Flow-injection technique for determination of uranium and thorium isotopes in urine by inductively coupled plasma mass spectrometry

被引:30
作者
Benkhedda, K
Epov, VN
Evans, RD
机构
[1] Trent Univ, Peterborough, ON K9J 7B8, Canada
[2] Vinogradov Inst Geochem SB RAS, Irkutsk 664033, Russia
关键词
uranium; thorium; flow injection; urine; SF-ICP-MS; isotope ratios;
D O I
10.1007/s00216-005-3147-y
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A sensitive and efficient flow-injection (FI) preconcentration and matrix-separation technique coupled to sector field ICP-mass spectrometry (SF-ICP-MS) has been developed and validated for simultaneous determination of ultra-low levels of uranium (U) and thorium (Th) in human urine. The method is based on selective retention of U and Th from a urine matrix, after microwave digestion, on an extraction chromatographic TRU resin, as an alternative to U/TEVA resin, and their subsequent elution with ammonium oxalate. Using a 10 mL sample, the limits of detection achieved for U-238 and Th-232 were 0.02 and 0.03 ng L-1, respectively. The accuracy of the method was checked by spike-recovery measurements. Levels of U and Th in human urine were found to be in the ranges 1.86-5.50 and 0.176-2.35 ng L-1, respectively, well in agreement with levels considered normal for non-occupationally exposed persons. The precision obtained for five replicate measurements of a urine sample was 2 and 3% for U and Th, respectively. The method also enables on-line measurements of the U-235/U-238 isotope ratios in urine. Precision of 0.82-1.04% (RSD) was obtained for U-235/U-238 at low ng L-1 levels, using the FI transient signal approach.
引用
收藏
页码:1596 / 1603
页数:8
相关论文
共 25 条
[1]   Thorium and uranium contents in human urine:: influence of age and residential area [J].
Al-Jundi, J ;
Werner, E ;
Roth, P ;
Höllriegl, V ;
Wendler, I ;
Schramel, P .
JOURNAL OF ENVIRONMENTAL RADIOACTIVITY, 2004, 71 (01) :61-70
[2]   Effect of collisional damping and reactions in a dynamic reaction cell on the precision of isotope ratio measurements [J].
Bandura, DR ;
Baranov, VI ;
Tanner, SD .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2000, 15 (08) :921-928
[3]  
Becker JS, 2002, ATOM SPECTROSC, V23, P177
[4]   Performance optimization of isotope ratio measurements in transient signals by FI-ICP-TOFMS [J].
Carrión, MC ;
Andrés, JR ;
Rubí, JAM ;
Emteborg, H .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2003, 18 (05) :437-443
[5]   Extraction chromatographic separations and analysis of actinides using sequential injection techniques with on-line inductively coupled plasma mass spectrometry (ICP MS) detection [J].
Egorov, OB ;
O'Hara, MJ ;
Farmer, OT ;
Grate, JW .
ANALYST, 2001, 126 (09) :1594-1601
[6]   Measurement of isotope ratios on transient signals by MC-ICP-MS [J].
Günther-Leopold, I ;
Wernli, B ;
Kopajtic, Z ;
Günther, D .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2004, 378 (02) :241-249
[7]   Detection of depleted uranium in urine of veterans from the 1991 Gulf War [J].
Gwiazda, RH ;
Squibb, K ;
McDiarmid, M ;
Smith, D .
HEALTH PHYSICS, 2004, 86 (01) :12-18
[8]   Determining picogram quantities of uranium in urine by isotope dilution inductively coupled plasma mass spectrometry.: Comparison with α-spectrometry [J].
Haldimann, M ;
Baduraux, M ;
Eastgate, A ;
Froidevaux, P ;
O'Donovan, S ;
Von Gunten, D ;
Zoller, O .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2001, 16 (12) :1364-1369
[9]  
HEUMANN KG, 1998, J ANAL ATOM SPECTROM, V13
[10]   SEPARATION AND PRECONCENTRATION OF ACTINIDES FROM ACIDIC MEDIA BY EXTRACTION CHROMATOGRAPHY [J].
HORWITZ, EP ;
CHIARIZIA, R ;
DIETZ, ML ;
DIAMOND, H .
ANALYTICA CHIMICA ACTA, 1993, 281 (02) :361-372