Consistency and reproducibility of the VMAT plan delivery using three independent validation methods

被引:7
作者
Chandraraj, Varatharaj [1 ,3 ]
Stathakis, Sotirios [2 ,3 ]
Manickam, Ravikumar [1 ]
Esquivel, Carlos [3 ]
Supe, Sanjay S. [1 ]
Papanikolaou, Nikos [3 ]
机构
[1] Kidwai Mem Inst Oncol, Dept Radiat Phys, Bangalore, Karnataka, India
[2] UTHSCSA, CTRC, Dept Radiat Oncol, San Antonio, TX 78229 USA
[3] Univ Texas Hlth Sci Ctr San Antonio, Dept CTRC, San Antonio, TX USA
基金
美国国家卫生研究院;
关键词
VMAT; quality assurance; DAVID; DynaLog files; PTW seven29 ARRAY; DYNAMIC MULTILEAF COLLIMATION; MODULATED RADIATION-THERAPY; CLINICAL IMPLEMENTATION; QUALITY-ASSURANCE; ARC THERAPY; IMRT; VERIFICATION; RADIOTHERAPY;
D O I
10.1120/jacmp.v12i1.3373
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The complexity of VMAT delivery requires new methods and potentially new tools for the commissioning of these systems. It appears that great consideration is needed for quality assurance (QA) of these treatments since there are limited devices that are dedicated to the QA of rotational delivery. In this present study, we have evaluated the consistency and reproducibility of one prostate and one lung VMAT plans for 31 consecutive days using three different approaches: 1) MLC DynaLog files, 2) in vivo measurements using the multiwire ionization chamber DAVID, and 3) using PTWseven29 2D ARRAY with the OCTAVIUS phantom at our Varian Clinac linear accelerator. Overall, the three methods of testing the reproducibility and consistency of the VMAT delivery were in agreement with each other. All methods showed minimal daily deviations that contributed to clinically insignificant dose variations from day to day. Based on our results, we conclude that the VMAT delivery using a Varian 2100CD linear accelerator equipped with 120 MLC is highly reproducible.
引用
收藏
页码:129 / 140
页数:12
相关论文
共 19 条
[1]   Evaluation of the Delta4 phantom for IMRT and VMAT verification [J].
Bedford, James L. ;
Lee, Young K. ;
Wai, Philip ;
South, Christopher P. ;
Warrington, Alan P. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (09) :N167-N176
[2]   REALIZATION AND VERIFICATION OF 3-DIMENSIONAL CONFORMAL RADIOTHERAPY WITH MODULATED FIELDS [J].
BORTFELD, T ;
BOYER, AL ;
SCHLEGEL, W ;
KAHLER, DL ;
WALDRON, TJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 30 (04) :899-908
[3]   Delivery of intensity-modulated radiation therapy with a conventional multileaf collimator: Comparison of dynamic and segmental methods [J].
Chui, CS ;
Chan, MF ;
Yorke, E ;
Spirou, S ;
Ling, CC .
MEDICAL PHYSICS, 2001, 28 (12) :2441-2449
[4]   Comparison of measured and computed portal dose for IMRT treatment [J].
Cilla, Savino ;
Viola, Pietro ;
Azario, Luigi ;
Grimaldi, Luca ;
Craus, Maurizio ;
D'Onofrio, Guido ;
Fidanzio, Andrea ;
Deodato, Francesco ;
Macchia, Gabriella ;
Digesu, Cinzia ;
Morganti, Alessio G. ;
Piermattei, Angelo .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2006, 7 (03) :65-79
[5]   Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committee [J].
Ezzell, GA ;
Galvin, JM ;
Low, D ;
Palta, JR ;
Rosen, I ;
Sharpe, MB ;
Xia, P ;
Xiao, Y ;
Xing, L ;
Yu, CX .
MEDICAL PHYSICS, 2003, 30 (08) :2089-2115
[6]   IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119 [J].
Ezzell, Gary A. ;
Burmeister, Jay W. ;
Dogan, Nesrin ;
LoSasso, Thomas J. ;
Mechalakos, James G. ;
Mihailidis, Dimitris ;
Molineu, Andrea ;
Palta, Jatinder R. ;
Ramsey, Chester R. ;
Salter, Bill J. ;
Shi, Jie ;
Xia, Ping ;
Yue, Ning J. ;
Xiao, Ying .
MEDICAL PHYSICS, 2009, 36 (11) :5359-5373
[7]  
FRAASS B, 2001, RADIOTHER ONCOL, V60, pS2
[8]   Quality assurance of volumetric modulated arc therapy using Elekta Synergy [J].
Haga, Akihiro ;
Nakagawa, Keiichi ;
Shiraishi, Kenshiro ;
Itoh, Saori ;
Terahara, Atsuro ;
Yamashita, Hideomi ;
Ohtomo, Kuni ;
Saegusa, Shigeki ;
Imae, Toshikazu ;
Yoda, Kiyoshi ;
Pellegrini, Roberto .
ACTA ONCOLOGICA, 2009, 48 (08) :1193-1197
[9]   Verification of dynamic multileaf collimation using an electronic portal imaging device [J].
James, HV ;
Atherton, S ;
Budgell, GJ ;
Kirby, MC ;
Williams, PC .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (02) :495-509
[10]   Commissioning and quality assurance of RapidArc radiotherapy delivery system [J].
Ling, C. Clifton ;
Zhang, Pengpeng ;
Archambault, Yves ;
Bocanek, Jim ;
Tang, Grace ;
LoSasso, Thomas .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (02) :575-581