3 kilograms of force or 4.8-18.9 Nm, percentage SEM = 7.0%22.0%, minimal detectable change = 3.6-18.8 kilograms of force or 13.2-52.4 Nm) and was moderate only for bilateral knee flexion and left hip internal rotation (intraclass correlation coefficient [ 2,3] = 0.62-0.66 and 0.70, respectively). Correlation with the isokinetic dynamometer was moderate to high (r = 0.60-0.90), but the absolute values did not demonstrate concordance between results using the Bland-Altman method. Conclusions: The belt-stabilized HHD measurements were reliable, and although they did not agree with those from the isokinetic dynamometer, the values were correlated for the hipand knee-muscle groups.
引用
收藏
页码:809 / 819
页数:11
相关论文
共 36 条
[1]   Hip Abductor Strength Reliability and Association With Physical Function After Unilateral Total Knee Arthroplasty: A Cross-Sectional Study [J].
Alnahdi, Ali H. ;
Zeni, Joseph A. ;
Snyder-Mackler, Lynn .
PHYSICAL THERAPY, 2014, 94 (08) :1154-1162
[2]  
[Anonymous], 2015, INTRO MED STAT
[3]   THE RELIABILITY AND VALIDITY OF HANDHELD DYNAMOMETRY FOR THE MEASUREMENT OF LOWER-EXTREMITY MUSCLE STRENGTH IN OLDER ADULTS [J].
Arnold, Cathy M. ;
Warkentin, Kathryn D. ;
Chilibeck, Phimp D. ;
Magnus, Charlene R. A. .
JOURNAL OF STRENGTH AND CONDITIONING RESEARCH, 2010, 24 (03) :815-824
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   A Prospective Investigation of Biomechanical Risk Factors for Patellofemoral Pain Syndrome The Joint Undertaking to Monitor and Prevent ACL Injury (JUMP-ACL) Cohort [J].
Boling, Michelle C. ;
Padua, Darin A. ;
Marshall, Stephen W. ;
Guskiewicz, Kevin ;
Pyne, Scott ;
Beutler, Anthony .
AMERICAN JOURNAL OF SPORTS MEDICINE, 2009, 37 (11) :2108-2116
[6]   Sample size requirements for estimating intraclass correlations with desired precision [J].
Bonett, DG .
STATISTICS IN MEDICINE, 2002, 21 (09) :1331-1335
[7]  
Domholdt E., 2000, PHYS THERAPY RES PRI, V2nd
[8]   Reliability and validity of the Biodex system 3 pro isokinetic dynamometer velocity, torque and position measurements [J].
Drouin, JM ;
Valovich-mcLeod, TC ;
Shultz, SJ ;
Gansneder, BM ;
Perrin, DH .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2004, 91 (01) :22-29
[9]   Preoperative quadriceps strength is a significant predictor of knee function two years after anterior cruciate ligament reconstruction [J].
Eitzen, I. ;
Holm, I. ;
Risberg, M. A. .
BRITISH JOURNAL OF SPORTS MEDICINE, 2009, 43 (05) :371-376
[10]  
Fleiss JL, 2003, DESIGN ANAL CLIN EXP, P7