Investigation on Effect of Image Lag in Fluoroscopic Images Obtained with a Dynamic Flat-panel Detector (FPD) on Accuracy of Target Tracking in Radiotherapy

被引:13
作者
Tanaka, Rie [1 ]
Ichikawa, Katsuhiro [1 ]
Mori, Shinichiro [2 ]
Dobashi, Suguru [2 ]
Kumagai, Motoki [2 ]
Kawashima, Hiroki [3 ]
Minohara, Shinichi [2 ]
Sanada, Sigeru [1 ]
机构
[1] Kanazawa Univ, Coll Med Pharmaceut & Hlth Sci, Sch Hlth Sci, Dept Radiol Technol, Kanazawa, Ishikawa 9200942, Japan
[2] Natl Inst Radiol Sci, Res Ctr Charged Particle Therapy, Inage Ku, Chiba 2638555, Japan
[3] Kanazawa Univ Hosp, Dept Radiol, Kanazawa, Ishikawa 9208641, Japan
关键词
Target tracking; Image lag; Fluoroscopic image; Flat-panel detector (FPD); Radiotherapy; GUIDED RADIOTHERAPY; SYSTEM;
D O I
10.1269/jrr.10059
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Real-time tumor tracking in external radiotherapy can be achieved by diagnostic (kV) X-ray imaging with a dynamic flat-panel detector (FPD). The purpose of this study was to address image lag in target tracking and its influence on the accuracy of tumor tracking. Fluoroscopic images were obtained using a direct type of dynamic FPD. Image lag properties were measured without test devices according to IEC 62220-1. Modulation transfer function (MTF) and profile curves were measured on the edges of a moving tungsten plate at movement rate of 10 and 20 mm/s, covering lung tumor movement of normal breathing. A lung tumor and metal sphere with blurred edge due to image lag was simulated using the results and then superimposed on breathing chest radiographs of a patient. The moving target with and without image lag was traced using a template-matching technique. In the results, the image lag for the first frame after X-ray cutoff was 2.0% and decreased to less than 0.1% in the fifth frame. In the measurement of profile curves on the edges of static and moving tungsten material plates, the effect of image lag was seen as blurred edges of the plate. The blurred edges of a moving target were indicated as reduction of MTF. However, the target could be traced within an error of +/- 5 mm. The results indicated that there was no effect of image lag on target tracking in usual breathing speed in a radiotherapy situation.
引用
收藏
页码:723 / 731
页数:9
相关论文
共 19 条
[1]   Experimental evaluation of a-Se and CdTe flat-panel X-ray detectors for digital radiography and fluoroscopy [J].
Adachi, S ;
Hori, N ;
Sato, K ;
Tokuda, S ;
Sato, T ;
Uehara, K ;
Izumi, Y ;
Nagata, H ;
Yoshimura, Y ;
Yamada, S .
MEDICAL IMAGING 2000: PHYSICS OF MEDICAL IMAGING, 2000, 3977 :38-47
[2]  
BALLARD DH, 1982, COMPUTER VISION, P195
[3]   Lag and ghosting in a clinical flat-panel selenium digital mammography system [J].
Bloomquist, Aili K. ;
Yaffe, Martin J. ;
Mawdsley, Gordon E. ;
Hunter, David M. ;
Beideck, Daniel J. .
MEDICAL PHYSICS, 2006, 33 (08) :2998-3005
[4]   Direct selenium x-ray detector for fluoroscopy, R&F, and radiography [J].
Choquette, M ;
Rougeot, H ;
Martin, JP ;
Laperrière, L ;
Shukri, Z ;
Polischuk, B .
MEDICAL IMAGING 2000: PHYSICS OF MEDICAL IMAGING, 2000, 3977 :128-136
[5]   A method to measure the temporal MTF to determine the DQE of fluoroscopy systems [J].
Friedman, S. N. ;
Cunningham, I. A. .
MEDICAL IMAGING 2006: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2006, 6142
[6]   A small-signal approach to temporal modulation transfer functions with exposure-rate dependence and its application to fluoroscopic detective quantum efficiency [J].
Friedman, S. N. ;
Cunningham, I. A. .
MEDICAL PHYSICS, 2009, 36 (08) :3775-3785
[7]   Dynamic targeting image-guided radiotherapy [J].
Huntzinger, C ;
Munro, P ;
Johnson, S ;
Miettinen, M ;
Zankowski, C ;
Ahlstrom, G ;
Glettig, R ;
Filliberti, R ;
Kaissl, W ;
Kamber, M ;
Amstutz, M ;
Bouchet, L ;
Klebanov, D ;
Mostafavi, H ;
Stark, R .
MEDICAL DOSIMETRY, 2006, 31 (02) :113-125
[8]  
*INT EL COMM, 2008, 622201 IEC
[9]  
International Electrotechnical Commission, 2003, 622201 IEC
[10]   Flat-panel cone-beam computed tomography for image-guided radiation therapy [J].
Jaffray, DA ;
Siewerdsen, JH ;
Wong, JW ;
Martinez, AA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (05) :1337-1349