Aortic Insufficiency During Contemporary Left Ventricular Assist Device Support Analysis of the INTERMACS Registry

被引:118
|
作者
Truby, Lauren K. [1 ]
Garan, A. Reshad [1 ]
Givens, Raymond C. [1 ]
Wayda, Brian [1 ]
Takeda, Koji [2 ]
Yuzefpolskaya, Melana [1 ]
Colombo, Paolo C. [1 ]
Naka, Yoshifumi [2 ]
Takayama, Hiroo [2 ]
Topkara, Veli K. [1 ]
机构
[1] Columbia Univ, Coll Phys & Surg, Dept Med, Div Cardiol, New York, NY 10032 USA
[2] Columbia Univ, Coll Phys & Surg, Dept Surg, Div Cardiothorac Surg, New York, NY 10032 USA
基金
美国国家卫生研究院;
关键词
aortic insufficiency; left ventricular assist device; mechanical circulatory support; MANAGEMENT; IMPLANTATION; CLOSURE;
D O I
10.1016/j.jchf.2018.07.012
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVES This study sought to evaluate the impact of moderate to severe aortic insufficiency (AI) on outcomes in patients with continuous flow left ventricular assist devices (CF-LVADs). BACKGROUND Development of worsening Al is a common complication of prolonged CF-LVAD support and portends poor prognosis in single-center studies. Predictors of worsening Al and its impact on clinical outcomes have not been examined in a large cohort. METHODS We conducted a retrospective analysis of patients with CF-LVAD in the INTERMACS (Interagency Registry for Mechanically Assisted Circulatory Support) study. Development of significant AI was defined as the first instance of at least moderate AI Primary outcomes of interest were survival after development of significant AI and time to adverse events, including device complications and rehospitatizations. RESULTS Among 10,603 eligible patients, 1,399 patients on CF-LVAD support developed moderate to severe AI. Prevalence of significant AI progressively increased over time. Predictors of worsening AI included older age, female sex, smaller body mass index, mild pre-implantation AI, and destination therapy strategy. Moderate to severe AI was associated with significantly higher left ventricular end-diastolic diameter, reduced cardiac output, and higher levels of brain natriuretic peptide. Significant AI was associated with higher rates of rehospitalization (32.1% vs. 26.6%, respectively, at 2 years; p 0.015) and mortality (77.2% vs. 71.4%, respectively, at 2 years; p 0.005), conditional upon survival to 1 year. CONCLUSIONS Development of moderate to severe AI has a negative impact on hemodynamics, hospitalizations, and survival on CF-LVAD support. Pre- and post-implantation management strategies should be developed to prevent and treat this complication. (C) 2018 by the American College of Cardiology Foundation.
引用
收藏
页码:951 / 960
页数:10
相关论文
共 50 条
  • [1] Aortic Insufficiency Causes Symptomatic Heart Failure during Left Ventricular Assist Device Support
    Naganuma, Masaaki
    Akiyama, Masatoshi
    Sasaki, Konosuke
    Maeda, Kay
    Ito, Koki
    Suzuki, Tomoyuki
    Katahira, Shintaro
    Suzuki, Yusuke
    Saiki, Yoshikatsu
    TOHOKU JOURNAL OF EXPERIMENTAL MEDICINE, 2021, 255 (03): : 229 - 237
  • [2] Modified Central Closure Technique for Treatment of Aortic Insufficiency in Patients on Left Ventricular Assist Device Support
    Morgan, Jeffrey A.
    Brewer, Robert J.
    ASAIO JOURNAL, 2012, 58 (06) : 626 - 628
  • [3] Residual mitral regurgitation in patients with left ventricular assist device support - An INTERMACS analysis
    Jain, Rashmi
    Truby, Lauren K.
    Topkara, Veli K.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2022, 41 (11): : 1638 - 1645
  • [4] Transcatheter Aortic Valve Replacement for Left Ventricular Assist Device-Induced Aortic Insufficiency
    Kozarek, Katherine
    Minhaj, Mohammed M.
    Chaney, Mark A.
    D'Ancona, Giuseppe
    Pasic, Miralem
    Carrel, Thierry
    Cotter, Elizabeth Healy
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2018, 32 (04) : 1982 - 1990
  • [5] In Vitro Evaluation of Aortic Insufficiency With a Rotary Left Ventricular Assist Device
    Gregory, Shaun D.
    Stevens, Michael C.
    Wu, Eric
    Fraser, John F.
    Timms, Daniel
    ARTIFICIAL ORGANS, 2013, 37 (09) : 802 - 809
  • [6] Inadequate left ventricular unloading during ramp is associated with hospitalization or death during left ventricular assist device support
    Rosenbaum, Andrew N.
    Stulak, John M.
    Clavell, Alfredo L.
    Behfar, Atta
    ARTIFICIAL ORGANS, 2021, 45 (02) : 115 - 123
  • [7] Aortic valve pathophysiology during left ventricular assist device support
    John, Ranjit
    Mantz, Katie
    Eckman, Peter
    Rose, Alan
    May-Newman, Karen
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2010, 29 (12): : 1321 - 1329
  • [8] Moderate Aortic Insufficiency with a Left Ventricular Assist Device Portends a Worse Long-Term Survival
    Auvil, Bryan
    Chung, Jennifer
    Ameer, Alyse
    Han, Jason
    Helmers, Mark
    Birati, Edo
    Acker, Michael
    Atluri, Pavan
    ASAIO JOURNAL, 2020, 66 (07) : 780 - 785
  • [9] Predictors of hospital length of stay after implantation of a left ventricular assist device: An analysis of the INTERMACS registry
    Cotts, William G.
    McGee, Edwin C., Jr.
    Myers, Susan L.
    Naftel, David C.
    Young, James B.
    Kirklin, James K.
    Grady, Kathleen L.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2014, 33 (07): : 682 - 688
  • [10] De Novo Aortic Insufficiency During Long-Term Support on a Left Ventricular Assist Device: A Systematic Review and Meta-Analysis
    Deo, Salil V.
    Sharma, Vikas
    Cho, Yang Hyun
    Shah, Ishan K.
    Park, Soon J.
    ASAIO JOURNAL, 2014, 60 (02) : 183 - 188