Monte Carlo simulation of activity measurement of 123I, 111In and 153Sm with a radionuclide calibrator

被引:6
作者
Olsovcova, Veronika [1 ]
机构
[1] Czech Metrol Inst, Prague 10200, Czech Republic
关键词
Monte Carlo; Radionuclide calibrator; I-123; In-111; (153)sm;
D O I
10.1016/j.apradiso.2009.11.016
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
The response of radionuclide calibrators for radionuclides with high abundances of high energy X-rays is very sensitive to changes in the source geometry. The magnitude of this effect was explored for Bqmetr (Konsorcium BQM, CR) calibrators with I-123, In-111 and Sm-153 in several vial and syringe geometries Dependencies of chamber responses on solution volume and container position were calculated using MCNP transport code. The possibility of usage of an additional copper filter to reduce chamber sensitivity was studied and found to be suitable (C) 2009 Elsevier Ltd All rights reserved.
引用
收藏
页码:1383 / 1387
页数:5
相关论文