Micro-CT assessment of comparative radiopacity of adhesive/composite materials in a cylindrical cavity

被引:11
作者
Sumitani, Yuta [1 ]
Hamba, Hidenori [1 ,2 ]
Nakamura, Keiki [1 ]
Sadr, Alireza [3 ]
Nikaido, Toru [1 ]
Tagami, Junji [1 ]
机构
[1] TMDU, Dept Cariol & Operat Dent, Div Oral Hlth Sci, Grad Sch Med & Dent Sci, Tokyo, Japan
[2] Tokyo Dent Coll, Dept Operat Dent Cariol & Pulp Biol, Tokyo, Japan
[3] Univ Washington, Sch Dent, Dept Restorat Dent, Seattle, WA 98195 USA
关键词
mu CT; Radiopacity; Composite restoration; Bovine teeth; OPTICAL COHERENCE TOMOGRAPHY; COMPUTED TOMOGRAPHY; RESIN COMPOSITE; RESTORATIONS; SHRINKAGE; ADAPTATION;
D O I
10.4012/dmj.2017-310
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
This study was performed to evaluate the comparative radiopacity of adhesive/resin composite materials in cylindrical cavities using micro-computed X-ray tomography (mu CT). The two-step self-etch adhesive systems, Clearfil SE Bond (SE) and FL-Bond II (FL), and flowable resin composites, Beautifil Flow F10 (BF) and Clearfil Majesty ES Flow High (MJ), were used. The radiopacity of bovine tooth structures and restorative materials was measured by mu CT. In addition, cylindrical cavities prepared in bovine teeth were restored with the following adhesive/composite combinations: SE-BF, SE-MJ, FL-BF, and FL-MJ. The mean gray values of the composite restorations were calculated. The threshold values of the mu CT images were evaluated using the Otsu's thresholding method. The current results show that the comparative radiopacity of the materials and tooth structure varied, which affected distinguishing the mu CT images of the composite restorations in the cylindrical cavity. The proper combination of restorative materials should be considered when conducting in vitro mu CT assessments of composite restorations.
引用
收藏
页码:634 / 641
页数:8
相关论文
共 21 条
  • [1] Polymerization composite shrinkage evaluation with 3D deformation analysis from μCT images
    Chiang, Yu-Chih
    Roesch, Peter
    Dabanoglu, Alp
    Lin, Chun-Pin
    Hickel, Reinhard
    Kunzelmann, Karl-Heinz
    [J]. DENTAL MATERIALS, 2010, 26 (03) : 223 - 231
  • [2] Mineral concentration of natural human teeth by a commercial micro-CT
    Clementino-Luedemann, Tatiana Nogueira Rocha
    Kunzelmann, Karl-Heinz
    [J]. DENTAL MATERIALS JOURNAL, 2006, 25 (01) : 113 - 119
  • [3] X-RAY MICROTOMOGRAPHY
    ELLIOTT, JC
    DOVER, SD
    [J]. JOURNAL OF MICROSCOPY, 1982, 126 (MAY) : 211 - 213
  • [4] RADIOPACITY OF RESTORATIONS AND DETECTION OF SECONDARY CARIES
    ESPELID, I
    TVEIT, AB
    ERICKSON, RL
    KECK, SC
    GLASSPOOLE, EA
    [J]. DENTAL MATERIALS, 1991, 7 (02) : 114 - 117
  • [5] Enamel Lesion Parameter Correlations between Polychromatic Micro-CT and TMR
    Hamba, H.
    Nikaido, T.
    Sadr, A.
    Nakashima, S.
    Tagami, J.
    [J]. JOURNAL OF DENTAL RESEARCH, 2012, 91 (06) : 586 - 591
  • [6] Effects of CPP-ACP with sodium fluoride on inhibition of bovine enamel demineralization: A quantitative assessment using micro-computed tomography
    Hamba, H.
    Nikaido, T.
    Inoue, G.
    Sadr, A.
    Tagami, J.
    [J]. JOURNAL OF DENTISTRY, 2011, 39 (06) : 405 - 413
  • [7] Non-destructive evaluation of an internal adaptation of resin composite restoration with swept-source optical coherence tomography and micro-CT
    Han, Seung-Hoon
    Sadr, Alireza
    Tagami, Junji
    Park, Sung-Ho
    [J]. DENTAL MATERIALS, 2016, 32 (01) : E1 - E7
  • [8] Shrinkage assessment of low shrinkage composites using micro-computed tomography
    Hirata, Ronaldo
    Clozza, Emanuele
    Giannini, Marcelo
    Farrokhmanesh, Ehsan
    Janal, Malvin
    Tovar, Nick
    Bonfante, Estevam A.
    Coelho, Paulo G.
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2015, 103 (04) : 798 - 806
  • [9] Hotta M, 2009, J ADHES DENT, V11, P207
  • [10] International Organization for Standardization, ISO 4049