Feasibility of 320-row area detector CT coronary angiography using 40 mL of contrast material: assessment of image quality and diagnostic accuracy

被引:6
|
作者
Kim, Rihyeon [1 ]
Park, Eun-Ah [1 ]
Lee, Whal [1 ]
Chung, Jin Wook [1 ]
机构
[1] Seoul Natl Univ Hosp, Dept Radiol, 101 Daehak Ro, Seoul 03080, South Korea
关键词
Computed tomography coronary angiography; Contrast material; Contrast-induced nephropathy; 320-row area detector CT; Diagnostic accuracy; BODY-MASS INDEX; COMPUTED-TOMOGRAPHY; INDUCED NEPHROPATHY; ENHANCEMENT; ATTENUATION; INJECTION; MEDIA; NEPHROTOXICITY; INTERVENTION; REDUCTION;
D O I
10.1007/s00330-016-4275-9
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To assess the image quality and diagnostic accuracy of 320-row area detector CT (320-ADCT) coronary angiography using 40 mL of contrast material in comparison with 60-mL protocol. This retrospective study included 183 patients who underwent 320-ADCT coronary angiography using 40 mL of contrast and additional 183 sex- and body mass index-matched patients using 60 mL of contrast constituting the control group. Both groups used the same 5-mL/sec injection rate. Quantitative image quality measurements and diagnostic accuracies were calculated and compared. Mean attenuation and contrast-to-noise ratio (CNR) at the aorta and all coronary arteries were lower in the 40-mL group than in the 60-mL group (all, p < 0.05), except for the CNR at proximal coronary arteries at 100 kVp (p = 0.073). However, the proportion of coronary segments with vessel attenuation > 250 HU was not different between groups (all, p > 0.05), except for distal coronary arteries at 80 kVp (p = 0.001). Furthermore, there were no differences in per-patient and per-segment diagnostic accuracies between the groups (all, p > 0.05). 320-ADCT coronary angiography using 40 mL of contrast showed image quality and diagnostic accuracy comparable to the 60-mL protocol, demonstrating the clinical feasibility of lowering the risk of contrast-induced nephropathy through contrast volume reduction. aEuro cent 320-ADCT might enable reduction of contrast material volume. aEuro cent A 40-mL contrast protocol for 320-ADCT provided acceptable image quality. aEuro cent A 40-mL contrast protocol for 320-ADCT demonstrated sufficient diagnostic accuracy.
引用
收藏
页码:3802 / 3810
页数:9
相关论文
共 50 条
  • [21] Diagnostic performance of coronary angiography utilizing intraprocedural 320-row computed tomography with minimal contrast medium
    Misato Chimura
    Tomohito Ohtani
    Kensuke Yokoi
    Tatsuya Shiraki
    Themistoklis Katsimichas
    Takashi Kitao
    Masaki Awata
    Isamu Mizote
    Shungo Hikoso
    Satoru Sumitsuji
    Yasushi Sakata
    Heart and Vessels, 2020, 35 : 1341 - 1348
  • [22] Continuous imaging of esophagogastric junction in patients with reflux esophagitis using 320-row area detector CT: A feasibility study
    Fukazawa, Kousuke
    Furuta, Kenji
    Adachi, Kyoichi
    Shimura, Shino
    Kamiyama, Kazutoshi
    Aimi, Masahito
    Ohara, Shunji
    Kajitani, Takafumi
    Tsurusaki, Masakatsu
    Kitagaki, Hajime
    Kinoshita, Yoshikazu
    JOURNAL OF GASTROENTEROLOGY AND HEPATOLOGY, 2013, 28 (10) : 1600 - 1607
  • [23] Application of Low Tube Voltage, Low-concentration Contrast Agent Using a 320-row CT in Coronary CT Angiography: Evaluation of Image Quality, Radiation Dose and Iodine Intake
    Pan, Yue-ying
    Zhou, Shu-chang
    Wang, Yu-jin
    Li, Qian
    Zhu, Ting-ting
    Liu, Chun-xia
    Guan, Han-xiong
    CURRENT MEDICAL SCIENCE, 2020, 40 (01) : 178 - 183
  • [24] Subtraction coronary CT angiography using second-generation 320-detector row CT
    Yoshioka, Kunihiro
    Tanaka, Ryoichi
    Muranaka, Kenta
    Sasaki, Tadashi
    Ueda, Takanori
    Chiba, Takuya
    Takeda, Kouta
    Sugawara, Tsuyoshi
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2015, 31 : 51 - 58
  • [25] The effect of adaptive iterative dose reduction on image quality in 320-detector row CT coronary angiography
    Tatsugami, F.
    Matsuki, M.
    Nakai, G.
    Inada, Y.
    Kanazawa, S.
    Takeda, Y.
    Morita, H.
    Takada, H.
    Yoshikawa, S.
    Fukumura, K.
    Narumi, Y.
    BRITISH JOURNAL OF RADIOLOGY, 2012, 85 (1016) : E378 - E382
  • [26] The Impact of Different Levels of Adaptive Iterative Dose Reduction 3D on Image Quality of 320-Row Coronary CT Angiography: A Clinical Trial
    Feger, Sarah
    Rief, Matthias
    Zimmermann, Elke
    Martus, Peter
    Schuijf, Joanne Desiree
    Blobel, Jrog
    Richter, Felicitas
    Dewey, Marc
    PLOS ONE, 2015, 10 (05):
  • [27] Can sufficient preoperative information of intracranial aneurysms be obtained by using 320-row detector CT angiography alone?
    Hayashida, Eri
    Sasao, Akira
    Hirai, Toshinori
    Hamasaki, Kiyotoshi
    Nishi, Toru
    Utsunomiya, Daisuke
    Oda, Seitaro
    Iryo, Yasuhiko
    Urata, Joji
    Yamashita, Yasuyuki
    JAPANESE JOURNAL OF RADIOLOGY, 2013, 31 (09) : 600 - 607
  • [28] 320-row CT: does beat-to-beat motion of the coronary arteries affect image quality?
    van der Wall, E. E.
    de Graaf, F. R.
    van Velzen, J. E.
    Jukema, J. W.
    Bax, J. J.
    Schuijf, J. D.
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2012, 28 (01) : 147 - 151
  • [29] The Effect of Bolus Viscosity on Laryngeal Closure in Swallowing: Kinematic Analysis Using 320-Row Area Detector CT
    Inamoto, Yoko
    Saitoh, Eiichi
    Okada, Sumiko
    Kagaya, Hitoshi
    Shibata, Seiko
    Ota, Kikuo
    Baba, Mikoto
    Fujii, Naoko
    Katada, Kazuhiro
    Wattanapan, Pattra
    Palmer, Jeffrey B.
    DYSPHAGIA, 2013, 28 (01) : 33 - 42
  • [30] The Effect of Bolus Viscosity on Laryngeal Closure in Swallowing: Kinematic Analysis Using 320-Row Area Detector CT
    Yoko Inamoto
    Eiichi Saitoh
    Sumiko Okada
    Hitoshi Kagaya
    Seiko Shibata
    Kikuo Ota
    Mikoto Baba
    Naoko Fujii
    Kazuhiro Katada
    Pattra Wattanapan
    Jeffrey B. Palmer
    Dysphagia, 2013, 28 : 33 - 42