Design and prototyping of a handheld 3-DOF laparoscopic ultrasound manipulator for liver surgery

被引:2
作者
Sato, Hideyuki [1 ]
Harada, Kanako [1 ]
Arata, Jumpei [2 ]
Oguri, Susumu [2 ]
Onogi, Shinya [2 ]
Ikeda, Tetsuo [2 ]
Hashizume, Makoto [2 ]
Mitsuishi, Mamoru [1 ]
机构
[1] Univ Tokyo, Sch Engn, Dept Mech Engn, Bunkyo Ku, 7-3-1 Hongo, Tokyo 1138656, Japan
[2] Kyushu Univ, Ctr Adv Med Innovat, Fukuoka 8128582, Japan
来源
SECOND CIRP CONFERENCE ON BIOMANUFACTURING | 2016年 / 49卷
关键词
Biomechatronics; Medical device; Design; NAVIGATION; SYSTEM;
D O I
10.1016/j.procir.2015.09.005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Laparoscopic ultrasound offers noninvasive, real-time, and low-cost intraoperative monitoring of the intra-abdominal organs. However, because of the lack of degrees of freedom in the positioning of the laparoscopic ultrasound probe, it is difficult to align an ultrasound imaging plane with the longitudinal section of a blood vessel in the liver. This paper proposes a handheld laparoscopic ultrasound manipulator with three degrees of freedom designed to manipulate a miniature laparoscopic ultrasound probe. First, an ideal range of motion, measured using sensors and quantified as the required minimum range of motion of the laparoscopic ultrasound probe, was demonstrated by a surgeon. Thereafter, a double-bevel-gear mechanism enabling a pitch motion of +/- 40 degrees and a yaw motion of +/- 30 degrees and a wire -driven mechanism enabling a roll motion of +/- 60 degrees were designed and implemented to the laparoscopic ultrasound manipulator with three degrees of freedom. A mechanism for assembling the miniature laparoscopic ultrasound probe with the shaft of the manipulator under a laparoscopic view was also designed to minimize the number and size of incisions in the abdomen. A prototype of the manipulator with a drive unit was fabricated and tested on an ultrasound liver phantom. A successful assembly, as well as successful visualization of the longitudinal section of a blood vessel in the liver model was demonstrated in a simulated laparoscopic environment. In future, the design will be revised, and the handheld laparoscopic ultrasound manipulator with three degrees of freedom will be tested for in vivo experiments. (C) 2015 The Authors. Published by Elsevier B.V.
引用
收藏
页码:121 / 124
页数:4
相关论文
共 15 条
  • [1] System for robot-assisted real-time laparoscopic ultrasound elastography
    Billings, Seth
    Deshmukh, Nishikant
    Kang, Hyun Jae
    Taylor, Russell
    Boctor, Emad M.
    [J]. MEDICAL IMAGING 2012: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2012, 8316
  • [2] Budde Ricardo P J, 2004, Interact Cardiovasc Thorac Surg, V3, P616, DOI 10.1016/j.icvts.2004.07.009
  • [3] The MIDSTEP system for ultrasound guided remote telesurgery.
    de Cunha, D
    Gravez, P
    Leroy, C
    Maillard, E
    Jouan, J
    Varley, P
    Jones, M
    Halliwell, M
    Hawkes, D
    Wells, PNT
    Angelini, L
    [J]. PROCEEDINGS OF THE 20TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOL 20, PTS 1-6: BIOMEDICAL ENGINEERING TOWARDS THE YEAR 2000 AND BEYOND, 1998, 20 : 1266 - 1269
  • [4] A new visualization technique for laparoscopic ultrasonography
    Ellsmere, J
    Stoll, M
    Wells, W
    Kikinis, R
    Vosburgh, K
    Kane, R
    Brooks, D
    Rattner, D
    [J]. SURGERY, 2004, 136 (01) : 84 - 92
  • [5] A novel approach to the design of a needle driver with multiple DOFs for pediatric laparoscopic surgery
    Fujii, Masahiro
    Sugita, Naohiko
    Ishimaru, Tetsuya
    Iwanaka, Tadashi
    Mitsuishi, Mamoru
    [J]. MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2013, 22 (01) : 9 - 16
  • [6] Harms J, 2001, SURG ENDOSC, V15, P1459, DOI 10.1007/s004640090071
  • [7] Laparoscopic ultrasound navigation in liver surgery: technical aspects and accuracy
    Kleemann, M.
    Hildebrand, P.
    Birth, M.
    Bruch, H. P.
    [J]. SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2006, 20 (05): : 726 - 729
  • [8] A real-time navigation system for laparoscopic surgery based on three-dimensional ultrasound using magneto-optic hybrid tracking configuration
    Konishi, Kozo
    Nakamoto, Masahiko
    Kakeji, Yoshihiro
    Tanoue, Kazuo
    Kawanaka, Hirofumi
    Yamaguchi, Shohei
    Ieiri, Satoshi
    Sato, Yoshinobu
    Maehara, Yoshihiko
    Tamura, Shinichi
    Hashizume, Makoto
    [J]. INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2007, 2 (01) : 1 - 10
  • [9] Leven J, 2005, LECT NOTES COMPUT SC, V3749, P811
  • [10] Oguri S, 2015, IJCARS, V10, pS243