Real-time dose-rate monitoring with gynecologic brachytherapy: Results of an initial clinical trial

被引:22
作者
Belley, Matthew D. [1 ]
Craciunescu, Oana [2 ]
Chang, Zheng [2 ]
Langloss, Brian W. [3 ]
Stanton, Ian N. [3 ]
Yoshizumi, Terry T. [2 ,4 ]
Therien, Michael J. [3 ]
Chino, Junzo P. [2 ]
机构
[1] Brown Univ, Alpert Med Sch, Providence, RI 02903 USA
[2] Duke Univ, Med Ctr, Dept Radiat Oncol, Durham, NC USA
[3] Duke Univ, Dept Chem, French Family Sci Ctr, Durham, NC 27706 USA
[4] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27710 USA
基金
美国国家科学基金会;
关键词
In vivo dosimetry; Nanoscale scintillation detector; Real-time brachytherapy quality assurance; vaginal cylinder; HDR; PLASTIC SCINTILLATION DETECTORS; EXTERNAL-BEAM RADIOTHERAPY; IN-VIVO DOSIMETRY; CANCER; THERAPY; IMPACT; LIGHT; BOOST;
D O I
10.1016/j.brachy.2018.07.014
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
PURPOSE: A nanoscintillator-based fiber-optic dosimeter (nanoFOD) was developed to measure real-time dose rate during high-dose-rate (HDR) brachytherapy. A trial was designed to prospectively test clinical feasibility in gynecologic implants. METHODS AND MATERIALS: A clinical trial enrolled women undergoing vaginal cylinder HDR brachytherapy. The nanoFOD was fixed to the cylinder alongside two thermoluminescent dosimeters (TLDs). Treatment was delivered and real-time dose rates captured by the nanoFOD. The nanoFOD and TLD positions were identified in CT images and used to extract the treatment planning system (TPS) calculated dose. The nanoFOD and TLD cumulative doses were compared with the TPS. RESULTS: Nine women were enrolled for 30 fractions, and real-time data were available in 27 treatments. The median ratio of nanoFOD/TPS dose was 1.00 (IQR 0.94-1.02), with a TLD/TPS ratio of 1.01 (IQR 0.98-1.04). Of the nanoFOD dose measurements, 63% were within 5% of the TPS, 26% between 5 and 10% of the TPS, and the remaining 11% between 10 and 20% of the TPS dose. Of the TLD measurements, 70% were within 5% of the TPS, 22% between 5 and 10% of the TPS, and 7% between 10 and 20% of the TPS dose. CONCLUSIONS: Real-time dose-rate measurements during HDR brachytherapy were feasible using the nanoFOD and cumulative dose per fraction showed reasonable agreement to TLD and TPS doses. Additional studies to determine dose thresholds that would yield a low false alarm rate and ongoing device development efforts to improve localization of the scintillator in CT images are needed before this detector should be used to inform clinical decisions.(C) 2018 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:1023 / 1029
页数:7
相关论文
共 31 条
[1]   Radiation-induced light in optical fibers and plastic scintillators: Application to brachytherapy dosimetry [J].
Arnfield, MR ;
Gaballa, HE ;
Zwicker, RD ;
Islam, Q ;
SchmidtUllrich, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (03) :2077-2084
[2]  
Baltas L.S. Dimos., 2006, PHYS MODERN BRACHYTH
[3]   Impact of a higher radiation dose on local control and survival in breast-conserving therapy of early breast cancer:: 10-year results of the randomized boost versus no boost EORTC 22881-10882 trial [J].
Bartelink, Harry ;
Horiot, Jean-Claude ;
Poortmans, Philip M. ;
Struikmans, Henk ;
Van den Bogaert, Walter ;
Fourquet, Alain ;
Jager, Jos J. ;
Hoogenraad, Willem J. ;
Oei, S. Bing ;
Warlam-Rodenhuis, Carla C. ;
Pierart, Marianne ;
Collette, Laurence .
JOURNAL OF CLINICAL ONCOLOGY, 2007, 25 (22) :3259-3265
[4]   WATER-EQUIVALENT PLASTIC SCINTILLATION DETECTORS FOR HIGH-ENERGY BEAM DOSIMETRY .1. PHYSICAL CHARACTERISTICS AND THEORETICAL CONSIDERATIONS [J].
BEDDAR, AS ;
MACKIE, TR ;
ATTIX, FH .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (10) :1883-1900
[5]   Fiber-optic detector for real time dosimetry of a micro-planar x-ray beam [J].
Belley, Matthew D. ;
Stanton, Ian N. ;
Hadsell, Mike ;
Ger, Rachel ;
Langloss, Brian W. ;
Lu, Jianping ;
Zhou, Otto ;
Chang, Sha X. ;
Therien, Michael J. ;
Yoshizumi, Terry T. .
MEDICAL PHYSICS, 2015, 42 (04) :1966-1972
[6]   Real-time in vivo dosimetry for SBRT prostate treatment using plastic scintillating dosimetry embedded in a rectal balloon: a case study [J].
Cantley, Justin L. ;
Cheng, Chee-Wai ;
Jesseph, Fredrick B. ;
Podder, Tarun K. ;
Colussi, Valdir C. ;
Traughber, Bryan J. ;
Ponsky, Lee E. ;
Ellis, Rodney J. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (06) :305-311
[7]   DOSE-VOLUME HISTOGRAM PARAMETERS AND LOCAL TUMOR CONTROL IN MAGNETIC RESONANCE IMAGE-GUIDED CERVICAL CANCER BRACHYTHERAPY [J].
Dimopoulos, Johannes C. A. ;
Lang, Stefan ;
Kirisits, Christian ;
Fidarova, Elena F. ;
Berger, Daniel ;
Georg, Petra ;
Doerr, Wolfgang ;
Poetter, Richard .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (01) :56-63
[8]   DOSE EFFECT RELATIONSHIP FOR LATE SIDE EFFECTS OF THE RECTUM AND URINARY BLADDER IN MAGNETIC RESONANCE IMAGE-GUIDED ADAPTIVE CERVIX CANCER BRACHYTHERAPY [J].
Georg, Petra ;
Poetter, Richard ;
Georg, Dietmar ;
Lang, Stefan ;
Dimopoulos, Johannes C. A. ;
Sturdza, Alina E. ;
Berger, Daniel ;
Kirisits, Christian ;
Doerr, Wolfgang .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 82 (02) :653-657
[9]  
Ghafoori P, 2008, ONCOLOGY-NY, V22, P37
[10]   National Cancer Data Base Analysis of Radiation Therapy Consolidation Modality for Cervical Cancer: The Impact of New Technological Advancements [J].
Gill, Beant S. ;
Lin, Jeff F. ;
Krivak, Thomas C. ;
Sukumvanich, Paniti ;
Laskey, Robin A. ;
Ross, Malcolm S. ;
Lesnock, Jamie L. ;
Beriwal, Sushil .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 (05) :1083-1090